The VERTICALE® Triangular Fixation System, when used in combination with the VERTICALE® Posterior Spinal Fixation System. is intended for immobilization of the spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine. The VERTICALE® Triangular Fixation System is intended to treat the following conditions: trauma (i.e., fracture or dislocation); tumor, pseudoarthrosis; and failed previous fusion. The VERTICALE® Triangular Fixation System is intended for iliac fixation and must be attached to the VERTICALE® Posterior Spinal Fixation System to treat the conditions listed above. Refer to the labeling for the VERTICALE® Posterior Spinal Fixation System for limitations and instructions for use.
Device Story
The VERTICALE® Triangular Fixation System is a spinal implant system consisting of iliac screws featuring a UHMWPE-lined slot and a polyaxial head; used in conjunction with the VERTICALE® Posterior Spinal Fixation System. The device is intended for surgical implantation by a surgeon to provide iliac fixation and spinal segment immobilization. It functions as a mechanical stabilizer to support fusion in patients with spinal instability or deformity. The system is used in an operating room setting. Clinical benefit is derived from the stabilization of the thoracic, lumbar, and sacral spine, facilitating fusion and treating conditions such as trauma, tumors, and pseudoarthrosis.
Clinical Evidence
No clinical data. Performance was demonstrated via bench testing, including ASTM F543 (insertion testing, torsional properties) and ASTM F1717 (static compression bending, static torsion, static tension bending, dynamic compression bending) to confirm mechanical equivalence.
Technological Characteristics
Implants manufactured from titanium alloy (ASTM F136 / ISO 5832-3) and UHMWPE (ASTM F648 / ISO 5834-1 and -2). System includes iliac screws with UHMWPE-lined slots and polyaxial heads. Mechanical performance validated per ASTM F543 and ASTM F1717.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion for acute/chronic instabilities or deformities of the thoracic, lumbar, and sacral spine, specifically for trauma (fracture/dislocation), tumor, pseudoarthrosis, and failed previous fusion. Requires attachment to the VERTICALE® Posterior Spinal Fixation System for iliac fixation.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
VERTICALE® Posterior Spinal Fixation System (K171421)
K100297 — PASS LP SPINAL SYSTEM · Medicrea Technologies · Mar 4, 2010
K062670 — VERTIFLEX SPINAL SCREW SYSTEM · Vertiflex (Tm), Incorporated · Jan 12, 2007
K202505 — GetSet Surgical GoPLF! Posterior Lateral Fusion System · Getset Surgical, SA · Jan 11, 2021
K161499 — VertiForm Posterior Fixation System · Next Orthosurgical · Jul 29, 2016
K022285 — MODIFICATION TO ISOLA SPINAL SYSTEM · Depuyacromed · Aug 13, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
July 6, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Silony Medical GmbH % Mr. Justin Eggleton Vice President, Spine Regulatory Affairs MCRA, LLC 803 7th St, NW Washington, District of Columbia 20001
Re: K223806
Trade/Device Name: VERTICALE® Triangular Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: June 13, 2023 Received: June 13, 2023
Dear Mr. Eggleton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Eileen
Digitally signed by
Eileen Cadel -S
Date: 2023.07.06
Cadel -S 10:14:39 -04'00'
for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223806
Device Name VERTICALE® Triangular Fixation System
#### Indications for Use (Describe)
The VERTICALE® Triangular Fixation System, when used in combination with the VERTICALE® Posterior Spinal Fixation System. is intended for immobilization of the spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine. The VERTICALE® Triangular Fixation System is intended to treat the following conditions: trauma (i.e., fracture or dislocation); tumor, pseudoarthrosis; and failed previous fusion.
The VERTICALE® Triangular Fixation System is intended for iliac fixation and must be attached to the VERTICALE® Posterior Spinal Fixation System to treat the conditions listed above.
Refer to the labeling for the VERTICALE® Posterior Spinal Fixation System for limitations and instructions for use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Contact: | Justin Eggleton<br>Vice President, Head of Musculoskeletal Regulatory Affairs<br>MCRA, LLC<br>803 7th Street NW<br>Washington, DC 20001<br>Phone: 202.552.5800<br>Email: jeggleton@mcra.com |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | June 13, 2023 |
| Device Trade Name: | VERTICALE® Triangular Fixation System |
| Manufacturer: | Silony Medical GmbH<br>Leinfelder Straße 60<br>70771 Leinfelden-Echterdingen<br>Germany |
| Common Name: | Thoracolumbosacral pedicle screw system |
| Classification: | 21 CFR §888.3070, Thoracolumbosacral pedicle screw<br>system |
| Class: | II |
| Product Code: | NKB |
## Indications For Use:
The VERTICALE® Triangular Fixation System, when used in combination with the VERTICALE® Posterior Spinal Fixation System, is intended for immobilization and stabilization of the spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The VERTICALE® Triangular Fixation System is intended to treat the following conditions: trauma (i.e., fracture or dislocation); tumor, pseudoarthrosis; and failed previous fusion.
The VERTICALE® Triangular Fixation System is intended for iliac fixation and must be attached to the VERTICALE® Posterior Spinal Fixation System to treat the conditions listed above.
Refer to the labeling for the VERTICALE® Posterior Spinal Fixation System for limitations and instructions for use.
{4}------------------------------------------------
## Device Description:
The VERTICALE® Triangular Fixation System consists of iliac screws with a UHMWPE lined slot, a polyaxial head as well as related instruments. The VERTICALE® Triangular Fixation System is intended to be used with VERTICALE® Posterior Spinal Fixation System (K171421).
The implants of the VERTICALE Triangular Fixation System are manufactured from titanium alloy conforming to ASTM F136 / ISO 5832-3 and Ultra-High-Molecular-Weight Polyethylene according to ASTM F648 / ISO 5834-1 and -2.
## Predicate Device:
Primary Predicate: VERTICALE® Posterior Spinal Fixation System, Silony Medical, K171421
Reference Device: Attune Knee System, DePuv Synthes, K101433
## Substantial Equivalence:
Non-clinical testing and engineering rationales were provided as justification for substantial equivalence of the VERTICALE® Triangular Fixation System to the predicate devices.
## Performance Testing Summary
The performance testing included tests conducted according to ASTM F543 (insertion testing, torsional properties) and ASTM F1717 (static compression bending, static torsion, static tension bending, dynamic compression bending) used to demonstrate the substantially equivalent mechanical performance of the subject device to legally marketed predicates.
#### Conclusion:
The subject device is substantially equivalent to the predicate devices with respect to indications for use, design, function, materials, and performance. Additionally, the provided performance data related to the subject device supports this traditional 510(k), for the VERTICALE® Triangular Fixation System.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.