← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K093668

# SYNTHES MATRIX MIS RODS (K093668)

_Synthes (Usa) · NKB · Mar 9, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K093668

## Device Facts

- **Applicant:** Synthes (Usa)
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Mar 9, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The Synthes USS (including Matrix, USS Side-Opening, USS Dual-Opening, USS Small Stature (which includes small stature and pediatric patients), USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, USS Iliosacral, and ClampFix) are non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation systems (T1-S2/ilium), a posterior hook fixation system (T1-L5), or as an anterolateral fixation system (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, and failed previous fusion (pseudoarthrosis). When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct. When used with the 3.5/6.0-mm parallel connectors, the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, USS Iliosacral, and ClampFix) can be linked to the CerviFix 3.5 mm Systems (including CerviFix, Axon, and Synapse). In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix 3.5 mm Systems (including CerviFix, Axon, and Synapse). When used with the 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, USS Iliosacral, and ClampFix). In addition, Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, USS Iliosacral, and ClampFix) can be interchanged with all USS 6.0 mm rods and transconnectors.

## Device Story

Synthes Matrix MIS Rods are titanium spinal rods designed for minimally invasive posterior pedicle screw fixation. Used as part of the Synthes Matrix System; rods are straight or curved with varying radii and lengths. Implanted by surgeons in clinical settings to stabilize the spine during fusion procedures. Output is mechanical stabilization of spinal segments. Benefits include reduced surgical invasiveness compared to traditional open approaches.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Titanium spinal rods; straight or curved configurations; various lengths and bend radii. Designed for minimally invasive implantation as part of the Synthes Matrix System. Non-cervical spinal fixation.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Reference Devices

- CerviFix 3.5 mm Systems
- Axon
- Synapse

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

. . . .

## @ SVNTHES® Spine

ﺔ,

MAR - 9 2010

| Name of Firm:                                            | Synthes Spine<br>1302 Wrights Lane East<br>West Chester, PA 19380                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact:                                          | Heather Guerin<br>Regulatory Affairs Specialist<br>Telephone: 610-719-5432<br>Facsimile: 610-719-5102<br>Email: querin.heather@synthes.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Date Prepared:                                           | February 8, 2010                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Trade Name:                                              | Synthes Matrix MIS Rods                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Classification:                                          | 21 CFR 888.3070 - Pedicle screw spinal system<br>Class III<br>Orthopaedic and Rehabilitation Devices Panel<br>Product Code: NKB, MNH, MNI, KWQ, KWP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Predicates:                                              | The Synthes Matrix MIS Rods are substantially equivalent to similar, previously<br>cleared pedicle screw spinal systems.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device Description:                                      | The Synthes Matrix MIS Rods are spinal rods to be used as part of the Synthes<br>Matrix System of posterior pedicle screw fixation implants. The titanium rods are<br>available straight or curved in two bend radii and a range of lengths. The Synthes<br>Matrix MIS rods are to be implanted via a minimally invasive approach.<br><br>The Synthes Matrix MIS Rods are intended to be used with the Synthes Matrix<br>System Implants. The Synthes Matrix Implants consist of a family of non-cervical<br>spinal fixation devices intended for use as a posterior pedicle screw fixation system<br>(T1-S2), posterior hook fixation system (T1-L5) or anterolateral fixation system (T8-L5).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Intended Use/<br>Indications for Use:                    | The Synthes USS (including Matrix, USS Side-Opening, USS Dual-Opening, USS<br>Small Stature (which includes small stature and pediatric patients), USS VAS<br>variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea,<br>Pangea Monoaxial, USS Polyaxial, USS Iliosacral, and ClampFix) are non-cervical<br>spinal fixation devices intended for use as posterior pedicle screw fixation systems<br>(T1-S2/ilium), a posterior hook fixation system (T1-L5), or as an anterolateral<br>fixation system (T8-L5). Pedicle screw fixation is limited to skeletally mature<br>patients with the exception of the Small Stature USS. These devices are indicated<br>as an adjunct to fusion for all of the following indications regardless of the intended<br>use: degenerative disc disease (defined as discogenic back pain with degeneration<br>of the disc confirmed by history and radiographic studies), spondylolisthesis,<br>trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis,<br>kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, and failed<br>previous fusion (pseudoarthrosis).<br><br>When treating patients with Degenerative Disc Disease (DDD), transverse bars are<br>not cleared for use as part of the posterior pedicle screw construct.<br><br>When used with the 3.5/6.0-mm parallel connectors, the Synthes USS (including<br>USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS<br>Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial,<br>USS Iliosacral, and ClampFix) can be linked to the CerviFix 3.5 mm Systems<br>(including CerviFix, Axon, and Synapse). In addition, when used with 3.5/5.0 mm<br>parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix |
|                                                          | 3.5 mm Systems (including CerviFix, Axon, and Synapse). When used with the<br>5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the<br>Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS<br>variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea,<br>Pangea Monoaxial, USS Polyaxial, USS Iliosacral, and ClampFix).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                          | In addition, Synthes USS (including USS Side-Opening, USS Dual-Opening, USS<br>VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea,<br>Pangea Monoaxial, USS Polyaxial, USS Iliosacral, and ClampFix) can be<br>interchanged with all USS 6.0 mm rods and transconnectors.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Comparison of the<br>device to predicate<br>device(s):   | The Synthes Matrix MIS Rods are a result of modifications to the design and<br>surgical technique of the predicate devices. They are substantially equivalent to<br>the predicates in design, function, material and intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| - Performance Date<br>(Non-Clinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions:<br>The information provided demonstrates that the Synthes Matrix MIS Rods are<br>substantially equivalent to the predicate devices.<br>Clinical Performance and Conclusions:<br>Clinical data and conclusions were not needed for this device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

{1}------------------------------------------------

## @ SYNTHES * Spine

.

and the comments of the comments of

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a stylized caduceus symbol. The caduceus is a symbol of medicine and health, often depicted as a staff with two snakes coiled around it.

## Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

Synthes Spine % Ms. Heather Guerin Regulatory Affairs Specialist 1302 Wrights Lane East Chester, Pennsylvania 19380

## MAR - 9 2010

Re: K093668

Trade/Device Name: Synthes Matrix MIS Rods Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI, KWQ, KWP Dated: February 08, 2010 Received: February 17, 2010

Dear Ms. Guerin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Ms. Heather Guerin

CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events} (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Barbara buchin

Mark N. Melk Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number: K093668 Device Name: Synthes Matrix MIS Rods

Indications for Use:

The Synthes USS (including Matrix, USS Side-Opening, USS Dual-Opening, USS Small Stature (which includes small stature and pediatric patients), USS VAS variable axis components, USS Fracture, Click'X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, USS Polyaxial, USS liiosacral, and ClampFix) are non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation systems (T1-S2/ilium), a posterior hook fixation system (T1-L5), or as an anterolateral fixation system (T8-L5). Pedicle scew fixation is limited to skeletally mature patients with the exception of the Small Stature USS. These devices are indicated as an adjunct to fusion for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), turnor, stenosis, and failed previous fusion (pseudoarthrosis).

When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.

When used with the 3.5/6.0-mm parallel connectors, the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea, Pangea Monoaxial, USS Polyaxial, USS Iliosacral, and ClampFix), can be linked to the CerviFix 3.5 mm Systems (including CerviFix, Axon, and Synapse). In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix 3.5 mm Systems (including CerviFix, Axon, and Synapse). When used with the 5.0/6.0mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea Monoaxial, USS Polyaxial, USS Polyaxial, USS Hibsacral, and ClampFix).

In addition, Synthes USS (including USS Side-Opening, USS Dual-Opening, USS VAS variable axis components, USS Fracture, Click'X, Click'X Monoaxial, Pangea Monoaxial, USS Fracture, Polyaxial, USS Iliosacral, and ClampFix) can be interchanged with all USS 6.0 mm ods and transconnectors.

Prescription Use × (21 CFR 801 Subpart D)

AND / OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

signature

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K093668

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K093668](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K093668)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
