QUANTUM/ STREAMLINE SPINAL SYSTEM MODEL 02-SERIES AND 10-SERIES
Applicant
Pioneer Surgical Technology
Product Code
NKB · Orthopedic
Decision Date
Aug 11, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Streamline TL Spinal System components are non- cervical spinal fixation devices intended as an adjunct to fusion for use as a pedicle screw (T1-S2), sacral/iliac screw fixation or as an anterolateral fixation system (T8 - L5). Pedicle screw fixation is limited to skeletally mature patients. These devices are indicated for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthosis, and failed previous fusion.
Device Story
Streamline TL Spinal System is a posterior spinal fixation system consisting of polyaxial screws, rods, transverse connectors, and set screws. Components are assembled into a tailor-made construct to provide immobilization and stabilization of spinal segments as an adjunct to fusion. Used in thoracic, lumbar, and sacral/iliac spine. Operated by surgeons in an OR setting. The system uses a threaded set screw/cap mechanism to lock the rod to the screw head. Mechanical performance is verified through static and fatigue compression bending and static torsion testing. The device provides structural support to the spine, facilitating fusion and potentially reducing pain and correcting deformities in patients with degenerative or traumatic spinal conditions.
Clinical Evidence
Bench testing only. Mechanical performance characterized via static and fatigue compression bending and static torsion per ASTM F1717. Additional internal testing evaluated crosslink disassociation, saddle/cap separation, and yoke disassociation from the screw head.
Technological Characteristics
Materials: Titanium Alloy (ASTM F136) for screws, set screws, and connectors; Titanium Alloy (ASTM F136) or Cobalt Chromium Molybdenum Alloy (ASTM F1537) for rods. Design: Polyaxial screw system with threaded locking mechanism. Connectivity: None. Sterilization: Not specified. Software: None.
Indications for Use
Indicated for skeletally mature patients requiring spinal fixation as an adjunct to fusion. Conditions include degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal deformities (scoliosis, kyphosis, lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion. Used for pedicle screw (T1-S2), sacral/iliac, or anterolateral (T8-L5) fixation.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}------------------------------------------------
## AUG 1 1 2010
| 3.0 | 510 (k) Summary | |
|-----|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Sponsor: | Pioneer Surgical Technology<br>375 River Park Circle<br>Marquette, MI 49855<br>(906) 225-5602<br>Contact: Emily M. Downs<br>Date prepared: August 10, 2010 |
| | Device Name: | Streamline TL Spinal System |
| | Classification Name: | $$888.3060$ , Spondylolisthesis Spinal Fixation Device System and<br>Pedicle Screw Spinal System - $$888.3070$ , Class III. |
| | Product Codes: | NKB, KWQ, MNH, MNI, Panel Code 87 |
| | Predicate Device: | K080518 - Quantum Spinal System (SE date - March 20, 2008)<br>K080504 - LowTop Spinal Rod System (SE date - March 20, 2008) |
| | | K003061 - SYNTHES USS (10.0mm Side Opening Screws)<br>K070973 - Quantum Spinal System (SE date - July 3, 2007) |
| | Description: | The Streamline TL Spinal System consists of a variety of rods,<br>polyaxial screws, transverse connectors, set screws and other<br>connecting components used to build a spinal construct.<br>Instrumentation is also available to facilitate implantation of the<br>device components. |
| | | The Streamline TL Spinal System is intended to help provide<br>immobilization and stabilization of spinal segments as an adjunct to<br>fusion of the thoracic, lumbar, and/or sacral/iliac spine. Sacral/iliac<br>screws are designed for posterior fixation. |
| | | The implant components can be rigidly locked into a variety of<br>configurations, with each construct being tailor-made for the<br>individual case. |
| | | Screws, Set Screws, and Connecting components are comprised of<br>Titanium Alloy per ASTM F 136. Rods are comprised of Titanium<br>Alloy per ASTM F 136 or Cobalt Chromium Molybdenum Alloy<br>per ASTM F 1537. |
| | Intended Use: | The Streamline TL Spinal System components are non-cervical<br>spinal fixation devices intended as an adjunct to fusion for use as a<br>pedicle screw (T1-S2) sacral/iliac screw fixation or as an<br>anterolateral fixation system (T8 - L5). Pedicle screw fixation is<br>limited to skeletally mature patients. These devices are indicated for<br>all of the following indications: degenerative disc disease (defined<br>as discogenic back pain with degeneration of the disc confirmed by<br>history and radiographic studies), spondylolisthesis, trauma, (i.e.,<br>fracture or dislocation), deformities or curvatures (i.e., scoliosis, |
.
{1}------------------------------------------------
.
| Technological Characteristics | kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthosis, and failed previous fusion. |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Streamline TL Spinal System consists of a range of polyaxial screw lengths and diameters that match that of predicate systems. The rods and connecting components are identical to that of the predicate system. Intended Use and Materials are also similar to that of the predicate systems. |
| | Both the Streamline TL and the predicate system use a component (set screw/ cap) that connects to the screw yoke to lock down the rod/ screw head. The mechanism by which this component locks to the screw head is different between the Streamline TL system (threaded) and that of the Pioneer predicate system (cam lock). |
| Performance Data: | Mechanical testing was presented to characterize construct and component performance, including testing static and fatigue compression bending and static torsion per recognized ASTM F1717 as outlined in FDA's "Guidance for Industry and Staff: Spinal Systems 510(k)s" issued May 3, 2004, |
| | Additional testing was performed per internal protocols to characterize component performance and included evaluation of crosslink disassociation, saddle/ cap separation, and yoke disassociation from screw head. |
| | The test results of verification testing demonstrate that the mechanical performance of the Pioneer Streamline TL Spinal System is substantially equivalent to the predicate devices. |
| Performance and SE Determination: | Comparisons of device performance data, materials, indications and design/function to predicate devices were provided in making a determination of substantial equivalence. |
.
:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird in flight, composed of three curved lines.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
AUG 11 2010
Pioneer Surgical Technology % Ms. Emily M. Downs Regulatory Affairs Associate 375 River Park Circle Marquette, Michigan 49855-0627
Re: K093692
Trade/Device Name: Streamline TL Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, KWQ, MNH, MNI Dated: June 08, 2010 Received: June 09, 2010
Dear Ms. Downs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
### Page 2 - Ms. Emily M. Downs
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Sincerely yours,
Mark N. Melkerson
Director
Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# K093692
## AUG 1 1 2010
#### 2.0 Indications for Use Statement
| 510(k) Number (if known): | K093692 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Streamline TL Spinal System |
| Indications for Use: | The Streamline TL Spinal System components are non- cervical spinal fixation devices intended as an adjunct to fusion for use as a pedicle screw (T1-S2), sacral/iliac screw fixation or as an anterolateral fixation system (T8 - L5). Pedicle screw fixation is limited to skeletally mature patients. These devices are indicated for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthosis, and failed previous fusion. |
Prescription Use
✓
OR
Over-the-Counter Use
(Per 21 CFR 801.109)
Over-the-Counter Use
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
(Division Sign-On) Division of Surgical, Orthopedic, and Hestorative Devices
510(k) Number_K093692
1