The SteriSpine™PS system is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion. SteriSpine™PS System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; tumor; pseudoarthrosis; and failed previous fusion.
Device Story
SteriSpine™ PS is a pedicle screw spinal system for posterior, non-cervical spinal fixation. This submission adds a Percutaneous Ancillary Kit to the previously cleared system. The kit includes surgical instruments: trocar needle, dilators, rod measurer, protection sleeve, funnel body, funnel shaft, and funnel impactor. These instruments are supplied as a sterile, single-use set to facilitate the implantation of multiaxial screws and rods. The system is used by surgeons in an operating room environment to provide spinal segment immobilization as an adjunct to fusion. The device components are made of Titanium Ta6V Eli. The system is supplied sterile.
Clinical Evidence
No clinical data provided. No testing was performed for the added components as they are considered substantially equivalent to the predicate system.
Technological Characteristics
System components (screws, rods) made of Titanium Ta6V Eli (ASTM F136). Percutaneous Ancillary Kit includes trocar needle, dilators, rod measurer, protection sleeve, funnel body, funnel shaft, and funnel impactor. Supplied sterile as single-use sets. Mechanical fixation via pedicle/non-pedicle screws and rods.
Indications for Use
Indicated for skeletally mature patients requiring posterior, non-cervical pedicle and non-pedicle fixation for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, tumor, pseudoarthrosis, and failed previous fusion.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 10, 2015
Safe Orthopaedics Mr. Pierre Dumouchel Quality Affairs & Regulatory Affairs Parc des Bellevues - Allée R. Luxembourg - Bat. Californie 95610 Eragny sur Oise FRANCE
Re: K150092
Trade/Device Name: SteriSpine™ PS Pedicle Screw Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH, KWP Dated: January 9, 2015 Received: January 16, 2015
Dear Mr. Dumouchel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### Indications for Use
### 510(k) Number (if known)
## K150092
Device Name SteriSpineTMPS Pedicle Screw
#### Indications for Use (Describe)
The SteriSpine™PS system is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion.
SteriSpine™PS System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; tumor; pseudoarthrosis; and failed previous fusion.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
## Special 510k STERISPINE® PS
Image /page/3/Picture/1 description: The image shows the logo for Safe Orthopaedics. The logo features a stylized letter "S" in a light green color, with a small dot to the right of the "S". Below the symbol, the words "Safe Orthopaedics" are written in a smaller, sans-serif font, also in a light green color.
# 510(k) SUMMARY
| 510k | Special 510k |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Basis for submission | Addition of a Percutaneous Ancillary Kit to the previously cleared<br>Sterispine™ PS range of products |
| Submitted by | Safe Orthopaedics<br>Parc des Bellevues<br>Allée R. Luxembourg - Le Californie<br>95610 Eragny sur Oise - FRANCE<br>Phone: +33 (0) 1 34 21 50 00 |
| Contacts | Pierre DUMOUCHEL – Quality affairs & Regulatory affairs Director<br>p.dumouchel@safeorthopaedics.com<br>Regulatory contact : Isabelle DRUBAIX (Idée Consulting)<br>idee-consulting@nordnet.fr |
| Date Prepared | January 9th 2015 |
| Common Name | Pedicle screw spinal system |
| Trade Name | Sterispine™ PS Pedicle Screw |
| Classification Name | Pedicle screw spinal system |
| Class | III |
| Product Code | NKB, MNI, MNH, KWP |
| CFR section | 888.3070 |
| Device panel | Orthopedic |
| Legally marketed predicate devices | Sterispine™ PS Pedicle Screw K112453 manufactured by Safe<br>Orthopaedics (Primary Predicate) |
| Indications for use | The SteriSpine™ PS system is intended to provide immobilization and<br>stabilization of spinal segments in skeletally mature patients as an<br>adjunct to fusion. SteriSpine™ PS System is intended for posterior, non-<br>cervical pedicle and non-pedicle fixation for the following indications:<br>degenerative disc disease (DDD) (defined as back pain of discogenic<br>origin with degeneration of the disc confirmed by history and<br>radiographic studies); spondylolisthesis; trauma (i.e., fracture or<br>dislocation); spinal stenosis; tumor; pseudoarthrosis; and failed previous<br>fusion. |
{4}------------------------------------------------
| Description of the<br>device & Technological<br>Characteristics | The cleared range of SteriSpineTM PS system include multiaxial screws and<br>cannulated multiaxial screws with or without extended head (ø5.5, 6.5<br>and 7.5mm, lengths from 25 to 60 mm) and straight and prebent rods (ø<br>5.5, 6.5 and 7.5mm, lengths from 30 to 380 mm). Components of<br>SteriSpineTM PS system are made of Titanium Ta6V Eli grade conforming<br>to ASTM F136. The SteriSpineTM PS range of products is supplied sterile<br>with a sterile single-use set of surgical instruments. The Percutaneous<br>Ancillary Kit added within this submission include (trocar needle, dilators,<br>rod measurer, protection sleeve, funnel body, funnel shaft and funnel<br>impactor). These instruments are supplied as a sterile single-use set to<br>be used with the previously cleared Ancillary Kits. |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Discussion of Testing | No testing has been performed for the added components. |
| Conclusion | The extended range of SteriSpineTM PS system is substantially equivalent<br>to its predicate device SteriSpineTM PS system (K112453) in terms of<br>intended use, material, design, mechanical properties and function.<br>Verification Activity and Validation Activity demonstrate that components<br>added to SteriSpineTM PS system are substantially equivalent to predicates. |