Iliad Pedicle Screw System and Zenius Pedicle Screw System
K142835 · Medyssey USA, Inc. · NKB · Jun 12, 2015 · Orthopedic
Device Facts
Record ID
K142835
Device Name
Iliad Pedicle Screw System and Zenius Pedicle Screw System
Applicant
Medyssey USA, Inc.
Product Code
NKB · Orthopedic
Decision Date
Jun 12, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Medyssey Co, Ltd. Iliad Pedicle Screw System and Zenius Pedicle Screw System are intended for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograft for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
Device Story
Iliad and Zenius Pedicle Screw Systems are spinal fixation implants used in orthopedic surgery. Systems consist of rods, pedicle screw assemblies, compression retaining assemblies, and transverse-link assemblies. Implants are surgically implanted by surgeons to provide stabilization during spinal fusion procedures. Components are available in various sizes to accommodate patient-specific spinal pathology. Device provides mechanical support to the spine to facilitate fusion in patients with degenerative or traumatic conditions. No software or algorithmic components are involved.
Clinical Evidence
Bench testing only. Mechanical performance testing conducted per ASTM F1717-13 and ASTM F1798-13, including flexural grip, torsional grip, static compression bending, and dynamic compression bending.
Technological Characteristics
Components constructed from Ti6Al4V ELI (ASTM F136) and wrought Co-Cr-Mo alloy (ASTM F1537). System includes rods, pedicle screws, compression retaining assemblies, and transverse-link assemblies. Mechanical fixation device; no energy source, software, or connectivity.
Indications for Use
Indicated for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograft/allograft for DDD, spondylolisthesis, trauma, spinal stenosis, spinal curvatures (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, and failed previous fusion.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
K050484 — BASIS SPINAL SYSTEM · Medtronic Sofamor Danek, Inc. · May 4, 2005
K101278 — INERTIA PEDICLE SCREW SYSTEM · Nexxt Spine, LLC · Dec 20, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other. The profiles are stylized and appear to be made of flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 12, 2015
Medyssey USA, Incorporated % Rich Jansen, Pharm.D. Silver Pine Consulting, LLC 11821 Bramble Cove Drive Fort Myers, Florida 33905
Re: K142835
Trade/Device Name: Iliad Pedicle Screw System and Zenius Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH, KWP Dated: May 12, 2015 Received: May 14, 2015
Dear Dr. Jansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Rich Jansen, Pharm.D.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K142835
#### Device Name
Iliad Pedicle Screw System and Zenius Pedicle Screw System
#### Indications for Use (Describe)
The Medyssey Co, Ltd. Iliad Pedicle Screw System and Zenius Pedicle Screw System are intended for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograft for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Date Prepared: | June 11, 2015 |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | John Kuczynski, VP R&D and RA<br>Medyssey USA Inc.<br>1550 E. Higgins Road<br>Suite 123<br>Skokie, IL 60077<br>847-427-0200<br>888-518-9070 (fax) |
| Regulatory Contact: | Rich Jansen, Pharm. D.<br>Silver Pine Consulting |
| Product | |
| Trade Names: | Iliad Pedicle Screw System and<br>Zenius Pedicle Screw System |
| Product Class: | Class III |
| Classification: | 21 CFR §888.3070 Pedicle Screw Spinal System; and<br>888.3050 - Spinal Interlaminal fixation orthosis |
| Common Name: | Pedicle Screw System |
| Product Codes: | NKB, MNI, MNH, KWP |
| Panel Code: | 87 |
#### Purpose:
The purpose of this Premarket Notification is to add components and addition to the previously cleared Iliad and Zenius pedicle screw systems to better facilitate use in deformity surgeries.
#### Device Descriptions:
The Iliad TM Spinal Fixation and Adjustable Bridge System, Internal Fixation Device for Spinal Surgery is comprised of: Rods, Pedicle Screw Assemblies, Compression Retaining Assemblies, and Transverse-Link Assemblies. Various forms and sizes of these implants are available so that adaptations can be utilized to take into account the unique pathology of individual patients.
Components are made of Ti6Al4V ELI, a titanium based alloy, which complies with ASTM F136. The Cobalt-Chrome rods are made from wrought Co-Cr-Mo alloy, which complies with ASTM F1537.
The Zenius™ Spinal System, Internal Fixation Device for Spinal Surgery is comprised of: Rods, Pedicle Screw Assemblies, Compression Retaining Assemblies, and Transverse-Link Assemblies. Various forms and sizes of these implants are available so that adaptations can be utilized to take into account the unique pathology of individual patients. Components are made of Ti6Al4V ELI, a titanium based alloy, which complies with ASTM F136. The Cobalt-Chrome rods are made from wrought Co-Cr-Mo alloy, which complies with ASTM F1537.
{4}------------------------------------------------
#### Indications for Use:
The Medyssey Co. Ltd. Iliad Pedicle Screw System and Zenius Pedicle Screw System are intended for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograft and/or allograft for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation): spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
#### Predicate Device(s):
Medyssey believes these devices are substantially equivalent to predicate devices previously cleared by FDA. The CD Horizon system by Medtronic (K042025) is the primary predicate.
Additional predicate devices include the lliad and Zenius systems (K131878), the Expedium Spine System by DePuy Spine (K102249), the Optima system by U&I (Zimmer) (K051971), and the Xia system by Stryker (K001272).
#### Technological Characteristics:
The Iliad and Zenius Pedicle Screw Systems share the same indications for use, design, material, performance, and fundamental technologies as the predicate devices.
#### Performance Standards:
The pre-clinical testing was performed by an independent laboratory. Testing was conducted per ASTM F1717-13 and ASTM F1798-13. Testing included:
- Flexural Grip .
- . Torsional Grip
- . Static Compression Bending
- . Dynamic Compression Bending
#### Conclusion:
Medyssey concludes that the additions to the previously cleared lliad and Zenius pedicle systems raise no new questions of safety or effectiveness and are substantially equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.