← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K243946

# EffortMed Spinal Fixation System (K243946)

_Effortmed, LLC · NKB · Jan 15, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K243946

## Device Facts

- **Applicant:** Effortmed, LLC
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Jan 15, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Intended Use

EffortMed Spinal Fixation System is intended for use in the noncervical spine. When used as a posterior, noncervical pedicle and non- pedicle fixation system, the EffortMed Spinal Fixation System is intended to provide additional support during fusion using autograft in skeletally mature patients in the following acute and chronic instabilities or deformities: • Degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). - · Spondylolisthesis. - · Trauma (i.e. fracture or dislocation) - · Spinal stenosis - · Curvatures (i.e. scoliosis, kyphosis, and/or lordosis) - · Tumor - · Pseudoarthrosis; and failed previous fusion When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EffortMed Spinal Fixation System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e. scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis. Additionally, the EffortMed Spinal Fixation System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/ spondylolysis,fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. Pediatric pedicle screw fixation is limited to a posterior approach. The EffortMed Spinal Fixation System is intended to be used with autograft and/or allograft.

## Device Story

Top-loading, posterior spinal fixation system; components include polyaxial, monoaxial, cannulated, and spondylolisthesis pedicle screws, rods, connectors, hooks, and setscrews. Used by surgeons in clinical settings to build spinal implant constructs for stabilization and fusion. System provides mechanical support to the spine; implants are non-sterile, single-use. Surgeon selects and assembles components to address specific spinal pathology; output is a rigid construct that promotes fusion. Benefits include stabilization of spinal segments and correction of deformities.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including ASTM F1717 (vertebrectomy model), ASTM F1798 (interconnection mechanisms), and ASTM F543 (bone screw testing), plus axial/torque gripping capacity, flexion-extension moment, and tulip-dissociation testing.

## Technological Characteristics

System components fabricated from titanium alloy (Ti6Al4V-ELI) conforming to ASTM F136. Includes polyaxial, monoaxial, cannulated, and spondylolisthesis screws, rods, connectors, hooks, and setscrews. Mechanical fixation system; non-sterile; single-use.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Msfx MIKRON Spinal Fixation System ([K241657](/device/K241657.md))
- Cortera™ Spinal Fixation System ([K241892](/device/K241892.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 15, 2025

EffortMed, LLC % J.D. Webb President The OrthoMedix Group, Inc. 4313 West 3800 South West Haven, Utah 84401

Re: K243946

Trade/Device Name: EffortMed Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB, KWP Dated: December 18, 2024 Received: December 20, 2024

Dear J.D. Webb:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jisun Yoo-S

for Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K243946

Device Name EffortMed Spinal Fixation System

#### Indications for Use (Describe)

EffortMed Spinal Fixation System is intended for use in the noncervical spine. When used as a posterior, noncervical pedicle and non- pedicle fixation system, the EffortMed Spinal Fixation System is intended to provide additional support during fusion using autograft in skeletally mature patients in the following acute and chronic instabilities or deformities:

• Degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies).

- · Spondylolisthesis.
- · Trauma (i.e. fracture or dislocation)
- · Spinal stenosis
- · Curvatures (i.e. scoliosis, kyphosis, and/or lordosis)
- · Tumor
- · Pseudoarthrosis; and failed previous fusion

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the EffortMed Spinal Fixation System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e. scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis. Additionally, the EffortMed Spinal Fixation System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/ spondylolysis,fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. Pediatric pedicle screw fixation is limited to a posterior approach.

The EffortMed Spinal Fixation System is intended to be used with autograft and/or allograft.

| Type of Use (Select one or both, as applicable)                                    |                                                                                   |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| 510(k) #: K243946                                                                                                                       |                                                          | 510(k) Summary | Prepared on: 2024-12-18 |
|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|----------------|-------------------------|
| Contact Details                                                                                                                         |                                                          |                | 21 CFR 807.92(a)(1)     |
| Applicant Name                                                                                                                          | EffortMed, LLC                                           |                |                         |
| Applicant Address                                                                                                                       | 4766 NW 91st Way Coral Springs FL 33067 United States    |                |                         |
| Applicant Contact Telephone                                                                                                             | 917-566-3479                                             |                |                         |
| Applicant Contact                                                                                                                       | Mr. Steven Brown                                         |                |                         |
| Applicant Contact Email                                                                                                                 | sbrown@effortmed.com                                     |                |                         |
| Correspondent Name                                                                                                                      | The OrthoMedix Group, Inc.                               |                |                         |
| Correspondent Address                                                                                                                   | 4313 W. 3800 S. West Haven UT 84401 United States        |                |                         |
| Correspondent Contact Telephone                                                                                                         | 512-590-5810                                             |                |                         |
| Correspondent Contact                                                                                                                   | Mr. J.D. Webb                                            |                |                         |
| Correspondent Contact Email                                                                                                             | jdwebb@orthomedix.net                                    |                |                         |
| Device Name                                                                                                                             | 21 CFR 807.92(a)(2)                                      |                |                         |
| Device Trade Name                                                                                                                       | EffortMed Spinal Fixation System                         |                |                         |
| Common Name                                                                                                                             | Thoracolumbosacral pedicle screw system                  |                |                         |
| Classification Name                                                                                                                     | Thoracolumbosacral Pedicle Screw System                  |                |                         |
| Regulation Number                                                                                                                       | 888.3070                                                 |                |                         |
| Product Code(s)                                                                                                                         | NKB, KWP                                                 |                |                         |
| Legally Marketed Predicate Devices                                                                                                      | 21 CFR 807.92(a)(3)                                      |                |                         |
| Predicate #                                                                                                                             | Predicate Trade Name (Primary Predicate is listed first) | Product Code   |                         |
| K241657                                                                                                                                 | Msfx MIKRON Spinal Fixation System                       | NKB            |                         |
| K241892                                                                                                                                 | Cortera™ Spinal Fixation System                          | NKB            |                         |
| Device Description Summary                                                                                                              | 21 CFR 807.92(a)(4)                                      |                |                         |
| EffortMed Spinal Fixation System is a top-loading, multiple component, posterior spinal fixation system consisting of polyaxial pedicle |                                                          |                |                         |

screws, monoaxial pedicle screws, cannulated and spondylolisthesis screws, rods (Straight and curved) , connectors , hooks and setscrews. It allows surgeons and functions to build a spinal implant construction to stabilize and promote spinal fusion. The System components are supplied non-sterile, for single use and fabricated from titanium alloy (Ti6Al4V-ELI) that conforms to ASTM F136. Various sizes of these components are available.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

{5}------------------------------------------------

EffortMed Spinal Fixation System is inthe noncervical spine. When used as a posterior, noncervical pedicle and nonpedicle fixation system, the EffortMed System is intended to provide additional support during fusion using autograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities:

Degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies).

- Spondylolisthesis.
- Trauma (i.e. fracture or dislocation)
- Spinal stenosis
- . Curvatures (i.e. scoliosis, kyphosis, and/or lordosis)
- Tumor
- Pseudoarthrosis; and failed previous fusion

When used for posterior non-cervical pedictic patients, the EffortMed Spinal Fixation System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e. scolosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the EffortMed Spinal Fixation System is intended to treat pediatic patients diagnosed with the following conditions: spondylolysis,fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. Pediatric pedicle screw fixation is limited to a posterior approach.

The EffortMed Spinal Fixation System is intended to be used with autograft and/or allograft.

# Indications for Use Comparison

## The subject system has the same indications for use as the Msfx system includes pediatric use. The subject system has the same indications for use as the Cortera system.

## Technological Comparison

### Indications for Use: no difference Materials: no difference Operating Principle: no difference Design/geometry: no difference Dimensions: no difference Material: no difference Sterility: no difference There are no differences between the EffortMed Spinal Fixation System and the predicates

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The following tests were performed in determining substantial equivalence of the EffortMed Spinal Fixation System:

-ASTM F 1717: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model

-ASTM F 1798: Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants

-Axial gripping capacity

-Axial torque gripping capacity

-Flexion-extension moment

-ASTM F543: Test Methods for Metallic Medical Bone Screws

-Annex A3-Test Method for Determining the Axial Pullout Strength of Medical Bone Screws.

-Determining the comparable shear failure force following the equation of Chapman, J.R., et al

-Tulip-dissociation testing on the worst-case bottom-loading screw construct at neutral and maximum screw angulations to evaluate the interconnection strength

Clinical data: Not applicable

The EffortMed Spinal Fixation System is substantially equivalent to the legally marketed predicate device identified above.

## 21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K243946](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K243946)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).