← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K200666

# Stryker Xia 3 Power Adaptor (K200666)

_Stryker Spine · NKB · Apr 23, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K200666

## Device Facts

- **Applicant:** Stryker Spine
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Apr 23, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Pediatric

## Intended Use

The intended use for the Xia 3 Power Adaptor is to facilitate the placement of pedicle screws using the power technique (corded and cordless), the use of the Xia 3 Power Adaptor is intended for exclusive use with the Stryker Cordless and Corded Power Drivers. When the power adaptors are attached, the Stryker Power Drivers (corded and cordless) provide power to rotate screwdrivers for the insertion of pedicle screws. Pedicle screws from select Stryker Spine implant systems may be implanted in the non-cervical spine using powered (corded and cordless) instrumentation. The systems included are the family of XIA® Spinal Systems (XIA® Stainless Steel, XIA® II, XIA® Anterior, and XIA® Precision), XIA® 3 Spinal System, XIA® 4.5 Spinal System, Radius® Spinal System, MANTIS® Spinal System, MANTIS® Redux Spinal System, and the ES2® Spinal System.

## Device Story

Xia 3 Power Adaptor is a mechanical instrument accessory; serves as an interface between Stryker corded/cordless power drivers and manual screwdriver instruments. Used in clinical settings by surgeons to facilitate pedicle screw insertion during spinal fixation procedures. Device transforms rotational power from the driver to the screwdriver; enables powered insertion of screws into the non-cervical spine. Benefits include improved efficiency in screw placement compared to manual techniques. Does not alter the indications for use of the associated spinal implant systems.

## Clinical Evidence

Bench testing only. Engineering analysis was performed to verify the mechanical interface and compatibility with Stryker power drivers, demonstrating that the adaptor does not represent a new worst-case scenario or affect system performance.

## Technological Characteristics

Mechanical instrument accessory; serves as a physical interface between power drivers and screwdrivers. Utilizes established materials consistent with predicate devices. Designed for use with Stryker corded and cordless power drivers. No electronic components or software algorithms; purely mechanical function.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- [K172724](/device/K172724.md)
- [K170496](/device/K170496.md)
- [K152632](/device/K152632.md)
- [K120434](/device/K120434.md)
- [K122845](/device/K122845.md)
- [K111478](/device/K111478.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 23, 2020

Stryker Spine Megan Callanan Regulatory Affairs Specialist 2 Pearl Court Allendale, New Jersey 07401

Re: K200666

Trade/Device Name: Stryker Xia 3 Power Adaptor Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: March 11, 2020 Received: March 13, 2020

Dear Megan Callanan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, MBE Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K200666

Device Name Xia 3 Power Adaptor

## Indications for Use (Describe)

The XIA® Spinal System is intended for anterior/anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvature (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

The XIA® 3, RADIUS® Spinal Systems are intended for use in the non-cervical spine. When used as an anterior/ anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the XIA® II, XIA® 3, and RADIUS® Spinal Systems are intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis,kyphosis); tumor; pseudoarthrosis; and failed previous fusion.

The XIA® 4.5 Spinal System is intended for anterolateral and posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); spondylolisthesis; trauma (i.e. fracture of dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudarthrosis; failed previous fusion.

The Stryker Spine DIAPASON® Spinal System, OPUS® Spinal System, and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion.

When used for posterior, non-cervical, pedicle screw fixation in pediatric patients, the Xia® 3 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Xia® 3 Spinal System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Except for the staples, when used for posterior non-cervical pedicle screw fixation in pediatric patients, the XIA® 4.5 Spinal System implants are indicated as an adjunct to treat progressive spinal deformities (i.e. scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenttal scoliosis. Additionally, the XIA® 4.5 Spinal System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The MANTIS® Spinal System, MANTIS® Redux Spinal System, and ES2® Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

{3}------------------------------------------------

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| 510(k) Summary: Xia 3 Power Adaptor                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                                         | Stryker Spine                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Contact Person :                                   | Name: Megan Callanan<br>Phone: (551)262-2429<br>Email: megan.callanan1@stryker.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Date Prepared:                                     | 3/11/2020                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Trade Name:                                        | Xia 3 Power Adaptor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Common Name:                                       | Spinal Fixation Appliances, Instrument Accessory                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Proposed Class:                                    | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification Name:                               | Thoracolumbosacral Pedicle Screw System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Regulation Number:                                 | 888.3070                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Product Code:                                      | NKB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Predicate Device:                                  | Primary Predicate is the most recently cleared system K172724.<br>Additional Predicate: K170496, K152632, K120434, K122845, K111478                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device Description:                                | This special 510(k) premarket notification is to expand the intended use of the Xia 3 Power Adaptor.<br><br>The Xia 3 Power Adaptor instrument accessory is intended to facilitate the insertion of pedicle screws using powered instrumentation. To facilitate the insertion of pedicle screws using the power technique, the use of the Xia 3 Power Adaptor is intended for use with Stryker power system drivers (corded and cordless). The adaptor serves as a mechanical interface between the power driver and screwdriver instrument. When the adaptor is attached, the driver (corded and cordless) provides appropriate power to rotate screw drivers for the insertion of pedicle screws. No changes have been made to the indications for use of the associated thoracolumbar spinal implant systems XIA® Spinal Systems (XIA® Stainless Steel, XIA® II, XIA® Anterior, and XIA® Precision), XIA® 3 Spinal System, XIA® 4.5 Spinal System, Radius® Spinal System, MANTIS® Spinal System, MANTIS® Redux Spinal System, and the ES2® Spinal System. The indications for use of each spinal system remain consistent with their most recent 510(k) clearance. |
| Intended Use                                       | The intended use for the Xia 3 Power Adaptor is to facilitate the placement of pedicle screws using the power technique (corded and cordless), the use of the Xia 3 Power Adaptor is intended for exclusive use with the Stryker Cordless and Corded Power Drivers. When the power adaptors are attached, the Stryker Power Drivers (corded and cordless) provide power to rotate screwdrivers for the insertion of pedicle screws.<br><br>Pedicle screws from select Stryker Spine implant systems may be implanted in the non-cervical spine using powered (corded and cordless) instrumentation. The                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Indications for Use:                               | systems included are the family of XIA® Spinal Systems (XIA® Stainless Steel, XIA® II, XIA® Anterior, and XIA® Precision), XIA® 3 Spinal System, XIA® 4.5 Spinal System, Radius® Spinal System, MANTIS® Spinal System, MANTIS® Redux Spinal System, and the ES2® Spinal System.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                    | The XIA® Spinal System is intended for anterior/anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvature (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                    | The XIA® 3, RADIUS® Spinal Systems are intended for use in the non-cervical spine. When used as an anterior/anteriolateral and posterior, non-cervical pedicle and non-pedicle fixation system, the XIA® II, XIA® 3, and RADIUS® Spinal Systems are intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis,kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.                                                                                                                                                                                                                                                                                                                                                                  |
|                                                    | The XIA® 4.5 Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); spondylolisthesis; trauma (i.e. fracture of dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; failed previous fusion.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                    | The Stryker Spine DIAPASON® Spinal System, OPUS® Spinal System, and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                    | When used for posterior, non-cervical, pedicle screw fixation in pediatric patients, the Xia® 3 Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Xia® 3 Spinal System is intended to treat pediatric patients diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                    | pseudarthrosis, and/or failed previous fusion. This system is intended to be used with<br>autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior<br>approach.<br>Except for the staples, when used for posterior non-cervical pedicle screw fixation in<br>pediatric patients, the XIA® 4.5 Spinal System implants are indicated as an adjunct to<br>fusion to treat progressive spinal deformities (i.e. scoliosis, kyphosis, or lordosis)<br>including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis.<br>Additionally, the XIA® 4.5 Spinal System is intended to treat pediatric patients<br>diagnosed with: spondylolisthesis/spondylolysis, fracture caused by tumor and/or<br>trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be<br>used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a<br>posterior approach.<br>The MANTIS® Spinal System, MANTIS® Redux Spinal System, and ES2® Spinal<br>System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle                                                 |
|                                                    | fixation of the spine to provide immobilization and stabilization of spinal segments in<br>skeletally mature patients as an adjunct to fusion for the following indications:<br>degenerative disc disease (defined as back pain of discogenic origin with<br>degeneration of the disc confirmed by history and radiographic studies);<br>spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e.<br>scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous<br>fusion.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Summary of the<br>Technological<br>Characteristics | The Xia 3 Power Adaptor shares the same materials, fundamental scientific<br>technologies and design as the predicate device.<br>The screws are also intended to be inserted manually or with Stryker's corded and<br>cordless power drivers. The power adaptor accessory instrument assembly aids in<br>the rotation of the bone screw to facilitate insertion.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Non-Clinical<br>Performance<br>Evaluation          | The Xia 3 Power Adaptor Instrument has demonstrated substantial equivalence to the<br>predicate device. Engineering analysis demonstrated connection of the Xia 3 Power<br>Adaptor to various Stryker Power Drivers does not represent a new worse case or<br>affect the performance of the Xia 3 Power Adaptor.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Conclusion                                         | Based on the design features, the use of established well known materials, feature<br>comparisons, and indications for use the subject devices have demonstrated<br>substantial equivalence to the identified predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{5}------------------------------------------------

{6}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K200666](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K200666)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
