← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K193219

# G Surgical Marksman System (K193219)

_G Surgical, LLC · NKB · Dec 17, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K193219

## Device Facts

- **Applicant:** G Surgical, LLC
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Dec 17, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The G Surgical Marksman System is intended to provide immobilization of the posterior non-cervical spine (T1-S2/Ilium) in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

## Device Story

The G Surgical Marksman System is a posterior pedicle screw system used for spinal stabilization. It consists of longitudinal rods, pedicle screws, crosslinks, and fasteners. The system is used by surgeons in an operating room setting to provide immobilization of the non-cervical spine (T1-S2/Ilium) as an adjunct to fusion. By securing the spine with these components, the device stabilizes the vertebral column to treat instabilities or deformities, potentially reducing pain and correcting spinal alignment in patients with conditions such as degenerative disc disease, trauma, or scoliosis.

## Clinical Evidence

Bench testing only. Mechanical testing of worst-case constructs was performed according to ASTM F1717, including static and dynamic compression bending and static torsion. Results demonstrate performance is substantially equivalent to predicate devices.

## Technological Characteristics

System components include longitudinal rods, anchors (screws), connectors (crosslinks), and fasteners. Materials: Ti-6AI-4V ELI titanium alloy (ASTM F136) and Cobalt Chrome (ASTM F1537). Design is a rod and screw configuration comparable in size and dimensions to predicate systems. Mechanical testing conducted per ASTM F1717.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- G Surgical Marksman MIS System ([K161516](/device/K161516.md))
- G Surgical Pedicle System ([K081041](/device/K081041.md))
- Tiger® Spine System ([K133369](/device/K133369.md))
- Moss Miami™ Spinal System ([K022623](/device/K022623.md))
- Optima™ Spinal System ([K051971](/device/K051971.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 17, 2019

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

G Surgical LLC % Karen E. Warden, PhD President BackRoads Consulting Inc. PO Box 566 Chesterland, Ohio 44026

Re: K193219

Trade/Device Name: G Surgical Marksman System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: November 20, 2019 Received: November 21, 2019

Dear Dr. Warden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, MBE Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

### K193219

Device Name G Surgical Marksman System

#### Indications for Use (Describe)

The G Surgical Marksman System is intended to provide immobilization of the posterior non-cervical spine (T1-S2/Ilium) in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

Type of Use (Select one or both, as applicable)

| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)    </span> |
|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)    </span>                 |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 7 – 510(k) Summary

| Date:                                                   | 20 November 2019                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor:                                                | G Surgical LLC<br>9433 Bee Cave Road<br>Building 3, Suite 101-A<br>Austin, Texas 78733 USA<br>Tel.: 512.494.4749                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sponsor Contact:                                        | Don Grafton, Managing Director                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| 510(k) Contact:                                         | Karen E. Warden, PhD<br>BackRoads Consulting Inc.<br>PO Box 566<br>Chesterland, OH 44026<br>Office: 440.729.8457                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Proposed Trade Name:                                    | G Surgical Marksman System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Common Name:                                            | Posterior pedicle screw system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Classification:                                  | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Regulation Name,<br>Regulation Number,<br>Product Code: | Thoracolumbosacral pedicle screw system, 888.3070, NKB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Submission Purpose:                                     | The subject 510(k) adds modified pedicle screws, rods, crosslinks and<br>fasteners to the cleared components of the Marksman System.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Description:                                     | The G Surgical Marksman MIS System consists of longitudinal members<br>(rods), anchors (screws), connectors (crosslinks) and fasteners in a variety<br>of sizes to accommodate differing anatomic requirements.                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Indications for Use:                                    | The G Surgical Marksman System is intended to provide immobilization and<br>stabilization of the posterior non-cervical spine (T1-S2/Ilium) in skeletally<br>mature patients as an adjunct to fusion in the treatment of the following<br>acute and chronic instabilities or deformities: degenerative disc disease<br>(defined as back pain of discogenic origin with degeneration of the disc<br>confirmed by history and radiographic studies), spondylolisthesis, trauma<br>(i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis,<br>kyphosis and/or lordosis), tumor, pseudarthrosis, and/or failed previous<br>fusion. |
| Materials:                                              | G Surgical Marksman System is manufactured from Ti-6AI-4V ELI titanium<br>alloy (ASTM F136) and Cobalt Chrome (per ASTM F1537).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Primary Predicate:                                      | G Surgical Marksman MIS System (G Surgical LLC - K161516)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Additional Predicates:                                  | G Surgical Pedicle System (G Surgical LLC - K081041), Tiger® Spine<br>System (CoreLink LLC - K133369), Moss Miami™ Spinal System (DePuy<br>AcroMed, Inc. - K022623) and Optima™ Spinal System (U&I Corporation -<br>K051971)                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Performance Data:                                       | Mechanical testing of worst case G Surgical Marksman System constructs<br>included static and dynamic compression bending and static torsion<br>according to ASTM F1717.<br>The mechanical test results demonstrate that G Surgical Marksman System<br>performance is substantially equivalent to the predicate devices.                                                                                                                                                                                                                                                                                                                              |
| Technological<br>Characteristics:                       | The G Surgical Marksman System possesses the same technological characteristics as one or more of the predicate devices. These include:<br>intended use (as described above) basic design (rod and screw configuration), material (titanium alloy) and sizes (dimensions are comparable to those offered by the predicate systems)              The fundamental scientific technology of the G Surgical Marksman System is the same as previously cleared devices.                                                                                                                                                                                    |
| Conclusion:                                             | The G Surgical Marksman System possesses the same intended use and technological characteristics as the predicate devices. Therefore G Surgical Marksman System is substantially equivalent for its intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                     |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K193219](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K193219)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
