← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K201648

# Sterispine PS (K201648)

_Safe Orthopaedics · NKB · Jul 30, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K201648

## Device Facts

- **Applicant:** Safe Orthopaedics
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Jul 30, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Sterispine™ PS system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct of fusion. Sterispine™ PS system is intended for posterior non-cervical, pedicle and non-pedicle fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation) ; spinal stenosis ; tumor ; pseudoarthrosis and failed previous fusion.

## Device Story

Sterispine™ PS is a thoracolumbosacral pedicle screw system; provides spinal immobilization/stabilization as an adjunct to fusion. System includes multiaxial/cannulated multiaxial screws (with/without extended heads), rods, cross-connectors, and single-use surgical instruments. Used in posterior non-cervical spinal surgery; operated by surgeons. Implants/instruments supplied sterile in single-use kits. Fenestrated screws are intended for use only with saline or radiopaque dye; safety not established for bone cement or poor bone quality patients. Device provides mechanical fixation to stabilize spinal segments; aids in fusion process; benefits patients by addressing spinal instability/deformity.

## Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F1798 (axial, flexural, torsional grip; dynamic flexion) and ASTM F1717 (static/dynamic compression, static tension, static torsion). Functional evaluation conducted on cross-connectors and single-use surgical instruments (Jamshidi, introducers, spatulas, screw extenders, screwdrivers, persuaders, handles). Results confirmed all components meet functional requirements, assemble correctly, and show no deterioration or blocking.

## Technological Characteristics

Thoracolumbosacral pedicle screw system. Components: multiaxial/cannulated screws (Ø4.5, 5.5, 6.5, 7.5mm), rods (Ø5.5, 6.5, 7.5mm), cross-connectors. Supplied sterile. Pyrogenicity testing per USP standard (Endotoxin limit 20 EU/device). Mechanical testing per ASTM F1798 and ASTM F1717. Single-use surgical instrumentation kit.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Sterispine™ PS ([K170528](/device/K170528.md))
- CD HORIZON™ Fenestrated Screw Set ([K170347](/device/K170347.md))
- Synergy 4.0 mm, 4.5 mm VLS screws ([K011437](/device/K011437.md))
- Sterispine™ PS ([K151921](/device/K151921.md))
- Sterispine™ PS ([K151747](/device/K151747.md))
- Sterispine™ PS ([K150092](/device/K150092.md))
- Sterispine™ PS ([K140802](/device/K140802.md))
- Sterispine™ PS ([K130632](/device/K130632.md))
- Sterispine™ PS ([K121299](/device/K121299.md))
- Sterispine™ PS ([K112453](/device/K112453.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized depiction of a human figure embracing a globe. The FDA acronym is in a blue square, and the full name, "U.S. Food & Drug Administration," is in blue text to the right of the square.

Safe Orthopaedics Pierre Dumouchel CEO Parc des Bellevues, Le Californie - Allee Rosa Luxemburg Eragny Sur Oise, 95610 France

Re: K201648

Trade/Device Name: Sterispine™ PS Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWQ Dated: June 16, 2020 Received: June 18, 2020

Dear Pierre Dumouchel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

July 30, 2020

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Colin O'Neill, M.B.E. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201648

Device Name Sterispine™ PS

#### Indications for Use (Describe)

The Sterispine™ PS system is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct of fusion. Sterispine™ PS system is intended for posterior non-cervical, pedicle and nonpedicle fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation) ; spinal stenosis ; tumor ; pseudoarthrosis and failed previous fusion.

| Type of Use (Select one or both, as applicable)                                                 |                                                                                     |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Traditional 510k STERISPINE™ PS

Image /page/3/Picture/2 description: The image shows the logo for Safe Orthopaedics. The logo features a stylized letter "S" in a light green color, with a small dot to the right of the "S". Below the "S", the words "Safe Orthopaedics" are written, with "Safe" in green and "Orthopaedics" in gray.

## 510(k) SUMMARY

| 510k                                  | Traditional                                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Basis for submission                  | Extension of the product line                                                                                                                                                                                                                                                                                                                                                                               |
| Submitted by                          | Safe Orthopaedics<br>Parc des Bellevues<br>Le Californie - Allée Rosa Luxemburg<br>95610 Eragny-sur-Oise, France                                                                                                                                                                                                                                                                                            |
| Contacts                              | Pierre DUMOUCHEL (CEO)<br>Phone number 33 (0) 1 84 28 01 79<br>e-mail p.dumouchel@safeorthopaedics.com<br>Regulatory contact: Dr Isabelle DRUBAIX (Idée Consulting) idrubaix@nordnet.fr                                                                                                                                                                                                                     |
| Date Prepared                         | July 23, 2020                                                                                                                                                                                                                                                                                                                                                                                               |
| Common Name                           | Pedicle screw spinal system                                                                                                                                                                                                                                                                                                                                                                                 |
| Trade Name                            | Sterispine™ PS                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification Name                   | Thoracolumbosacral pedicle screw system                                                                                                                                                                                                                                                                                                                                                                     |
| Class                                 | II                                                                                                                                                                                                                                                                                                                                                                                                          |
| Product Code                          | NKB, KWQ                                                                                                                                                                                                                                                                                                                                                                                                    |
| CFR section                           | 888.3070                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device panel                          | ORTHOPEDIC                                                                                                                                                                                                                                                                                                                                                                                                  |
| Legally marketed<br>predicate devices | Primary predicate: Sterispine™ PS manufactured by SAFE ORTHOPAEDICS (K170528)<br>Additional predicates: CD HORIZON™ Fenestrated Screw Set manufactured by<br>Medtronic Sofamor Danek (K170347), Synergy 4.0 mm, 4.5 mm VLS screws<br>manufactured by Interpore Cross Intl (K011437), Sterispine™ Ps manufactured by<br>SAFE ORTHOPAEDICS (K151921, K151747, K150092, K140802, K130632, K121299,<br>K112453) |

{4}------------------------------------------------

| Indications for use                                                      | The Sterispine™ PS system is intended to provide immobilization and stabilization<br>of spinal segments in skeletally mature patients as an adjunct of fusion.<br>Sterispine™ PS system is intended for posterior non-cervical, pedicle and non-<br>pedicle fixation for the following indications: degenerative disc disease (DDD)<br>(defined as back pain of discogenic origin with degeneration of the disc confirmed<br>by history and radiographic studies); spondylolisthesis; trauma (i.e, fracture or<br>dislocation) ; spinal stenosis ; tumor ; pseudoarthrosis and failed previous fusion.                                                                                                                                                                                                                                                                                  |
|--------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description of the<br>device                                             | The cleared range of Sterispine™ PS system includes multiaxial screws and<br>cannulated multiaxial screws with or without extended head (Ø5.5, 6.5 and<br>7.5mm, lengths from 25 to 85mm) and straight and prebent rods (Ø5.5, 6.5 and<br>7.5mm, lengths from 30 to 380mm). Sterispine™ PS implants are supplied sterile<br>with a sterile single-use set of surgical instruments supplied in kits. Bacterial<br>endotoxin testing as specified in USP standard is used for pyrogenicity testing to<br>achieve the Endotoxin limit of 20 EU / device.                                                                                                                                                                                                                                                                                                                                   |
| Technological<br>characteristics<br>compared to the<br>predicate devices | The present submission is an extension of the product line that includes<br>- Modified and new sterile single-use surgical instruments<br>- Add of a cross-connector kit<br>- Add of multi-axial fenestrated screws (with and without extended head)<br>- Add of Ø4.5 mm for all STERISPINE™ PS screws<br>As established in this submission, the modified or added components are<br>substantially equivalent to predicate devices in areas including indications for use,<br>function materials in contact with patients and technological and mechanical<br>characteristics.<br>The safety and effectiveness of Sterispine™ PS fenestrated screws have not been<br>established when used in conjunction with bone cement or for use in patients with<br>poor bone quality (e.g., osteoporosis, osteopenia). This device is intended only to<br>be used with saline or radiopaque dye. |
| Discussion of Testing                                                    | The following non-clinical tests were conducted on the STERISPINE™ PS system<br>Mechanical testing of fenestrated screw φ4.5 mm / with cross -connector where<br>applicable):<br>- Axial Grip, Flexural Grip, Torsional Grip and Dynamic Flexion per ASTM F1798<br>- Static Compression Bending, Static Tension Bending, Static Torsion and<br>Dynamic compression per ASTM F1717.<br>Results demonstrate that the Sterispine™ PS system meets or exceeds functional<br>requirements and suitability for use. Additionally, the Sterispine™ PS system<br>performs equivalent or better than predicates systems.<br>- Functional evaluation of the cross-connector and associated surgical<br>instrumentation<br>All tests met the acceptance criteria<br>- Functional testing of the single-use surgical instruments                                                                    |

{5}------------------------------------------------

| Overall functional testing demonstrates that the new added surgical instruments can be assembled with each component of the Sterispine PS instrument/implant kit. No deterioration or blocking were observed during the functional testing. Additional mechanical and / or functional testings conducted on Jamshidi, threaded rod introducer, spatula, screw extender arm, fenestrated screwdriver, persuader and handle show that surgical instruments are suitable and fulfill requirements for single-use. |                                                                                                                                                                                                                                                                                                                           |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Based on the design features, technological characteristics, feature comparisons, indications for use, and non-clinical performance testing, the added surgical implants and the added or modified sterile single-use surgical instruments have demonstrated substantial equivalence to the identified predicate devices. |

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K201648](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K201648)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
