MANTIS SPINAL SYSTEM & MANTIS REDUX SPINAL SYSTEM

K102235 · Stryker Spine · NKB · Aug 24, 2010 · Orthopedic

Device Facts

Record IDK102235
Device NameMANTIS SPINAL SYSTEM & MANTIS REDUX SPINAL SYSTEM
ApplicantStryker Spine
Product CodeNKB · Orthopedic
Decision DateAug 24, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

The MANTIS® Spinal System and MANTIS® Redux Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: - Degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); - Spondylolisthesis; - Trauma (i.e., fracture or dislocation); - Spinal stenosis; - Curvatures (i.e., scoliosis, kyphosis, and/or lordosis); - Tumor; - Pseudoarthrosis; and - Failed previous fusion. The Titanium and Vitallium® rods from the Stryker Spine RADIUS® Spinal System are intended to be used with the other components of MANTIS® Spinal System and MANTIS® Redux Spinal System.

Device Story

MANTIS Spinal System and MANTIS Redux Spinal System are spinal fixation appliances; line extension adds 7.5 mm and 8.5 mm diameter long arm polyaxial screws (30-90 mm length). System provides immobilization and stabilization of spinal segments; used as adjunct to fusion. Components include pedicle screws and rods (including Titanium and Vitallium rods from RADIUS system). Used in surgical settings by surgeons for posterior, non-cervical spinal fixation. Device stabilizes spine to facilitate fusion; benefits patient by correcting deformities, treating trauma, or addressing degenerative conditions.

Clinical Evidence

Bench testing only. Static compression bending, static torsion, and fatigue compression bending tests were performed in accordance with ASTM F1717 to verify the mechanical integrity of the additional screw sizes compared to the predicate system.

Technological Characteristics

Spinal fixation system consisting of polyaxial screws and rods. Materials include Titanium and Vitallium. Mechanical testing performed per ASTM F1717. System is a non-active, implantable orthopedic device.

Indications for Use

Indicated for skeletally mature patients requiring percutaneous, posterior, non-cervical pedicle or non-pedicle spinal fixation as an adjunct to fusion. Conditions include degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal curvatures (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K10 2235 Page 1 of 2 # AUG 2 4 2010 ## STRYKER SPINE Special 510(k) Summary of Safety and Effectiveness: Line Extension | Proprietary Name: | MANTIS® Spinal System and MANTIS® Redux Spinal System | |---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Spinal Fixation Appliances | | Classification Name and<br>Reference: | 1) Spinal Intervertebral Body Fixation Orthosis, 21 CFR §888.3060<br>2) Pedicle Screw Spinal System, 21 CFR §888.3070 (b) (1) & (b) (2) | | Device Product Code: | NKB, MNH, MNI, KWQ | | Proposed Regulatory Class: | Class III | | For Information contact: | Curtis Truesdale<br>Regulatory Affairs Project Manager<br>2 Pearl Court<br>Allendale, NJ 07401<br>Telephone: (201) 760-8296<br>Fax: (201) 760-8496<br>Email: Curtis.Truesdale@Stryker.com | | Date Summary Prepared: | August 6, 2010 | | Predicate Devices | Stryker Spine MANTIS® Spinal System and MANTIS® Redux Spinal System, K092631; Stryker Spine Xia® II Spinal System, K063428; Stryker Spine Xia Spinal System, K002858, K013823, K043473; Stryker Spine Osteonics Spinal System, K951725 | . . . . . {1}------------------------------------------------ # K102235 #### Description of Device Modification This 510(k) is intended to introduce a line extension to the existing MANTIS® Redux Spinal System. The line extension consists of additional sizes of the MANTIS® Redux Long Arm Polyaxial Screws. The additional sizes include the 7.5 mm and 8.5 screw diameter, ranging 30 - 90 mm in length for the MANTIS Redux Spinal System. No modifications were made to the MANTIS Spinal System. #### Intended Use The MANTIS® Spinal System and MANTIS® Redux Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: - u Degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); - Spondylolisthesis; - 트 Trauma (i.e., fracture or dislocation); - Spinal stenosis; - Curvatures (i.e., scoliosis, kyphosis, and/or lordosis); ■ - Tumor: . - . Pseudoarthrosis; and - . Failed previous fusion. The Titanium and Vitallium® rods from the Stryker Spine RADIUS® Spinal System are intended to be used with the other components of MANTIS® Spinal System and MANTIS® Redux Spinal System. Summary of the Technological Characteristics The Stryker Spine MANTIS® Redux Spinal System, with the incorporation of the subject components, is substantially equivalent to the predicate devices in terms of material, design, and indications for use. Static Compression Bending testing, Static Torsion testing and Fatigue Compression Bending testing per ASTM F1717 were conducted on the subject components. The results obtained from these tests were compared to those of a predicate system to demonstrate substantial equivalence, as recommended by the "Guidance for Industry & FDA Staff Spinal System 510(k)s, May 3, 2004." {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, with three distinct lines forming the body and wings. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Stryker Spine % Mr. Curtis Truesdale Regulatory Affairs Project Manager 2 Pearl Court Allendale, New Jersey 07401 ### AUG 2 4 2010 Re: K102235 Trade/Device Name: MANTIS® Spinal System and MANTIS® Redux Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI, KWQ Dated: August 06, 2010 Received: August 09, 2010 Dear Mr. Truesdale: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ Page 2 - Mr. Curtis Truesdale or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Barbara Buettner Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 1112235 # Indications for Use 510(k) Number (if known): K102235 Device Name: MANTIS® Spinal System and MANTIS® Redux Spinal System Indications for Use: The MANTIS® Spinal System and MANTIS® Redux Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: - degenerative disc disease (defined as back pain of discogenic origin with . degeneration of the disc confirmed by history and radiographic studies); - . spondylolisthesis; - trauma (i.e. fracture or dislocation); . - spinal stenosis; . - curvatures (i.e. scoliosis, kyphosis, and/or lordosis); . - . tumor: - Pseudoarthrosis; and . - . failed previous fusion. The Titanium and Vital!ium rods from the Stryker Spine RADIUS® Spinal System are also intended to be used with other components of the MANTIS® Spinal System and MANTIS® Redux Spinal System. AND/OR Prescription Use (21 CFR 801 Subpart D) Over-The-Counter Use _ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) signature (Division Sign-Off) Division of Surgical. Orthopedic, and Restorative Devices K102235 510(k) Number_
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...