← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K151747

# STERISPINE PS (K151747)

_Safe Orthopaedics · NKB · Oct 5, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K151747

## Device Facts

- **Applicant:** Safe Orthopaedics
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Oct 5, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The SteriSpine™PS system is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion. SteriSpine™PS System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; tumor; pseudoarthrosis; and failed previous fusion.

## Device Story

SteriSpine™PS is a pedicle screw spinal system consisting of multiaxial/cannulated pedicle screws and rods. Supplied sterile with single-use surgical instruments. Used by surgeons in clinical settings for spinal immobilization and stabilization as an adjunct to fusion. System provides mechanical support to spinal segments; aids in correcting alignment and stabilizing vertebrae during fusion procedures. Benefits include reduced risk of cross-contamination due to single-use sterile instrumentation. This submission covers modifications to set screw guide and universal handle, plus packaging updates.

## Clinical Evidence

Bench testing only. Biocompatibility testing per ISO 10993-5, -10, -11; packaging validation per ISO 11607-1, -2; sterilization revalidation per ISO 11137-1, -2, -3.

## Technological Characteristics

Implants: Titanium Ta6V Eli (ASTM F136). System: Multiaxial/cannulated pedicle screws, straight/prebent rods. Supplied sterile with single-use surgical instruments. Modifications: Set screw guide and universal handle. Standards: ISO 10993 (biocompatibility), ISO 11607 (packaging), ISO 11137 (sterilization).

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- SteriSpine™PS ([K130632](/device/K130632.md))
- SteriSpine™PS ([K121299](/device/K121299.md))
- SteriSpine™PS ([K112453](/device/K112453.md))
- SteriSpine™PS ([K140802](/device/K140802.md))
- SteriSpine™PS ([K150092](/device/K150092.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 5, 2015

Safe Orthopaedics % Mr. Pierre Dumouchel Industrialization Director / OARA Director Parc des Bellevues - Allée R. Luxembourg - Le Californie 95610 Eragny sur Oise FRANCE

Re: K151747

Trade/Device Name: SteriSpine™ PS Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI, KWP Dated: September 1, 2015 Received: September 4, 2015

Dear Mr. Dumouchel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

### Page 2 - Pierre Dumouchel

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K151747

Device Name SteriSpineTMPS

#### Indications for Use (Describe)

The SteriSpine™PS system is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion. SteriSpine™PS System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; tumor; pseudoarthrosis; and failed previous fusion.

Type of Use (Select one or both, as applicable)

| <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)             </span> |
|-----------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)             </span>  |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Abbreviated 510k STERISPINE™ PS

Image /page/3/Picture/1 description: The image shows the logo for Safe Orthopaedics. The logo features a stylized, cursive letter "S" in a light green color. Below the "S" are the words "Safe Orthopaedics" with "Safe" in a lighter green and "Orthopaedics" in a darker color.

## 510(k) SUMMARY

| 510k                               | Abbreviated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by                       | Safe Orthopaedics<br>Parc des Bellevues<br>Allée R. Luxembourg - Le Californie<br>95610 Eragny sur Oise – FRANCE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Contacts                           | Pierre DUMOUCHEL Industrialization Director / QARA Director<br>p.dumouchel@safeorthopaedics.com<br>+33 (0) 1 34 21 50 00<br>Isabelle DRUBAIX Regulatory contact<br>idee-consulting@nordnet.fr<br>+33 (0)3 21 05 64 23                                                                                                                                                                                                                                                                                                                                                                                       |
| Date Prepared                      | June 8th 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Common Name                        | Pedicle screw spinal system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Trade Name                         | SteriSpineTMPS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification Name                | Pedicle screw spinal system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Class                              | III                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Product Code                       | NKB, MNI, MNH, KWP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| CFR section                        | 888.3070                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device panel                       | Orthopedic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Legally marketed predicate devices | SteriSpineTMPS manufactured by Safe Orthopaedics and cleared under<br>K130632, K121299, K112453, K140802 and K150092                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Indications for use                | The SteriSpineTMPS system is intended to provide immobilization and<br>stabilization of spinal segments in skeletally mature patients as an<br>adjunct to fusion. SteriSpineTMPS System is intended for posterior, non-<br>cervical pedicle and non-pedicle fixation for the following indications:<br>degenerative disc disease (DDD) (defined as back pain of discogenic<br>origin with degeneration of the disc confirmed by history and<br>radiographic studies); spondylolisthesis; trauma (i.e., fracture or<br>dislocation); spinal stenosis; tumor; pseudoarthrosis; and failed previous<br>fusion. |
| Description of the device          | SteriSpineTMPS system includes pedicle screws (multiaxial and<br>cannulated) and rods (straight and prebent). Implantable components of<br>SteriSpineTMPS system are made of Titanium Ta6V Eli grade conforming<br>to ASTM F136. The SteriSpineTMPS range of products is supplied sterile<br>with a sterile single-use set of surgical instruments.                                                                                                                                                                                                                                                         |
| Technological<br>Characteristics   | There have been no changes made to the SteriSpine™PS implants<br>(cleared under K130632, K121299, K112453, K140802 and K150092).<br>The purpose of this 510(k) submission is to obtain clearance for modified<br>set screw guide and universal handle as well as the addition of a<br>"window" to the external packaging of SteriSpine™PS devices supplied<br>sterile. This new packaging will be used for all devices manufactured by<br>Safe Orthopaedics.                                                                                                                                                |
| Discussion of Testing              | The following non-clinical tests were conducted: biocompatibility testing<br>according to ISO 10993-5, ISO 10993-10 and ISO 10993-11 for the<br>modified raw material of the Universal handle; packaging validation<br>according to ISO 11607-1 and 11607-2 and sterilization revalidation<br>according to ISO 11137-1, 11137-2 and 11137-3. Results demonstrate<br>the conformance to above mentioned standards.                                                                                                                                                                                           |
| Conclusion                         | Non clinical performance testing demonstrates that the modified surgical<br>instruments are substantially equivalent to predicate devices in terms of<br>intended use, materials, design and function.                                                                                                                                                                                                                                                                                                                                                                                                      |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K151747](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K151747)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
