← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K140696

# STREAMLINE TL SPINAL FIXATION SYSTEM (K140696)

_Pioneer Surgical Technology, Inc. · NKB · Apr 10, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K140696

## Device Facts

- **Applicant:** Pioneer Surgical Technology, Inc.
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Apr 10, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Streamline TL Spinal System components are non-cervical spinal fixation devices intended as an adjunct to fusion for use as a pedicle screw (T1-S2), sacral/iliac screw fixation or as an anterolateral fixation system (T8 - L5). Pedicle screw fixation is limited to skeletally mature patients. These devices are indicated for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion.

## Device Story

Streamline TL Spinal System is a pedicle screw system consisting of rods, poly-axial/fixed/reduction screws, transverse connectors, and set screws. Used for posterior or anterolateral spinal fixation; components are rigidly locked into tailor-made constructs for individual patients. Intended for use in OR by orthopedic or neurosurgeons to stabilize spine during fusion procedures. System includes manual instrumentation for implantation. Line extension adds additional screw diameters to existing system. Benefits include spinal stabilization and support for fusion in patients with degenerative conditions, deformities, or trauma.

## Clinical Evidence

No clinical data provided; bench testing only (engineering analysis).

## Technological Characteristics

Implant components manufactured from Titanium Alloy (ASTM F136) and Cobalt Chromium Alloy (ASTM F1537). System includes poly-axial, fixed, and reduction screws, rods, and transverse connectors. Manual instrumentation used for implantation. No software or electronic components.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Pioneer Streamline TL Spinal System ([K131100](/device/K131100.md))
- Blackstone Medical, Inc. Firebird Spinal Fixation System ([K093926](/device/K093926.md))

## Reference Devices

- Streamline TL Crosslink
- SpineWorks FixxSure Crosslink
- Quantum® Spinal System X-Link®

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary Pursuant to 21 CFR 807.92

Pioneer Surgical Technology, Inc. Sponsor: (RTI Surgical, Inc.) 375 River Park Circle Marquette, MI. 49855 USA Contact: Sarah McIntyre Ph: (906) 225-5861 Fx: (906) 226-4459 Prepared: March 18, 2014 Name: Streamline TL Spinal System Trade name: Streamline TL Spinal Fixation System Common name: Pedicle screw system 21 CFR 888.3060, Spondylolisthesis Spinal Fixation Device System and 21 Classifications: CFR 888.3070 Pedicle Screw Spinal System, Class III Product Codes: NKB, KWQ, MNI, MNH Panel/ Branch: Orthopaedic and Rehabilitation Devices Panel; Panel Code 87 Predicates: Pioneer Streamline TL Spinal System (K131100) Blackstone Medical, Inc. Firebird Spinal Fixation System (K093926) The Streamline TL Spinal System consists of a variety of rods, screws Description: (poly-axial, fixed, and reduction), transverse connectors, set screws and other connecting components used to build a spinal construct. The implant components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. Sacral/iliac screws are designed for posterior fixation. The Streamline TL Spinal System includes Class I manual instrumentation to facilitate implantation of the device components. The Streamline TL Spinal System may be used with the Streamline TL Crosslink, SpineWorks FixxSure Crosslink or the Quantum® Spinal System X-Link® and rods. Description of This 510(k) is intended to introduce a line extension to the existing Streamline TL Spinal System. The line extension consists of additional Device Modification: diameters for the poly-axial screws.

APR 1 0 2014

{1}------------------------------------------------

The Streamline TL Spinal System components are non-cervical spinal Intended Use: fixation devices intended as an adjunct to fusion for use as a pedicle screw (T1-S2), sacral/iliac screw fixation or as an anterolateral fixation system (T8 - L5). Pedicle screw fixation is limited to skeletally mature patients. These devices are indicated for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion.

The implant components of the Streamline TL Spinal System are Materials: manufactured from implant grade Titanium Alloy per ASTM F136. Spinal rods are also available in Cobalt Chromium Alloy per ASTM F1537.

Substantial The Streamline TL Spinal System, with incorporation of the subject Equivalence components, is substantially equivalent to the predicate in terms of material, design, and indications for use. Engineering analysis was completed for the subject components and demonstrated no pre-clinical performance data was required to demonstrate equivalence of the product. There are no significant differences between the Streamline TL Spinal System and the predicate devices which would adversely affect the use of the product.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 10, 2014

Pioneer Surgical Technology, Incorporated Ms. Sarah McIntyre Regulatory Affairs Associate II 375 River Park Circle Marquette, Michigan 49855

Re: K140696

Trade/Device Name: Streamline TL Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH, KWO Dated: March 18, 2014 Received: March 19, 2014

Dear Ms. McIntyre:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Sarah Meintyre

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# RonaldiPhylean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use

510(k) Number (if known)

#### K140696

Device Name

Streamline TL Spinal System

# Indications for Use (Describe)

The Streamline 71. Spinal System components are non-cervical spinal fixation devices intended as an adjunct to fusion for use as a pedicle screw (TI-S2). sacral/iliac screw fixation or as an anterolateral fixation system (T8 - L5). Pedicle screw fixation is limited to skeletally mature patients. These devices are indicated for all of the following indications: degenerative disc discase (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), turnor, stenosis, pseudoarthrosis, and failed previous fusion.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# 2014.04.10

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden lime for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K140696](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K140696)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
