NextGen Pedicle Screw System

K221776 · Corelink, LLC · NKB · Aug 9, 2022 · Orthopedic

Device Facts

Record IDK221776
Device NameNextGen Pedicle Screw System
ApplicantCorelink, LLC
Product CodeNKB · Orthopedic
Decision DateAug 9, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The CoreLink NextGen Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral/iliac spine (T1 - S1/Ilium): degenerative disc disease (defined as discogenic back pain with degeneration of disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). When used for posterior non-cervical screw fixation in pediatric patients, the CoreLink NextGen Pedicle Screw System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally, the CoreLink NextGen Pedicle Screw System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis, and fracture caused by tumor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft.

Device Story

Thoracolumbosacral pedicle screw system; provides spinal immobilization/stabilization. Components include pedicle screw assemblies, modular tulip heads, modular screw shanks, cross-link connectors, set screws, and spinal rods. Used by surgeons in clinical settings to treat spinal instabilities/deformities. Implants provide mechanical support to spinal segments; facilitate fusion when used with autograft/allograft. System offers various shapes/sizes to accommodate patient anatomy.

Clinical Evidence

Bench testing only. No clinical data provided. Mechanical performance validated via static/dynamic compression bending (ASTM F1717), static torsion (ASTM F1717), axial grip (ASTM F1798), torsion grip (ASTM F1798), and tulip shank disassociation (ASTM F1798).

Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V, ASTM F136) and cobalt chromium alloy (Co-28Cr-6Mo, ASTM F1537). Components: Pedicle screws, modular tulip heads, modular shanks, cross-links, set screws, spinal rods. Mechanical testing per ASTM F1717 and F1798. Non-sterile, intended for surgical implantation.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion for thoracic, lumbar, and sacral/iliac (T1-S1/Ilium) instabilities/deformities (DDD, spondylolisthesis, fracture, dislocation, scoliosis, kyphosis, tumor, pseudarthrosis). Also indicated for pediatric patients (posterior approach only) for adolescent idiopathic scoliosis, spondylolisthesis/spondylolysis, and fracture due to tumor/trauma. Must be used with autograft/allograft.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ August 9, 2022 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. CoreLink, LLC % Nathan Wright Engineer and Regulatory Specialist Empirical Technologies 4628 Northpark Drive Colorado Springs, Colorado 80918 Re: K221776 Trade/Device Name: NextGen Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: June 17, 2022 Received: June 21, 2022 Dear Nathan Wright: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K221776 Device Name NextGen Pedicle Screw System #### Indications for Use (Describe) The CoreLink NextGen Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral/iliac spine (T1 - S1/Ilium): degenerative disc disease (defined as discogenic back pain with degeneration of disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). When used for posterior non-cervical screw fixation in pediatric patients, the CoreLink NextGen Pedicle Screw System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally, the CoreLink NextGen Pedicle Screw System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis, and fracture caused by tumor and/or trauma. Pediatric edicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft. Type of Use (Select one or both, as applicable) � Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection including suggestions for reducing this burden to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) SUMMARY | Submitter's Name: | CoreLink, LLC | | |----------------------------|-------------------------------------------------------------------------------------------|------------------------------------| | Submitter's Address: | 2072 Fenton Logistics Park<br>St. Louis, Missouri 63026 | | | Submitter's Telephone: | 888-349-7808 | | | Contact Person: | Nathan Wright MS<br>Empirical Technologies<br>1-719-351-0248<br>nwright@empiricaltech.com | Image: Empirical Technologies logo | | Date Summary was Prepared: | June 17, 2022 | | | Trade or Proprietary Name: | NextGen Pedicle Screw System | | | Common or Usual Name: | Thoracolumbosacral Pedicle Screw System | | | Classification: | Class II per 21 CFR §888.3070 | | | Product Code: | NKB | | | Classification Panel: | Orthopedic - Spinal Devices (DHT6B) | | ## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION: The NextGen Pedicle Screw System is a thoracolumbosacral pedicle screw system containing metallic implants intended to provide immobilization and stabilization of spinal segments. The system consists of traditional pedicle screw assemblies or modular tulip heads with modular screw shanks with extended tab reduction tulip or closed head tulip options, cross-link connectors, set screws, and spinal rods. Components are offered in various shapes and sizes to meet the requirements of the individual patient anatomy. Implants in the NextGen Pedicle Screw System are manufactured from titanium alloy Ti-6Al-4V per ASTM F136 and cobalt chromium alloy Co-28Cr-6Mo per ASTM F1537. ## INDICATIONS FOR USE The CoreLink NextGen Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral/iliac spine (T1 - S1/Ilium): degenerative disc disease (defined as discogenic back pain with degeneration of disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). When used for posterior non-cervical screw fixation in pediatric patients, the CoreLink NextGen Pedicle Screw System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally, the CoreLink NextGen Pedicle Screw System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis, and fracture caused by tumor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft. ## TECHNOLOGICAL CHARACTERISTICS The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. The subject device is the same or similar to the predicate devices in the following ways: {4}------------------------------------------------ #### ● Indications for Use - Materials - Component Selection ● - Component Sizes . - . Biocompatibility The differences in modular tulip head components do not affect the safety and efficacy of the subject device since mechanical testing was conducted to show substantial equivalence in device bench testing performance. #### Predicate Devices | 510k Number | Trade or Proprietary or Model Name | Manufacturer | Product Code | Predicate Type | |-------------|------------------------------------|-----------------|---------------|----------------| | K210539 | CoreLink Midline Fixation System | CoreLink, LLC | NKB | Primary | | K211323 | ASTRA Spine System | SpineCraft, LLC | NKB, KWP, KWQ | Additional | | K180179 | Firebird Spinal Fixation System | Orthofix Inc. | NKB | Additional | # PERFORMANCE DATA The NextGen Pedicle Screw System has been tested in the following test modes: - Static compression bending per ASTM F1717 ● - Static torsion per ASTM F1717 - Dynamic compression bending per ASTM F1717 - Axial grip per ASTM F1798 - Torsion grip per ASTM F1798 - . Tulip shank disassociation per ASTM F1798 The results of this non-clinical testing show that the strength of the NextGen Pedicle Screw System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices. # CONCLUSION The overall technology characteristics and mechanical performance data lead to the conclusion that the NextGen Pedicle Screw System is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...