← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K202505

# GetSet Surgical GoPLF! Posterior Lateral Fusion System (K202505)

_Getset Surgical, SA · NKB · Jan 11, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K202505

## Device Facts

- **Applicant:** Getset Surgical, SA
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Jan 11, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The GetSet Surgical GoPLF! Posterior Lateral Fusion System is intended for posterior, non-cervical fixation in skeletally mature patients as an adjunct to fusion in the treatment of the following: degenerative disc disease (DDD; defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; tumor; pseudoarthrosis; and failed previous fusion. The GoPLF! Posterior Lateral Fusion System is for prescription use only and is to be used in an open, posterior approach, and can be used with autograft and/or allograft.

## Device Story

System consists of pedicle screws, rods, and associated surgical instruments; used for immobilization and stabilization of spinal segments in thoracolumbar spine and iliosacral fixation. Implants inserted bilaterally into vertebral pedicles and connected via rods to facilitate fusion. Used in open, posterior surgical approach by surgeons. Provides mechanical stabilization to support bone healing; benefits patient by correcting spinal alignment and stabilizing segments affected by degenerative disease, trauma, or tumor. System is MR Conditional.

## Clinical Evidence

No clinical data or animal studies performed. Evidence based on bench testing, including biocompatibility (ISO 10993-1), bacterial endotoxin testing (USP <85>, EP 2.6.14), and mechanical testing (ASTM F1717-15).

## Technological Characteristics

Implants manufactured from Ti-6Al-4V ELI (ASTM F136). System includes pedicle screws, rods, and instruments. MR Conditional per ASTM F2503-20. Provided sterile.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Medtronic CD HORIZON® Spinal System ([K090390](/device/K090390.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 11, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The words "U.S. FOOD & DRUG" are on the first line, and "ADMINISTRATION" is on the second line.

GetSet Surgical, SA % Carolyn Guthrie Director of Regulatory & Quality Kapstone Medical, LLC 520 Elliot St. Charlotte, North Carolina 28202

Re: K202505

Trade/Device Name: GetSet Surgical GoPLF! Posterior Lateral Fusion System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: January 6, 2021 Received: January 7, 2021

#### Dear Carolyn Guthrie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K202505

#### Device Name

GetSet Surgical GoPLF! Posterior Lateral Fusion System

Indications for Use (Describe)

The GetSet Surgical GoPLF! Posterior Lateral Fusion System is intended for posterior, non-cervical fixation in skeletally mature patients as an adjunct to fusion in the treatment of the following: degenerative disc disease (DDD; defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; tumor; pseudoarthrosis; and failed previous fusion.

The GoPLF! Posterior Lateral Fusion System is for prescription use only and is to be used in an open, posterior approach, and can be used with autograft and/or allograft.

| Type of Use (Select one or both, as applicable)                                                          |                                                                                                         |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a logo for GetSet Surgical. The word "Get" is in gray, and the word "Set" is in green. The word "Surgical" is in small gray letters below the word "Set". To the left of the word "Get" is a green graphic of three people holding hands in a circle.

Traditional 510(k) Submission GetSet Surgical GoPLF! Posterior Lateral Fusion System

# 510(k) Summary

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR 807.92, the following summary of information is provided.

#### 1. Date of Preparation:

November 17, 2020

#### 2. Applicant

GetSet Surgical, SA Route de la Corniche 4 1066 Epalinges Switzerland Contact: Ole Stoklund, CEO Phone: +31 733035535.

#### 3. Official Correspondent

Kapstone Medical LLC 520 Elliot St. Charlotte, NC 28202

## 4. Contact Person:

Carolyn Guthrie Email: cguthrie@kapstonemedical.com Phone: +1 704-737-2866

#### 5. Device Name

Trade Name: GetSet Surgical GoPLF! Posterior Lateral Fusion System Common Name: Pedicle Screw System Regulation Description: Thoracolumbosacral Pedicle Screw System Requlations Number: 21 CFR 888.3070 Product Code: NKB Classification: Class II Panel: Orthopedic

## 6. Primary Predicate Device

Medtronic CD HORIZON® Spinal System (K090390)

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for GetSet Surgical. The word "Get" is in gray, and the word "Set" is in green. The word "Surgical" is in smaller green font below the word "Set". To the left of the word "Get" is a green graphic of three people holding hands in a circle.

#### Traditional 510(k) Submission GetSet Surgical GoPLF! Posterior Lateral Fusion System

## 7. Indications for Use:

The GetSet Surgical GoPLF! Posterior Lateral Fusion System is intended for posterior, non-cervical fixation in skeletally mature patients as an adjunct to fusion in the treatment of the following: degenerative disc disease (DDD; defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; tumor; pseudoarthrosis; and failed previous fusion.

The GoPLF! Posterior Lateral Fusion System is for prescription use only and is to be used in an open, posterior approach and can be used with autograft and/or allograft.

## 8. Device Description

#### 8.1 Product Names

GetSet Surgical (GetSet) GoPLF! Posterior Lateral Fusion System (Pedicle screws, rods, and instruments)

#### 8.2 Intended Performance

The GetSet Surgical GoPLF! Posterior Lateral Fusion System is offered sterile and consists of implants and instruments.

The GoPLF! Posterior Lateral Fusion System Implants consist of a variety of sizes of pedicle screws, and rods designed to provide immobilization and stabilization of spinal segments as an adjunct to fusion in skeletally mature patients. The pedicle screws are intended to be inserted bilaterally into the pedicles of each vertebral segment, connected with the rod, and intended for spinal fusion at the selected level. The system conforms to patient anatomy for placement in the thoracolumbar human spine and for iliosacral fixation.

The GoPLF! Posterior Lateral Fusion System is offered with GetSet GoPLF! general and system-specific instruments as accessories. The method of use of the instruments is provided in the Surgical Technique Guide.

## 9. Technological Characteristics

As was established in this submission, the subject device GoPLF! Posterior Lateral Fusion System is substantially equivalent to the predicate device, cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate device through comparison in areas including design, intended use, material composition, function and sterilization method.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a logo for GetSet Surgical. The word "Get" is in gray, and the word "Set" is in green. The word "Surgical" is in small gray letters below the word "Set". To the left of the word "Get" is a graphic of three people in green and gray.

## 10. Performance Data

There are no clinical data generated and held by the manufacturer, i.e., no premarketing or post-market clinical studies or animal studies have been performed. The following information is provided in support of substantial equivalence.

#### 10.1 Biocompatibility

The subject device GoPLF! Posterior Lateral Fusion System is classified as an implant device with tissue/bone contact and permanent contact. Therefore, according to ISO 10993-1 the biological evaluation was assessed for potential effects. The GetSet Surqical GoPLF! Posterior Lateral Fusion devices are manufactured from Ti-6AI-4V ELI conforming to ASTM standard F136. The evaluation was based on product-specific tests and on published literature data. The results show that implants made of Ti-6AI-4V ELI have a high demonstrable biological safety. No concerns arose that would preclude clinical use of GoPLF! Posterior Lateral Fusion System. The instruments are classified as externally communicating medical devices with tissue/bone contact and less than 24 hours contact. All used materials have a medical grade. The requirements of the ISO 10993 standard are fulfilled.

## 10.2 Pyrogenicity / Endotoxin Testing

The bacterial endotoxin test was performed utilizing worst case subject implants to verify that the subject implants meet the acceptance criteria of ≤ 20 EU/kit. Testing was successfully performed, and it was confirmed that the GoPLF! Posterior Lateral Fusion System meets the acceptance criteria of ≤ 20 EU/kit according to USP. General Chapter <85>, Bacterial Endotoxins Test, and EP Ch 2.6.14 Bacterial Endotoxins as recommended in ISO 10993-11:2009 Biological Evaluation of Medical Devices, Part 11: Tests for Systemic Toxicity.

#### 10.3 Mechanical Testing

To demonstrate equivalence to the predicate device, the CD HORIZON® Spinal System, testing was performed according to ASTM F1717-15, "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model". The GoPLF! Pedicle Screw System is geometrically and mechanically equivalent to the predicate device.

#### Magnetic Resonance Safety Testing 10.4

Nonclinical testing has demonstrated that the GetSet Surgical GoPLF! Posterior Lateral Fusion Device range is "MR Conditional" in accordance with the ASTM F2503-20 standard definitions.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image is a logo for GetSet Surgical. The logo features the word "Get" in gray and the word "Set" in green. To the left of the word "Get" is a green graphic of three people holding hands in a circle. Below the word "Set" is the word "Surgical" in small gray letters.

#### 11 Conclusion

The GoPLF! Posterior Lateral Fusion System and the predicate device Medtronic CD HORIZON® Spinal System (K090390) have the same "Indications for Use," are available by prescription only, and are provided sterile. Any technical differences, which were identified, do not result in new questions of safety or effectiveness.

Through assessment of technological characteristics, indications for use and performance data, it can be concluded that GoPLF! Posterior Lateral Fusion System is is substantially equivalent to the CD HORIZON® Spinal System.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K202505](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K202505)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
