Rexious Spinal Fixation System
K222415 · Dio Medical Corporation · NKB · Oct 25, 2022 · Orthopedic
Device Facts
| Record ID | K222415 |
| Device Name | Rexious Spinal Fixation System |
| Applicant | Dio Medical Corporation |
| Product Code | NKB · Orthopedic |
| Decision Date | Oct 25, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3070 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Rexious Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). The Rexious Spinal Fixation System is intended for use as a posterior, non-pedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
Device Story
Top-loading, posterior spinal fixation system; components include pedicle screws, rods, set screws, connectors, hooks, and cross-linking mechanisms. Surgeons assemble implant constructs to stabilize spine and promote fusion. Used in clinical settings by surgeons. System provides mechanical immobilization of spinal segments. Benefits patients by stabilizing acute/chronic instabilities or deformities.
Clinical Evidence
No clinical data or performance testing required; the device is identical to previously cleared predicates (K100765, K113324, K111362, K173131).
Technological Characteristics
Top-loading posterior spinal fixation system. Materials: Titanium/Titanium alloy (ASTM F67, F136, F1295) and Cobalt-Chromium-Molybdenum (ASTM F1537). Components: pedicle screws, rods, set screws, connectors, hooks, cross-links. Non-sterile, single-use.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization and stabilization as an adjunct to fusion for thoracic, lumbar, and sacral spine conditions including degenerative disc disease, spondylolisthesis, fracture, dislocation, scoliosis, kyphosis, spinal tumor, spinal stenosis, and pseudarthrosis.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
- Huvexel Co. Ltd. - Rexious Spinal Fixation System (K173131)
- Huvexel Co. Ltd. - Rexious Hook System (K113324)
- Huvexel Co. Ltd. - Rexious Spinal Fixation System (K100765)
- Huvexel Co. Ltd. - Rexious Spinal Fixation System (K111362)
Related Devices
- K212220 — ArtFX Spinal Fixation System · Artfx Medical, LLC · Apr 20, 2023
- K103147 — NOVEL III SPINAL FIXATION AND ADJUSTABLE BRIDGE SYSTEM · Medyssey Co, Ltd. · Dec 16, 2010
- K092128 — MODIFICATION TO: PSS SYSTEM · Spinal USA · Sep 14, 2009
- K131522 — CORE PEDICLE SCREW SYSTEM · Core-Nexus · Aug 9, 2013
- K210420 — Goblin and Goblin LS Pedicle Screw Systems · Medynus, Inc. · Apr 19, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
October 25, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym and the words "U.S. FOOD & DRUG" are in a blue sans-serif font, with "ADMINISTRATION" in a smaller font below.
Dio Medical Corporation Milan George Vice President of R&D 2100 Campus Lane, Suite 100 East Norriton, Pennsylvania 19403
## Re: K222415
Trade/Device Name: Rexious Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWP Dated: August 5, 2022 Received: August 10, 2022
### Dear Milan George:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222415
Device Name Rexious Spinal Fixation System
#### Indications for Use (Describe)
The Rexious Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
The Rexious Spinal Fixation System is intended for use as a posterior, non-pedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K222415 Page 1 of 2
## 510(k) SUMMARY
# Dio Medical Rexious Spinal Fixation System
| Sponsor: | Manufacturer: | Dio Medical Corp.<br>2100 Campus Lane, Suite 100<br>East Norriton, PA 19403 |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| | Official Contact: | Milan George<br>Email: mgeorge@dio-us.com<br>Phone: 1-877-394-5407 ext.102 |
| Date Prepared: | | August 5, 2022 |
| Device Name: | | Rexious Spinal Fixation System |
| Common Name: | | Pedicle Screw Spinal Fixation System |
| Classification Name: | | Thoracolumbosacral Pedicle Screw System |
| Classification Number: | | 21 CFR 888.3070 |
| Product Code/<br>Classification: | | NKB, KWP, Class II |
| Description: | The Rexious Spinal Fixation System is a top-loading multiple<br>component, posterior spinal fixation system which consists of<br>fixation system which consists of pedicle screws, rods, set<br>screws, connectors, hooks, and transverse (cross) linking<br>mechanisms. | |
| | The Rexious Spinal Fixation System will allow surgeons to build a<br>spinal implant construct to stabilize and promote spinal fusion. The<br>Rexious implant components are supplied non-sterile, single use<br>and fabricated from titanium or titanium alloy (Ti-6Al-4V ELI) as<br>specified in ASTM F67, F136, and F1295 and from Cobalt-<br>Chromium-Molybdenum (CoCr) as specified in ASTM F1537.<br>Various sizes of these implants are available. | |
| Indications<br>for Use: | The Rexious Spinal Fixation System is intended to provide<br>immobilization and stabilization of spinal segments in skeletally<br>mature patients as an adjunct to fusion in the treatment of the<br>following acute and chronic instabilities or deformities of the<br>thoracic, lumbar, and sacral spine: degenerative disc disease; | |
| | spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal<br>tumor; and failed previous fusion (pseudarthrosis). | |
| | The Rexious Spinal Fixation System is intended for use as a<br>posterior, noncervical, non-pedicle system indicated for<br>degenerative disc disease (defined as back pain of discogenic<br>origin with degeneration of the disc confirmed by history and<br>radiographic studies); spondylolisthesis; trauma (i.e., fracture or<br>dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis,<br>and/or lordosis); tumor; pseudoarthrosis; and failed previous<br>fusion. | |
| Predicate<br>Device: | Primary predicate:<br>Huvexel Co. Ltd. - Rexious Spinal Fixation System (K173131) | |
| | Additional predicates:<br>Huvexel Co. Ltd. : Rexious Hook System (K113324)<br>Huvexel Co. Ltd. - Rexious Spinal Fixation System (K100765,<br>K111362) | |
| Substantial<br>Equivalence: | The Rexious Spinal Fixation System is identical to the predicate<br>device, the Huvexel - Rexious Spinal Fixation System and Rexious<br>Hook System. The Subject device has the same intended uses and<br>similar indications, technological characteristics, and principles of<br>operation as its predicate device. There are no technological<br>differences between the Subject device and its predicate devices<br>resulting in no new issues of safety or effectiveness. Thus, the<br>Dio Medical Rexious Spinal Fixation System is identical/<br>substantially equivalent to the predicates. | |
| Performance<br>Data: | The subject and predicate devices are identical and therefore, no<br>performance testing is required. Submission is only transferring<br>name of a system that has already been cleared under K100765,<br>K113324, K111362, and K173131. No testing is required. | |
| Conclusion: | The Dio Medical Rexious Spinal Fixation System has the same<br>intended uses and similar indications, technological<br>characteristics, and principles of operation as its predicate device.<br>Thus, the subject device is identical/substantially equivalent to the<br>predicate device. | |
{4}------------------------------------------------