4CIS® Chiron Spinal Fixation System

K190471 · Solco Biomedical Co., Ltd. · NKB · Jul 24, 2019 · Orthopedic

Device Facts

Record IDK190471
Device Name4CIS® Chiron Spinal Fixation System
ApplicantSolco Biomedical Co., Ltd.
Product CodeNKB · Orthopedic
Decision DateJul 24, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

The 4CIS® Chiron Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); deformities or curvatures (i.e., scoliosis, and/or lordosis); tumor; stenosis; and failed previous fusion (pseudoarthrosis).

Device Story

Top-loading posterior spinal fixation system; consists of pedicle screws, rods, nuts, and transverse links. Used by surgeons in open or minimally invasive procedures to stabilize spinal segments and promote fusion. Implant components inserted into vertebral bodies via pedicles; rigidly locked into configurations tailored to patient anatomy and pathology. System provides mechanical immobilization of thoracic, lumbar, and sacral spine segments. Benefits include stabilization of unstable or deformed spinal segments to facilitate fusion and alleviate symptoms associated with degenerative disease, trauma, or deformity.

Clinical Evidence

Bench testing only. Mechanical performance evaluated via static and dynamic compression bending, static tension bending, and static torsion testing in accordance with ASTM F1717. No clinical data provided.

Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V ELI) per ASTM F136 and CoCr alloy per ASTM F1537. Components: Pedicle screws, rods, nuts, transverse links. Form factor: Top-loading posterior spinal fixation system. Sterilization: Non-sterile, single-use. Connectivity: None.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion for thoracic, lumbar, and sacral spine conditions including degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), deformities (scoliosis, lordosis), tumor, stenosis, and pseudoarthrosis.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. July 24, 2019 Solco Biomedical Co., Ltd. % Hwi-Geun Yu 154, Seotan-ro, Seotan-myeon Pyeongtaek-si, Gyeonggi-do Republic of Korea 17704 Re: K190471 Trade/Device Name: 4CIS® Chiron Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: July 8, 2019 Received: July 11, 2019 Dear Hwi-Geun Yu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Ronald P. Jean, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K190471 Device Name 4CIS® Chiron Spinal Fixation System #### Indications for Use (Describe) The 4CIS® Chiron Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); deformities or curvatures (i.e., scoliosis, and/or lordosis); tumor; stenosis; and failed previous fusion (pseudoarthrosis). | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|----------------------------------------------------------------------| | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Submitter | Solco Biomedical Co., Ltd.<br>154 Seotan-ro, Seotan-myeon, Pyeongtaek-si, Gyeonggi-do,<br>17704 Republic of Korea<br>Phone. +82-31-664-4101<br>Fax. +82-31-663-8983 | |-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Hwi-geun Yu<br>Solco Biomedical Co., Ltd.<br>154 Seotan-ro, Seotan-myeon, Pyeongtaek-si, Gyeonggi-do,<br>17704 Republic of Korea<br>Phone: +82)31-610-4091<br>Fax: +82)31-663-8983 | | Submission Date | Feb 22, 2019 | | Trade / Proprietary name | 4CIS® Chiron Spinal Fixation System | | Common / Usual Name | Spinal Fixation System | | Classification Name<br>Classification Code<br>Regulatory Class<br>Regulation Number | Thoracolumbosacral pedicle screw system<br>NKB<br>Class II<br>888.3070 | | Predicate Device | EXPEDIUM SPINE SYSTEM, VIPER SYSTEM, VIPER 2 SYSTEM<br>(K111136) [DEPUY SPINE, INC.] - Primary Predicate<br>MOSS MIAMI SPINAL SYSTEM (K030383) [DEPUY AcroMed Inc.]<br>- Reference Predicate<br>SYNERGY™ TI INTEGRAL OPEN SCREW SYSTEM (K012871)<br>[Interpore Cross International, LLC] - Reference Predicate | | Description of Device | The Spinal Fixation System is a top-loading posterior spinal fixation<br>system which consists of pedicle screws, rods, nuts, transverse<br>(cross) link and associated instruments. Rigid fixation is provided by<br>pedicle screws inserted into the vertebral body through pedicle of the<br>lumbar spine via posterior approach. This system will allow surgeons<br>to build a spinal implant construct to stabilize and promote spinal<br>fusion through open surgery or minimally invasive surgery.<br>Implant components can be rigidly locked into a variety of different<br>configurations to suit the individual pathology and anatomical<br>conditions of the mature patient. The implant components are<br>supplied non-sterile single use and are fabricated from titanium alloy<br>(Ti-6A1-4V ELI) that conforms to ASTM F 136 and CoCr Alloy per<br>ASTM F1537. Also, Specialized instruments are available for the<br>application and removal of the Spinal Fixation System. | | Indication for Use | The 4CIS® Chiron Spinal Fixation System is intended to provide<br>immobilization and stabilization of spinal segments in skeletally mature<br>patients as an adjunct to fusion in the treatment of the following acute<br>and chronic instabilities or deformities of the thoracic, lumbar and sacral<br>spine: degenerative disc disease (defined as discogenic back pain with<br>degeneration of the disc confirmed by history and radiographic studies);<br>spondylolisthesis; trauma (i.e., fracture or dislocation); deformities or<br>curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; stenosis;<br>and failed previous fusion (pseudoarthrosis). | | Comparison of<br>Technological<br>Characteristics with the<br>Predicate Devices | The 4CIS® Chiron Spinal Fixation System and all the predicates have<br>the same or similar indications for use statements. The system is<br>composed of the same material as the predicate devices conforming to<br>recognized industry standards for permanent implants and surgical<br>orthopedic instruments. The 4CIS® Chiron Spinal Fixation System and<br>cited predicate devices share similar basic design features and functions<br>as well as their dimensions. Also they are provided non-sterile for single<br>use only. Mechanical testing confirmed the 4CIS® Chiron Spinal<br>Fixation System demonstrated equivalent performance to the cited<br>predicate device under the same test conditions. | | Performance Data | Mechanical testing (static and dynamic compression bending, static<br>tension bending, static torsion) was conducted in accordance with<br>ASTM F1717.<br>Above non-clinical performance data in the form of a comprehensive<br>literature review was provided in support of substantial equivalence of<br>the subject device to the predicate devices. | | Conclusion | The overall technology characteristics and mechanical performance data<br>lead to the conclusion that Spinal Fixation System is substantially<br>equivalent to legally marketed predicate devices for intended use,<br>material composition, principles of operation, and design. | {4}------------------------------------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...