Palisade Pedicular Fixation System

K153323 · Spineology, Inc. · NKB · Feb 23, 2016 · Orthopedic

Device Facts

Record IDK153323
Device NamePalisade Pedicular Fixation System
ApplicantSpineology, Inc.
Product CodeNKB · Orthopedic
Decision DateFeb 23, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Palisade Pedicular Fixation System is intended for posterior, non-cervical fixation as an adjunct to fusion in skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis; trauma, (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion.

Device Story

Palisade Pedicular Fixation System; spinal implant construct for immobilization and stabilization of thoracic, lumbar, and sacral spine segments. Surgeon-assembled construct using titanium screws and cobalt chrome rods; accommodates patient-specific anatomical and physiological needs. Used in surgical settings by orthopedic or neurosurgeons. Provides rigid fixation to facilitate spinal fusion; treats acute and chronic instabilities or deformities. Benefits patient by stabilizing spinal segments, reducing pain, and correcting deformities.

Clinical Evidence

Bench testing only. Performance testing conducted per ASTM F1717 (static compression bending, static torsion, dynamic compression bending) and ASTM F1798 (flexural grip, axial grip, torsional grip) to verify mechanical integrity. No clinical data provided.

Technological Characteristics

Materials: Titanium screws, cobalt chrome rods. Standards: ASTM F1717, ASTM F1798. Principle: Posterior pedicular spinal fixation/stabilization. Form factor: Modular screw and rod system. Sterilization: Unchanged from predicate.

Indications for Use

Indicated for skeletally mature patients requiring posterior, non-cervical spinal fixation as an adjunct to fusion for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal curvatures (scoliosis/lordosis), tumor, pseudoarthrosis, or failed previous fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, overlapping each other to create a sense of depth and unity. The faces are simple and abstract, with no distinct features other than the outline of the head and neck. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 23, 2016 Spineology, Incorporated Ms. Jacqueline A. Hauge Regulatory Affairs Manager 7800 3rd Street North, Suite 600 Saint Paul, Minnesota 55128 Re: K153323 Trade/Device Name: Palisade Pedicular Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH Dated: December 3, 2015 Received: December 4, 2015 Dear Ms. Hauge: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K153323 Device Name Palisade Pedicular Fixation System #### Indications for Use (Describe) The Palisade Pedicular Fixation System is intended for posterior, non-cervical fixation as an adjunct to fusion in skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis; trauma, (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion. Type of Use (Select one or both, as applicable) | <div><span style="font-size:16px">☑</span> Prescription Use (Part 21 CER 801 Subpart D)</div> | |-----------------------------------------------------------------------------------------------| | <div>□ Over-The-Counter Use (21 CER 801 Subpart C)</div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "Spineology" in a bold, sans-serif font. To the right of the word is a graphic of a cluster of purple dots arranged in a circular pattern. The word "Spineology" is the main focus of the image, and the purple dots serve as a visual element that complements the text. | Date Prepared: | November 17, 2015 | |------------------------------------|-------------------------------------------------------------------------------| | Submitter: | Spineology Inc.<br>7800 3rd Street North<br>Suite 600<br>Saint Paul, MN 55128 | | Establishment Registration Number: | 2135156 | | Contact Person: | Jacqueline A. Hauge<br>Regulatory Affairs Manger | | Phone: | 651.256.8534 | | Fax: | 651.256.8505 | | Email: | jhauge@spineology.com | ## Device Name and Classification | Trade Name: | Palisade Pedicular Fixation System | |----------------------|------------------------------------------------| | Common Name: | Pedicle Screw System | | Classification Name: | Orthosis, Spinal Pedicle Fixation | | Product Codes: | NKB, MNI, MNH | | Regulatory Class: | MNI, MNH: Class II NKB: Class III | | Regulation Number: | MNI, MNH: 888.3070(b)(1) NKB: 888.3070(b)(2) | | Panel: | Orthopedic | Predicate Devices | Primary: | K143403 | Threshold Pedicular Fixation System | |-------------|--------------------|---------------------------------------------------------------------------------------------------------| | Additional: | K140010<br>K122901 | Fortress Pedicle Screw System<br>Firebird Spinal Fixation System, Phoenix<br>MIS Spinal Fixation System | ### Device Description The Palisade Pedicular Fixation System consists of titanium screws and cobalt chrome rods which allow a surgeon to build a spinal implant construct to accommodate the anatomical and physiological needs of the patient. The system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the word "Spineology" in a bold, sans-serif font. The word is black and is followed by a registered trademark symbol. To the right of the word is a graphic of a cluster of purple dots arranged in a circular pattern. ## Indications for Use The Palisade Pedicular Fixation System is intended for posterior, non-cervical fixation as an adjunct to fusion in skeletally mature patients for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis; trauma, (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion. ## Comparison to Predicate When compared to the predicate devices, the Palisade Pedicular Fixation System has the same: - Intended Use - Indications for Use - Fundamental Scientific Technology ● - · Biological Safety - Packaging System - Sterilization Method ## Summary of Performance and Biological Testing The following performance tests were conducted in accordance with applicable standards and test methods to determine the endurance, strength, and durability of the Palisade Pedicular Fixation System: ## ASTM F1798 - ASTMF1717 - · Static Compression Bending - Flexural Grip • Axial Grip - · Static Torsion - Torsional Grip - - Dynamic Compression Bending Packaging and sterilization testing was not required as there was no change to the sterile barrier system, sterilization method, sterilization parameters, or sterility assurance level (SAL) from the predicate device. Biological safety testing was not required as there was no change in materials from the predicate device. ## Conclusion Spineology has demonstrated that the Palisade Pedicular Fixation System is substantially equivalent to the predicate devices. The fundamental scientific principle, primary technological characteristics, and intended use are unchanged from the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...