← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K161225

# Paonan Armstrong Posterior Spinal Fixation System (K161225)

_Paonan Biotech Co., Ltd. · NKB · Aug 11, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K161225

## Device Facts

- **Applicant:** Paonan Biotech Co., Ltd.
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Aug 11, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Paonan" Armstrong Posterior Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion using autograft or allograft. This system is intended for posterior, noncervical pedicle fixation of the thoracic, lumbar, and sacral/iliac spine (T1 –Sacrum/Ilium) for the following indications: - Degenerative disc diseases (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) - Degenerative Spondylolisthesis with objective evidence of neurologic impairment. - Fractures - Dislocation - Scoliosis - Kyphosis - Spinal tumor - Failed previous fusion (pseudarthrosis)

## Device Story

The Paonan Armstrong Posterior Spinal Fixation System is a set of spinal implants including pedicle screws (cannulated, mono-axial, poly-axial), rods, cross connectors, and associated instruments. Used by surgeons in clinical settings to provide immobilization and stabilization of spinal segments during fusion procedures. The system is implanted posteriorly in the thoracic, lumbar, and sacral/iliac spine. By securing the spine with these components, the device facilitates fusion, potentially reducing pain and correcting spinal deformities or instability caused by trauma, degeneration, or tumors. The system is non-sterile and requires sterilization by the user.

## Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F1717 (static/dynamic axial compression bending, static torsion) and ASTM F1798 (axial/torsional gripping capacity of cross-link and connecting block mechanisms) to demonstrate substantial equivalence.

## Technological Characteristics

Components: cannulated, mono-axial, and poly-axial pedicle screws, rods, blocks, washers, links, connectors, and set screws. Material: Ti-6Al-4V (ASTM F136/ISO 5832-3). Non-sterile. Mechanical testing per ASTM F1717 and ASTM F1798.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- MYKRES Spinal System ([K051704](/device/K051704.md))
- SYNERGY Spinal System VLS Open ([K950099](/device/K950099.md), [K973836](/device/K973836.md), [K993808](/device/K993808.md))
- Ibis™ Pedicle Screw System ([K142146](/device/K142146.md), [K162995](/device/K162995.md))
- Depuy MOSS Miami Spinal System ([K955348](/device/K955348.md), [K983583](/device/K983583.md), [K011182](/device/K011182.md), [K030383](/device/K030383.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with healthcare, featuring a staff with a serpent entwined around it.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Paonan Biotech Co., Ltd. Ms. Vivi Tsai Regulatory Affairs 3F, No. 50, Lane 258, Rueiguang Road Neihu District, Taipei City 114 TAIWAN

August 11, 2017

Re: K161225

Trade/Device Name: "Paonan" Armstrong Posterior Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: May 16, 2017 Received: July 18, 2017

Dear Ms. Tsai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K161225

#### Device Name

"Paonan" Armstrong Posterior Spinal Fixation System

Indications for Use (Describe)

"Paonan" Armstrong Posterior System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion using autograft. This system is intended for posterior, non-cervical pedicle fixation of the thoracic, lumbar, and sacraliliac spine (T1-Sacrum/Tium) for the following indications:

- Degenerative disc diseases (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) degeneration of the disc confirmed by history and radies)

- Degenerative Spondylolisthesis with objective evidence of neurologic impairment.
- Fractures
- Dislocation
- Scoliosis
- Kyphosis
- Spinal tumor
- Failed previous fusion (pseudarthrosis)

| Type of Use (Select one or both, as applicable)                                                                                                                          |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Registration Use (Part 21 CFR 601 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 601 Subpart C)</label> |

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# ADMINISTRATIVE INFORMATION

| Manufacturer Name            | Paonan Biotech Co., Ltd.                                                       |
|------------------------------|--------------------------------------------------------------------------------|
| Manufacturer Address         | 3F, No.50, Lane.258, Rueiguang Rd.,<br>Neihu District, Taipei City 114, Taiwan |
| DEVICE NAME                  | "Paonan" Armstrong Posterior Spinal Fixation System                            |
| Contact Person/Date Prepared | Vivi Tsai / August 3, 2017                                                     |
| Common/Classification Name   | Thoracolumbosacral Pedicle Screw System. 21 CFR 888.3070. Class II             |

Product code: NKB

## Predicate Device

Primary predicate device: MYKRES Spinal System (K051704)

|                                | Subjective Device                                                                                                                                                                         | Primary Predicate<br>Device                                                                                                                                                                                    | Additional predicates                                                                                                                                                                                         | Additional<br>predicates                                                                                                                                                   | Additional<br>predicates                                                                                                                                                                                      |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| System name                    | "Paonan" Armstrong<br>Posterior Spinal<br>Fixation System                                                                                                                                 | MYKRES Spinal<br>System                                                                                                                                                                                        | SYNERGY Spinal<br>System<br>VLS Open                                                                                                                                                                          | Ibis™ Pedicle<br>Screw System                                                                                                                                              | Depuy MOSS<br>Miami Spinal<br>System                                                                                                                                                                          |
| Manufacture<br>r               | Paonan Biotech. Co.                                                                                                                                                                       | Showa Ika Kohgyo<br>Co.                                                                                                                                                                                        | Cross Medical<br>Products, Inc.<br>(Interpore Cross<br>International)                                                                                                                                         | Ortho<br>Development Corp.                                                                                                                                                 | Depuy Acromed                                                                                                                                                                                                 |
| 510(k)<br>Number               | K161225                                                                                                                                                                                   | K051704                                                                                                                                                                                                        | K950099, K973836,<br>K993808                                                                                                                                                                                  | K142146,<br>K162995                                                                                                                                                        | K955348, K983583,<br>K011182, K030383                                                                                                                                                                         |
| Classificatio<br>n             | 888.3070                                                                                                                                                                                  | 888.3070                                                                                                                                                                                                       | 888.3070                                                                                                                                                                                                      | 888.3070                                                                                                                                                                   | 888.3070, 888.3050                                                                                                                                                                                            |
| Product<br>Code                | NKB                                                                                                                                                                                       | MNI                                                                                                                                                                                                            | MNH                                                                                                                                                                                                           | NKB                                                                                                                                                                        | MNH                                                                                                                                                                                                           |
| Subsequent<br>Product<br>Codes | N/A                                                                                                                                                                                       | KWP, MNH                                                                                                                                                                                                       | KWP, KWQ                                                                                                                                                                                                      | MNH, MNI                                                                                                                                                                   | KWP, MNI, KWQ                                                                                                                                                                                                 |
| Intended Use                   | To provide<br>immobilization and<br>stabilization for<br>posterior, non-<br>cervical, pedicle<br>fixation of the<br>thoracic, lumbar, and<br>sacral/iliac spine (T1<br>– Sacrum/Ileum) in | To provide<br>immobilization and<br>stabilization for<br>posterior, non-<br>cervical, pedicle and<br>non-pedicle fixation<br>of the thoracic,<br>lumbar, and sacral<br>spine in skeletally<br>mature patients. | To provide<br>immobilization and<br>stabilization for<br>posterior, anterior,<br>non-cervical, pedicle<br>and non-pedicle<br>fixation of the<br>thoracic, lumbar, and<br>sacral spine in<br>skeletally mature | To provide<br>immobilization and<br>stabilization for<br>posterior, non-<br>cervical<br>pedicle fixation of<br>the thoracic,<br>lumbar, and<br>sacral/iliac spine<br>(T1 - | To provide<br>immobilization and<br>stabilization for<br>posterior, anterior,<br>non-cervical, pedicle<br>and non-pedicle<br>fixation of the<br>thoracic, lumbar, and<br>sacral spine in<br>skeletally mature |

{4}------------------------------------------------

#### K161225 Page 2 of 3

|                       | skeletally mature<br>patients.                                                                                                                                                                      |                                                                                                                                 | patients.                                                                                                                                                                                         | Sacrum/Ileum) in<br>skeletally<br>mature patients                                                                                                                                                                                                 | patients.                                                                                                                                                                                              |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications           | Degenerative disc diseases Degenerative Spondylolisthesis Fractures Dislocation Scoliosis kyphosis Spinal tumor Failed previous fusion<br>(pseudarthrosis)                                          | Degenerative Spondylolisthesis Fractures Dislocation Scoliosis kyphosis Spinal tumor Failed previous fusion<br>(pseudarthrosis) | Indications for<br>posterior non-cervical,<br>pedicle fixation:<br>Degenerative Spondylolisthesis Fracture Dislocation Scoliosis kyphosis Spinal tumor Failed previous fusion<br>(pseudarthrosis) | Degenerative disc disease (DDD) Degenerative Spondylolisthesis Trauma (fracture or dislocation) Spinal tumor Failed previous fusion<br>(pseudarthrosis) Spinal stenosis Spinal deformities or curvatures such as scoliosis, kyphosis, or lordosis | Indications for<br>posterior non-<br>cervical, pedicle<br>fixation:<br>Degenerative Spondylolisthesis Fractures Dislocation Scoliosis kyphosis Spinal tumor Failed previous fusion<br>(pseudarthrosis) |
| Component             | Cannulated screw,<br>mono-axial screw,<br>poly-axial screw,<br>block, washer, rod,<br>link, connector, set<br>screw                                                                                 | rods, mono-axial<br>screw, poly-axial<br>screw, hooks and<br>connectors                                                         | rods, mono-axial<br>screw, poly-axial<br>screw, hooks and<br>connectors, washers                                                                                                                  | Cannulated screw,<br>mono-axial screw,<br>poly-axial screw,<br>set screw, rod,                                                                                                                                                                    | Rods, screws, hook,<br>transverse connector,<br>axial connectors,<br>staple washer and<br>sacral extender                                                                                              |
| Materials             | Ti-6Al-4V                                                                                                                                                                                           | Ti-6Al-4V                                                                                                                       | Ti-6Al-4V, CP Ti,<br>Stainless Steel                                                                                                                                                              | Ti-6Al-4V                                                                                                                                                                                                                                         | Ti-6Al-4V, Stainless<br>Steel                                                                                                                                                                          |
| Mechanical<br>testing | Static and dynamic<br>compression<br>bending Test, Static<br>Torsional Test per<br>ASTM F1717<br>Axial gripping<br>capacity test and<br>Axial torque<br>gripping capacity<br>test per ASTM<br>F1798 | Static compression,<br>static torsion, and<br>dynamic<br>compression per<br>ASTM F1717                                          | Static compression,<br>static torsion, and<br>dynamic compression<br>per ASTM F1717                                                                                                               | Static compression,<br>static torsion, and<br>dynamic<br>compression per<br>ASTM F1717<br>Axial torque<br>gripping capacity<br>test per ASTM<br>F1798                                                                                             | Static compression,<br>static torsion, and<br>dynamic<br>compression per<br>ASTM F1717                                                                                                                 |
| Sterilization         | Non-sterile                                                                                                                                                                                         | Non-sterile                                                                                                                     | Non-sterile                                                                                                                                                                                       | Non-sterile                                                                                                                                                                                                                                       | Non-sterile                                                                                                                                                                                            |

#### Device Description

"Paonan" Armstrong Posterior Spinal Fixation System includes pedicle screws, rods, cross connectors, and associated instruments. Different types or sizes of these implants are available so that adaptations can be utilized to take into account the unique pathology of individual patients. The system components are manufactured from Ti6Al4V based which complies with ASTM F136 /ISO 5832-3.

{5}------------------------------------------------

"Paonan" Armstrong Posterior Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion using autograft or allograft. This system is intended for posterior, noncervical pedicle fixation of the thoracic, lumbar, and sacral/iliac spine (T1 –Sacrum/Ilium) for the following indications:

- Degenerative disc diseases (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
- Degenerative Spondylolisthesis with objective evidence of neurologic impairment.
- Fractures
- Dislocation
- Scoliosis
- Kyphosis
- Spinal tumor
- Failed previous fusion (pseudarthrosis)

## Performance Data:

Mechanical testing including static/dynamic axial compression bending test and static torsion test for spinal cropectomy model were conducted referring to ASTM F1717 to demonstrate substantial equivalence to the predicate system. In addition, axial and torsional gripping capacity of Cross Link and Connecting Block interconnection mechanism were performed according to ASTM F1798. The results demonstrate substantial equivalence of the "Paonan" Armstrong Posterior Fixation System to the predicate device.

#### Conclusion of Substantial Equivalence:

The "Paonan" Armstrong Posterior Fixation System has been demonstrated to be substantially equivalent to predicate system with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence of the subject device to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K161225](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K161225)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
