← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K162764

# JAZZ FRAME SYSTEM (K162764)

_Implanet, S.A. · NKB · Jan 19, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K162764

## Device Facts

- **Applicant:** Implanet, S.A.
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Jan 19, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The Jazz Claw System (hooks and rods) and the Jazz Frame System are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The Jazz Claw System (hooks and rods) and the Jazz Frame System are intended for posterior fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, spinal deformities (i.e., scoliosis and/or lordosis), tumor, pseudarthrosis or revision of a failed fusion attempt. The Jazz Claw System (hooks and rods) and the Jazz Frame System are indicated as an adjunct to fusion to treat adolescent idiopathic scolosis. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

## Device Story

Jazz Frame System is a spinal fixation implant set; components include T-bars, straight rods, and connectors. Used in posterior spinal surgery to immobilize and stabilize spinal segments. Surgeon implants pedicle screws; rods are bent to desired shape; T-bars are positioned at top and middle of rods and tightened; cranial end secured with previously cleared JAZZ Connectors. System provides structural support to facilitate spinal fusion. Used in clinical settings by orthopedic or neurosurgeons. Benefits include stabilization of spine in patients with degenerative disease, deformity, or trauma.

## Clinical Evidence

Bench testing only. Static axial, rotational, and bending testing performed per ASTM F1798. Bacterial endotoxin testing performed per ANSI/AAMI:ST72:2011 to ensure levels <20EU/Device.

## Technological Characteristics

Spinal fixation system consisting of T-bars, straight rods (500/600 mm), and connectors. Anchored via pedicle screws. Mechanical stabilization principle. Testing conducted per ASTM F1798 for static axial, rotational, and bending strength. Pyrogenicity testing per ANSI/AAMI:ST72:2011.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Implanet Spine System (ISS) ([K132303](/device/K132303.md))
- JAZZ System ([K151740](/device/K151740.md))
- Renovis S100 Pedicle Screw System ([K101682](/device/K101682.md), [K111940](/device/K111940.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three figures in profile facing right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 19, 2017

Implanet S.A. % Janice Hogan Regulatory Counsel Hogan Lovells US LLP 1835 Market Street 29th Floor Philadelphia, Pennsylvania 19103

Re: K162764

Trade/Device Name: Jazz Frame System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: December 23, 2016 Received: December 23, 2016

Dear Janice Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below

□ Over-The-Counter Use (21 CFR 801

510(k) Number (if known)

#### K162764

Device Name

#### JAZZ Frame System

Indications for Use (Describe)

The Jazz Claw System (hooks and rods) and the Jazz Frame System are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The Jazz Claw System (hooks and rods) and the Jazz Frame System are intended for posterior fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, spinal deformities (i.e., scoliosis and/or lordosis), tumor, pseudarthrosis or revision of a failed fusion attempt.

The Jazz Claw System (hooks and rods) and the Jazz Frame System are indicated as an adjunct to fusion to treat adolescent idiopathic scolosis. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Type of Use (Select one or both, as applicable)

� Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

# Implanet, S.A.'s JAZZ Frame System

# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Implanet S.A. Technopole Bordeaux Montesquieu Allée François Magendie 33650 Martillac France

Phone: +33 557 995 555 Facsimile: +33 557 995 700

Contact Person: Janice M. Hogan, Regulatory Counsel, Hogan Lovells US LLP

Date Prepared: January 18, 2016

#### Name of Device

JAZZ Frame System

# Common or Usual Name

Thoracolumbosacral Pedicle Screw System

#### Classification Name

Thoracolumbosacral Pedicle Screw System

# Primary Product Code

NKB

#### Requlation Number

21 CFR § 888.3070

#### Device Class

Class II

#### Predicate Devices

Implanet S.A.'s Implanet Spine System (ISS) (K132303) (primary) Implanet S.A.'s JAZZ System (K151740) (additional) Renovis S100 Pedicle Screw System (K101682, K111940) (additional)

#### Intended Use / Indications for Use

{4}------------------------------------------------

The Jazz Claw System (hooks and rods) and the Jazz Frame System are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The Jazz Claw System (hooks and rods) and the Jazz Frame System are intended for posterior fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal deformities (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudarthrosis or revision of a failed fusion attempt.

The Jazz Claw System (hooks and rods) and the Jazz Frame System are indicated as an adiunct to fusion to treat adolescent idiopathic scoliosis. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

# Device Description

The JAZZ Frame spinal implants are spinal fixation devices consisting of the following components: T-bars, straight rods, connectors, and surgical instruments. The T-bars are fixed closed transverse connectors. The rods have been previously cleared as part of the ISS system and are used with the JAZZ Frame. The JAZZ Connector and JAZZ Braid have been previously cleared and are also used with the JAZZ Frame. There is no change to the previously cleared JAZZ instruments.

Once the fusion rods have been bent to the desired shape, a T-bar is positioned at the top and middle of the two bars, and then tightened in place using the screwdriver. The caudal parts of the rods are inserted into heads of pedicle screws that have been previously implanted. The cranial end of the construct is then secured with JAZZ Connectors, and the screws of the T-bars undergo final tightening.

#### Technological Characteristics

The purpose of this 510(k) is to add the JAZZ Frame System to the JAZZ Systems. The JAZZ Frame consists of T-bars and 500 or 600 mm straight rods of various diameters. The construct is anchored to the spine using the previously cleared JAZZ Connectors. The only components not previously cleared are the T-bar connectors.

#### Performance Data

Static axial, rotational, and bending testing per ASTM F1798. Bacterial endotoxin testing (BET) as specified in ANSI/AAMI:ST72:2011 is used for pyrogenicity testing to achieve the Endotoxin limit of <20EU/Device.

#### Substantial Equivalence

The JAZZ Frame is substantially equivalent to existing ISS and JAZZ Systems. Interconnection strength testing has shown the T-bars to possess at least equivalent gripping strength to the crosslink components of the JAZZ Frame has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the JAZZ Frame and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the JAZZ Frame is substantially equivalent to its predicates.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K162764](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K162764)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
