← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K103748

# LAGUNA PEDICLE SCREW SYSTEM (K103748)

_Phygen, LLC · NKB · Mar 2, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K103748

## Device Facts

- **Applicant:** Phygen, LLC
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Mar 2, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The LAGUNA® Pedicle Screw System is intended to be used as an adjunct to fusion in skeletally mature patients using autograft or allograft in posterior, non-cervical fixation for the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration disc confirmed by history and radiographic studies); the of spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumors; pseudarthrosis; and/or failed previous fusion.

## Device Story

LAGUNA® Pedicle Screw System is a spinal fixation implant set consisting of polyaxial screws and connecting rods. Used by surgeons in clinical settings to provide immobilization and stabilization of thoracic, lumbar, and sacral spinal segments as an adjunct to fusion. Components are assembled into custom configurations for individual patient anatomy. System provides mechanical support to the spine to facilitate fusion; benefits include correction and stabilization of spinal segments in patients with degenerative or traumatic conditions.

## Clinical Evidence

Bench testing only. Biomechanical performance evaluated per ASTM F1717-10, including static compression bending, static torsion, and dynamic compression bending. Results confirmed equivalence to predicate devices.

## Technological Characteristics

Materials: Medical grade titanium alloy per ASTM F136. Components: Polyaxial screws (winged/non-winged, 30-100mm length, 5-8mm diameter) and 5.5mm connecting rods. Mechanical fixation via rigid locking configurations. Non-sterile (implied).

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- LAGUNA® Pedicle Screw System ([K083826](/device/K083826.md))
- LAGUNA® Pedicle Screw System ([K091995](/device/K091995.md))
- LAGUNA® Pedicle Screw System ([K050060](/device/K050060.md))
- LAGUNA® Pedicle Screw System ([K072678](/device/K072678.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K103748

MAR - 2 2011

### 510(k) Summary

[in accordance with SMDA of 1990, 21 CFR 807.92(c)]

Contact: Mr. Hartmut Loch Vice President, Regulatory Affairs & Quality Assurance Phygen. LLC. 2301 Dupont Drive. Suite 510 Irvine CA 92612

February 23, 2011 Date Prepared:

LAGUNA® Pedicle Screw System Trade name:

Common name: Spinal Fixation System

Classification Appliance, Fixation, Spinal Interlaminal -§ 888.3050 (KWP) - class II name: Orthosis, Spinal Pedicle Fixation - - -§ 888.3070 (MNI) - class II Orthosis, Spondylolisthesis Spinal Fixation § 888.3070 (MNH) - class II Orthosis, Spinal Pedicle Fixation, For Degenerative Disc Disease - § 888.3070 (NKB) - class III

- All Orthopedic Device Panel
#### Product Code (s): KWP, MNI, MNH & NKB

Device Description and Characteristics:

The LAGUNA® Pedicle Screw System is intended to help provide correction, immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar and/or sacral space. The LAGUNA® Pedicle Screw System consists of a variety of rods and screws, which can be rigidly locked into a variety of configurations, with each construct being tailor made for the individual case. Polyaxial screws are supplied in winged and non-winged configurations, in a variety of different length, ranging from 30 mm to 100 mm and 5 mm, 6 mm, 7 mm and 8mm diameter sizes. All sizes are able to receive 5.5mm connecting rods only.

The LAGUNA® Pedicle Screw System implant components are fabricated from medical grade titanium alloy per ASTM F136.

The modified LAGUNA® Pedicle Screw System is substantially equivalent to Equivalence: the original LAGUNA® Pedicle Screw System (K083826, K091995, K050060, K072678), which are manufactured and marketed by Phygen,LLC.

- The LAGUNA® Pedicle Screw System is intended to be used as an adjunct Indications: to fusion in skeletally mature patients using autograft in posterior, non-cervical fixation for the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumors; pseudarthrosis; and/or failed previous fusion.
### Biomechanical tests per ASTM F1717-10 (Static Compression Bending, Performance data: Static Torsion, and Dynamic Compression Bending) have been performed. The test results were equivalent to other similar implants and are sufficient for in vivo loading.

page 1 of 1

{1}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, depicted in black, with its wings forming a shape that resembles human profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Phygen, LLC % Mr. Hartmut Loch Vice President, Regulatory Affairs and Quality Assurance 2301 Dupont Drive, Suite 510 · Irvine, California 92612

MAR - 2 2011

Re: K103748

Trade/Device Name: LAGUNA® Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH, KWP, KWQ Dated: December 22, 2010 Received: December 23, 2010

Dear Mr. Loch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 – Mr. Hartmut Loch

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Aling B. Rth
for

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known): K103748

Device Name(s): LAGUNA® Pedicle Screw System

Indications for Use:

The LAGUNA® Pedicle Screw System is intended to be used as an adjunct to fusion in skeletally mature patients using autograft or allograft in posterior, non-cervical fixation for the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration disc confirmed by history and radiographic studies); the of spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumors; pseudarthrosis; and/or failed previous fusion.

Prescription Use

Over-The-Counter-Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

KI03748 510(k) Number_

Page _________________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K103748](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K103748)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).