← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K220486

# Kentrospine PSS Pedicle Screw System (K220486)

_Rivarp Medical Private Limited · NKB · Jul 22, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K220486

## Device Facts

- **Applicant:** Rivarp Medical Private Limited
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Jul 22, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Kentrospine PSS Pedicle Screw System is intended for posterior, non-cervical fixation as an adjunct to fusion for skeletally mature patients for rigid fixation and stability with following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenoss; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthritis; and/or failed previous fusion.

## Device Story

Kentrospine PSS Pedicle Screw System is a top-loading, posterior spinal fixation system; components include monoaxial/polyaxial pedicle screws, rods, locking caps, and crosslink connectors. Used by surgeons in clinical settings to build spinal implant constructs to stabilize the spine and promote fusion. Implants are made of Ti6Al4V ELI titanium alloy; specialized instruments are made of 316L stainless steel. System provides rigid fixation and stability for various spinal pathologies. Surgeons select appropriate screw/rod sizes to construct the fixation assembly based on patient anatomy and clinical needs.

## Clinical Evidence

No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including static compression bending, static torsional testing, static tension bending, and fatigue compression testing per ASTM F1717-21.

## Technological Characteristics

Top-loading posterior spinal fixation system. Materials: Ti6Al4V ELI titanium alloy (ASTM F136-13(2021)) for implants; 316L stainless steel (ASTM F899-20) for instruments. Components: monoaxial/polyaxial screws, rods, locking caps, crosslink connectors. Supplied non-sterile, single-use. Mechanical testing performed per ASTM F1717-21.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- CD HORIZON® Spinal System ([K091442](/device/K091442.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The full name reads "U.S. Food & Drug Administration".

July 22, 2022

Rivarp Medical Private Limited Sushma N.R Manager - RA No.34, Azeez Sait Industrial Town, 6th Mile, Mysore Road,Nayandahalli. Bangalore, Karnataka 560039 India

Re: K220486

Trade/Device Name: Kentrospine PSS Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: June 29, 2022 Received: July 6, 2022

Dear Sushma N.R:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K220486

Device Name Kentrospine PSS Pedicle Screw System

### Indications for Use (Describe)

The Kentrospine PSS Pedicle Screw System is intended for posterior, non-cervical fixation as an adjunct to fusion for skeletally mature patients for rigid fixation and stability with following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenoss; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthritis; and/or failed previous fusion.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

## Premarket Notification 510(k) Summary as required by Section 21 CFR 807.92

# General Company Information as required by 807.92 (a)

| Submitter's Name /<br>Contact details: | Rivarp Medical Pvt.Ltd.,<br># 34, Azeez Sait Industrial Town, 6th Mile<br>Mysore Road, Nayandahalli,<br>Bangalore -560039, INDIA<br>+91 80 2339 1136<br>info@rivarpmedical.com<br>www.rivarpmedical.com    |                                                                                                                                                                     |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                        | Contact Details:<br>Sushma N.R<br>Manager- Regulatory affairs<br>+91 80 2339 1136                                                                                                                          |                                                                                                                                                                     |
| Summary preparation<br>date:           | 20th July 2022                                                                                                                                                                                             |                                                                                                                                                                     |
| Device name and<br>classification:     | Proprietary/ trade name:<br>Common or Usual Name:<br>Classification Name:<br>Product Code(s):<br>Device Class:<br>Regulation Number:<br>Review Panel:                                                      | Kentrospine PSS Pedicle Screw System<br>Pedicle Screw System<br>Thoracolumbosacral Pedicle Screw System<br>NKB<br>Class II<br>21 CFR Section 888.3070<br>Orthopedic |
| Variants/Types:                        | Kentrospine PSS Pedicle Screw System consists of following<br>components.<br>• PSS Monoaxial Pedicle Screw<br>• PSS Polyaxial Pedicle Screw<br>• PSS Locking Cap<br>• PSS Crosslink Connector<br>• PSS Rod |                                                                                                                                                                     |

{4}------------------------------------------------

#### The identified primary predicate device within this submission is as follows: Identification of the predicate device: CD HORIZON® Spinal System (K091442) manufactured by Medtronic Sofamor Danek USA, dated July 15 2009. The design features, material and indications for use of the Kentrospine PSS Pedicle Screw System are substantially equivalent to the predicate CD HORIZON® Spinal System.

The purpose of this Traditional 510(k) submission is to obtain market clearance Device Description: for the Kentrospine PSS Pedicle Screw System. The Kentrospine PSS Pedicle Screw System is a top-loading multiple component, posterior spinal fixation system which consists of PSS Pedicle Screws (Monoaxial, polyaxial), PSS Rods, PSS Locking Cap, and a PSS Crosslink Connectors.

> The Kentrospine PSS Pedicle Screw System will allow surgeons to build a spinal implant construct to stabilize and promote spinal fusion. Kentrospine PSS Pedicle Screw System components are supplied non-sterile, single use and are made from medical grade. biocompatible Titanium alloy (Ti6Al4V ELI) conforming to ASTM F136-13(2021)- Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications. Various sizes of these implants are available.

> The PSS Pedicle Screws are available in diameters from Ø4.50mm to Ø7.00mm and in lengths of 25mm to 55mm. PSS Rods are straight, made from titanium and available in 05.50mm diameter with length from 45mm, 55mm, 65mm, 75mm, 85mm, 100mm, 125mm, 200mm, 250mm, 300mm, 350mm, 450mm and 500mm.

> Specialized instruments are available for the application and removal of the Kentrospine PSS Pedicle Screw System. The instruments are made from SAE 316L stainless steel which is complied with ASTM F899 - 20 Standard Specification for Wrought Stainless Steels for Surgical Instruments.

## Indications for use:

The Kentrospine PSS Pedicle Screw System is intended for posterior, noncervical fixation as an adjunct to fusion for skeletally mature patients for rigid fixation and stability with following indications: Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed studies): Spondylolisthesis; Trauma (i.e., fracture or dislocation); Spinal stenosis; Curvatures (i.e., scoliosis, kyphosis and/or lordosis); Tumor; Pseudarthritis; and/or failed previous fusion.

{5}------------------------------------------------

| Comparison to<br>Technological<br>Characteristics:  | The Kentrospine PSS Pedicle Screw System is equivalent to the predicate<br>device with respect to intended use, device design and material. All the<br>necessary performance test to determine the mechanical characteristics have<br>been performed on the Kentrospine PSS Pedicle Screw System                                                                                                                                    |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance data:<br>1.Non-clinical<br>Performance: | The following tests have been conducted on the Kentrospine PSS Pedicle<br>Screw System and the device performance has been demonstrated against<br>following applicable standards,                                                                                                                                                                                                                                                  |
|                                                     | 1. Static compression bending testing (ASTM F 1717-21)<br>2. Static Torsional Testing (ASTM F 1717-21)<br>3. Static Tension bending Testing (ASTM F 1717-21)<br>4. Fatigue Compression testing (ASTM F 1717-21)                                                                                                                                                                                                                     |
|                                                     | The results of this non-clinical testing shows that the strength and performance<br>of the Kentrospine PSS Pedicle Screw System is sufficient for its intended use<br>and is substantially equivalent to legally marketed predicate devices.                                                                                                                                                                                        |
|                                                     | Following are the applicable material standards considered for Kentrospine<br>PSS Pedicle Screw System:<br>1. ASTM F136-13(2021)-Standard Specification for Wrought Titanium-<br>6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy (Ti6Al4V ELI) for<br>Surgical Implant Applications.<br>2. ASTM F899 – 20 Standard Specification for Wrought Stainless Steels for<br>Surgical Instruments.                                   |
| 2. Clinical<br>Performance<br>Data/Information:     | No clinical studies were performed.                                                                                                                                                                                                                                                                                                                                                                                                 |
| Conclusion:                                         | The Kentrospine PSS Pedicle Screw System is equivalent to the identified<br>primary predicate device. From the data available we can justify that the<br>Kentrospine PSS Pedicle Screw System has the same indications for use<br>and similar technological characteristics as the legally marketed predicate<br>device. Hence Kentrospine PSS Pedicle Screw System can be considered<br>substantially equivalent to the predicate. |

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K220486](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K220486)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
