EXPEDIUM SPINE SYSTEM

K101070 · Medos International SARL · NKB · Aug 31, 2010 · Orthopedic

Device Facts

Record IDK101070
Device NameEXPEDIUM SPINE SYSTEM
ApplicantMedos International SARL
Product CodeNKB · Orthopedic
Decision DateAug 31, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

The EXPEDIUM Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The EXPEDIUM Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

Device Story

EXPEDIUM Spine System is a rod-based spinal fixation system; components include monoaxial/polyaxial screws, reduction screws/hooks, fixed/polyaxial bolts, plates, connectors, nuts, washers, and wires. System provides immobilization and stabilization of spinal segments as an adjunct to fusion. Used in thoracic, lumbar, and sacral spine procedures. Components are available in various geometries/sizes to accommodate patient anatomy; provided sterile or non-sterile. Surgeon implants hardware to stabilize spine; aids in treatment of instabilities or deformities. Benefits include structural support for spinal fusion, potentially reducing pain and correcting spinal alignment.

Clinical Evidence

No clinical tests were performed. Bench testing included static compression bending, static torsion, and dynamic compression bending per ASTM F1717-04, all of which met acceptance criteria.

Technological Characteristics

Rod-based spinal fixation system (5.5mm and 6.35mm). Materials: ASTM F138 stainless steel, ASTM F136 titanium alloy, and ASTM F562 cobalt-nickel-chromium-molybdenum alloy. Components include screws, bolts, plates, connectors, nuts, washers, and wires. Non-sterile and sterile options. Mechanical fixation via pedicle or non-pedicle approach.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion for thoracic, lumbar, and sacral spine conditions including degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, curvatures (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | 5. 510(K) SUMMARY | AUG 31 2010 | |-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Medos International Sárl<br>Chemin-Blanc 38<br>Le Locle, CH-NE 2400, Switzerland | | Contact Person: | Daphney Germain<br>Regulatory Affairs Associate<br>DePuy Spine, Inc.<br>Voice: (508) 880-8395<br>Fax: (508) 828-3797<br>E-Mail: dgermain@its.jnj.com | | Date Prepared: | August 16, 2010 | | Device Class: | Class III | | Tradename: | EXPEDIUM® Spine System | | Common Name: | Appliance, Fixation, Spinal Interlaminal;<br>Orthosis, Spondyloisthesis Spinal Fixation;<br>Orthosis, Spinal Pedicle Fixation;<br>Orthosis, Spinal Pedicle Fixation, For Degenerative Disc Disease | | Classification Name: | Spinal interlaminar fixation orthosis<br>per 21 CFR §888.3050<br>Spinal intervertebral body fixation orthosis<br>per 21 CFR §888.3060<br>Pedicle screw spinal fixation<br>per 21 CFR §888.3070 | | Classification Panel: | Orthopedics | | FDA Panel Number: | 87 | | Product Code(s): | NKB, KWQ, KWP, MNH, MNI | | Proprietary Name: | EXPEDIUM® Spine System | | Device Description: | The EXPEDIUM Spine System is a 5.5mm and 6.35mm rod based<br>system offered in both titanium and stainless steel. Both systems<br>consist of the following:<br>Monoaxial Screws<br>Polyaxial Screws | {1}------------------------------------------------ - . Reduction screws - . Reduction hooks - Hooks . - Extended Tab Implants . - Sacral Extenders . - Lateral connectors . - Washers . The proposed addition to the EXPEDIUM Spine System is a 5.5 and 6.35mm rod and plate offset system offered in both titanium and stainless steel. The EXPEDIUM Spine System will also include the following: - Fixed Bolts ● - . Polyaxial Bolts - Closed screws . - Slotted connectors . - Plates . - . Nuts - . Washers - . Drop-entry connectors - . Modular cross connectors - Transverse rod connectors . - . Wires These additional components are available in various geometries and sizes to accommodate patient anatomy. They will be provided non-sterile and sterile. The EXPEDIUM Spine System is intended to provide Intended Use: immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. > The EXPEDIUM Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients. {2}------------------------------------------------ | Materials: | Manufactured from ASTM F138 implant grade stainless steel,<br>ASTM F136 implant grade titanium alloy, and ASTM F 562<br>implant grade cobalt-nickel-chromium-molybdenum alloy. | |---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Predicate Devices: | EXPEDIUM Spine System - K092473, K090230, K080313<br>K062174, and K071495<br>ISOLA System - K013441, K010972, and K980485<br>MONARCH Spine System - K024348 and K010576<br>VSP System - K984350<br>TIMX System – K981274 | | Summary of<br>Technological<br>Differences: | The purpose of this submission is to obtain market clearance for<br>the proposed additional components to the EXPEDIUM 5.5mm<br>and 6.35mm Spine Systems which consist of closed screws, fixed<br>bolts, polyaxial bolts, cross connectors, nuts, washers, wires, and<br>plates. These proposed components have the same design<br>characteristics, performance, intended use, and packaging as the<br>predicate devices. The key differences between the subject and<br>predicate devices are:<br>The addition of various sizes of closed screws, fixed bolts,<br>polyaxial bolts, cross connectors, nuts, washers, wires, and<br>plates. The availability of sterile components. | | Nonclinical Test<br>Summary: | The following mechanical tests were conducted:<br>Static compression bending in accordance with ASTM<br>F1717-04 test standard Standard Test Methods for Spinal<br>Implant Constructs in a Vertebrectomy Model. The<br>acceptance criteria was/were met. Static torsion in accordance with ASTM F1717-04 test<br>standard Standard Test Methods for Spinal Implant<br>Constructs in a Vertebrectomy Model. The acceptance<br>criteria was/were met. Dynamic compression bending in accordance with ASTM<br>F1717-04 test standard Standard Test Methods for Spinal<br>Implant Constructs in a Vertebrectomy Model. The<br>acceptance criteria was/were met. | | Clinical Test<br>Summary: | No clinical tests were performed. | {3}------------------------------------------------ Traditional 510(k) Submission – Additions to the EXPEDIUM® 5.5mm & 6.35mm Spine Systems . ## Conclusion: Based on the predicate comparison and testing, the subject additions to the EXPEDIUM 5.5mm and 6.35mm Spine Systems are substantially equivalent to the predicate devices. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Medos International Sarl % DePuy Spine, Inc. Ms. Daphney Germain Regulatory Affairs Associate 325 Paramount Drive Raynham, Massachusetts 02767 AUG 3 1 2010 Re: K101070 K101070 Trade/Device Name: EXPEDIUM® Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Regulatory Chass!" B, MNH, MNI, KWQ, KWP Dated: August 16, 2010 Received: August 17, 2010 Dear Ms. Germain: We have reviewed your Section 510(k) premarket notification of intent to market the device ting indication in the charge is and the device is whetantially equivalent (for the We have reviewed your Section > IV(K) premium in the minuted (for the indications referenced above and have determined the device is substantially equivalent in interstate referenced above and have determined the devices marketed in interstate for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosure) to legally market predical Device Amendments, or to commerce prior to May 28, 1976, the enactment date of the Federal Food, Drug, commerce prior to May 28, 1976, the enactment use of the Federal Food, Drug, Drug, devices that have been reclassified in accordance with the provisions of the Federal Food, devices that have been reclassined in accordance will and proval application and Cosmetic Act (Act) that do not require approval of a premarket approval application and Cosmetic Act (Act) that do not require approvation to the general controls provisions of the (PMA). You may, therefore, market the device, subject to the general control (PMA). You may, therefore, market the device, suchie menual registration, Act. The general controls provisions of the Act include requirements for annual misbrandin Act. The general controls provisions of the Act Include Lequintini Son against misbranding listing of devices, good manufacturing practice, labeling, and prohibitions agains listing of devices, good manufacturing practice, labelion related to contract liability and adulteration. Please note: CDRH does not evaluate information related ing. and adulteration. Please note: CDRH does not evaluate information on the misleading. warranties. We remind you; however, that device labeling must be truthful and not mis If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) into eimer class if (optents of control your device can be may be subject to additional controls. Existing major regulations affecti may be subject to additional controls. Existing major regardent of the Soles of Anaddition, FDA may found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In found in the Code of Federal Regulations, This ETF and Collection publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ ## Page 2 - Ms. Daphney Germain Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s Issualic of a subscribes with other requirements of the Act that FDA has made a determination that your device only - Econoles, You must that FDA has made a delerimiation and Jour and Journer Federal agencies. You must or any Federal statures and regulations administers or registration and listing (21 comply with all the Act's requirements, including, but not setting of medical comply with all the Act S requirements, includes reporting (to ording of medical CFR Part 807); labeling (21 CFR Pall 801); pood manufacturing practice requirements as set device-related adverse events) (21 CFR 803); good manufacturing practice requiremen device-related adverse events) (21 CFR 803); good fif applicable, the electronic forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the elect forth in the quality systems (Sections (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please r (1580), please (1) (1) = ================================================ If you desire specific advice for your device of var adolfices/ucm115809.htm for any for for for for for for for the go to http://www.ida.gov/Aboult12/0CentersOries/Office of Compliance. Also, please the Center for Devices and Radiological Health's (CDBH's) Office of Compliance. (21CFR Par the Center for Devices and Radional realling by reference to oremarket notification" (21CFR Part note the regulation entitled, "Misbranding by reference under the MDR regulat note the regulation entitled, " Misoraliung by reference to premiers" CFR Part 803), please go to CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of http://www.ida.gov/Micdicalles/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the You may obtain other general informational and Consumer Assistance at its toll-free number Division of Small Manazzone of its Internet address (800) 038-2041 01 (501) 770-7100 01 at the researcesfor You/Industry/default.htm. Sincerely yours, Carlare Buelu Mark N. Melkers Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Traditional 510(k) Submission - Additions to the EXPEDIUM® 5.5mm & 6.35mm Spine Systems ## INDICATIONS FOR USE STATEMENT 4. 510(k) Number (if known): K101070 AUG 31 2010 10/070 Device Name: EXPEDIUM® Spine System Indications For Use: The EXPEDIUM Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The EXPEDIUM Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion in skeletally mature patients. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices KI01070 510(k) Number_
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...