← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K091797

# TSRH SPINAL SYSTEM (K091797)

_Medtronic Sofamor Danek · NKB · Apr 30, 2010 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K091797

## Device Facts

- **Applicant:** Medtronic Sofamor Danek
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Apr 30, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The TSRH® Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine.

## Device Story

TSRH® Spinal System is a modular orthopedic implant set for spinal stabilization and fusion. Components include rods, hooks, screws, cross connectors, staples, and plates; fabricated from stainless steel, titanium alloy, or cobalt-chromium-molybdenum alloy; some components feature hydroxyapatite coating. System is used by surgeons in clinical/OR settings to create patient-specific constructs for thoracic, lumbar, and sacral spine stabilization. Device provides rigid immobilization to facilitate bone fusion. Surgeons select and assemble components intraoperatively to address specific spinal deformities or instabilities. Benefits include structural support during the fusion process, potentially reducing pain and correcting spinal alignment.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing only, specifically mechanical testing of hydroxyapatite coating properties per ASTM standards.

## Technological Characteristics

Materials: medical grade stainless steel, titanium alloy (ASTM F 136/ISO 5832-3), and cobalt-chromium-molybdenum alloy. Components: rods, hooks, screws, cross connectors, staples, plates. Coating: hydroxyapatite. Mechanical testing: Static Tensile Testing (ASTM F1147-05), Coating Thickness Analysis (ASTM F1854-01). Sterilization: sterile or non-sterile. Connectivity: N/A (mechanical implant).

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- TSRH® Spinal System ([K081080](/device/K081080.md))
- TSRH® Spinal System ([K052054](/device/K052054.md))
- CD HORIZON® Spinal System ([K043151](/device/K043151.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K091797 page 1 of 3

## TSRH® Spinal System 510(k) Summary

APR 3 0 2010

### April 2010

| I. | Company: | Medtronic Sofamor Danek USA<br>1800 Pyramid Place<br>Memphis, Tennessee 38132<br>Telephone: (901) 396-3133<br>Fax: (901) 346-9738 |
|----|----------|-----------------------------------------------------------------------------------------------------------------------------------|
|    | Contact: | Brad Sheals<br>Senior Regulatory Affairs Specialist                                                                               |

- Proposed Proprietary Trade Name: TSRH® Spinal System ll.
- III. Classification Name(s): Spinal Interlaminal Fixation Orthosis, Spinal Intervertebral Body Fixation Orthosis, Spondylolisthesis Spinal Fixation Device System, and Pedicle Screw Spinal System (per 21CFR Section 888.3050, 888.3060, and/or 888.3070); Product Code(s): KWP, MNI, MNH, NKB
- Description: The TSRH® Spinal System is intended to help provide immobilization and III. stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine.

The TSRH® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, cross connectors, staples, plates and connecting components as well as implant components from other Medtronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

Certain implant components from other Medtronic spinal systems can be used with the TSRH® Spinal System. These components include GDLH® rods, GDLH® rod/bolt connectors, GDLH® Variable Angle T-Bolts, GDLH® set screws and locking screws, DYNALOK® PLUS™ bolts, CD HORIZON® Low Profile MULTI-SPAN® CROSSLINK® Plates, VANTAGE™ Anterior Fixation System screws, and CD HORIZON® rods, set screws and locking screws.

The hooks are intended for posterior use only and the staples are for anterior use only. The TSRH-3D® and TSRH-3Dx™ connectors and TSRH-3D® and TSRH-3Dx™ screws are intended for posterior use only. Within the TSRH® family, the cobalt chromium rods should only be used with TSRH® 3Dx™ Spinal System. All CROSSLINK® Plates are for posterior use and the CROSSLINK® Axial and Offset Plates may be used anteriorly as well.

The TSRH® Spinal System components are fabricated from medical grade stainless steel, medical grade titanium alloy, and/or medical grade cobalt-chromiummolybdenum alloy. Medical grade titanium, titanium alloy, and/or cobalt-chromiummolybdenum alloy may be used together. Certain TSRH® Spinal System components may be coated with hydroxyapatite. The subject components will be manufactured from medical grade titanium alloy described by such standards as ASTM F 136 or ISO 5832-3. The TSRH® Spinal System may be sold sterile or non-sterile.

{1}------------------------------------------------

# K091797 Page Zof 3

The purpose of this 510(k) submission is to include hydroxyapatite coated screws into the TSRH® Spinal System.

#### IV. Indications for Use:

When used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, the TSRH® Spinal System is indicated as an adjunct to fusion for one or more of the following: {} degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) degenerative spondylolisthesis with objective evidence of neurologic impairment, (3) fracture, (4) dislocation, (5) scoliosis, (6) kyphosis. (7) spinal tumor, and/or (8) failed previous fusion (pseudarthrosis).

In addition, when used as a pedicle screw fixation system, the TSRH® Spinal System is indicated as an adjunct to fusion for skeletally mature patients: (1) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint: (2) who are receiving fusions using autogenous bone graft only: (3) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (4) who are having the device removed after the development of a solid fusion mass.

When used as a posterior, non-cervical, non-pedicle screw fixation system, the TSRH& Spinal System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spondylolisthesis, (3) fracture, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) spinal stenosis, (6) pseudarthrosis, (7) tumor resection, and/or (8) unsuccessful previous attempts at spinal fusion.

When used as a unilateral supplemental fixation device in the antero-lateral thoracic/Jumbar region, the TSRH® L-Plate and VANTAGE™ screws are intended for the following indications: spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthritis; and/or failed previous fusion.

For anterior use only the TSRH® Spinal System has the additional indication of: spondylolysis.

- Identification of the Legally Marketed Predicate Devices Use to Claim Substantial V. Equivalence: The design features, materials and indications for use of the TSRH& Spinal System are substantially equivalent to predicate TSRH® Spinal System components previously cleared in K081080 (S.E. 11/21/2008), K052054 (S.E. 08/16/2005) and predicate CD HORIZON® Spinal System components cleared in K043151 (S.E. 02/01/2005).

{2}------------------------------------------------

K091797 page 3 of 3

#### VI. Brief Discussion of the Non-Clinical Tests Submitted

For a determination of substantial equivalence, the following non-clinical mechanical tests were performed:

> Static Tensile Testing per ASTM F1147-05 (FDA Recognition No. 8-113) > Average Coating Thickness Analysis per ASTM F1854-01 (FDA Recognition No. 8-85)

Existing testing was also provided to demonstrate the requirements of the FDA guidance document "510(k) Information Needed for Hydroxyapatite Coated Implants".

#### VII. Conclusions Drawn from the Non-Clinical Tests

Based on the supporting documentation within this premarket notification, the subject device demonstrates substantial equivalence to the listed predicate devices.

©2010 MEDTRONIC SOFAMOR DANEK USA, Inc. All rights reserved.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

# APR 3 0 2010

Medtronic Sofamor Danek USA % Mr. Brad Sheals Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132

Re: K091797

Trade/Device Name: TSRH® Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH, KWP Dated: March 01, 2010 Received: March 02, 2010

Dear Mr. Sheals:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 - Mr. Brad Sheals

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Barbara Buehrle

Mark N. Melke Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K091797

510(k) Number (if known):

TSRH® Spinal System Device Name:

Indications for Use:

When used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients using bone graft, the TSRH® Spinal System is indicated as an adjunct to fusion for one of the following: (1) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) degenerative spondylolisthesis with objective evidence of neurologic impairment, (3) fracture, (4) dislocation, (5) scoliosis, (6) kyphosis, (7) spinal tumor, and/or (8) failed previous fusion (pseudarthrosis).

In addition, when used as a pedicle screw fixation system, the TSRH® Spinal System is indicated as an adjunct to fusion for skeletally mature patients using bone graft: (1) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint: (2) who are receiving fusions using autogenous bone graft only: (3) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (4) who are having the device removed after the development of a solid fusion mass.

When used as a posterior, non-cervical, non-pedicle screw fixation system, the TSRH® Spinal System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spondylolisthesis, (3) fracture, (4) spinal deformities (i.e., scoliosis, and/or lordosis), (5) spinal stenosis, (6) pseudarthrosis, (7) tumor resection, and/or (8) unsuccessful previous attempts at spinal fusion.

When used as a unilateral supplemental fixation device in the antero-lateral thoracic/lumbar region, the TSRH® L-Plate and VANTAGE™ screws are intended for the following indications: spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthritis; and/or failed previous fusion.

For anterior use only the TSRH® Spinal System has the additional indication of: spondylolysis.

AND/OR

Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

W

(Division Sign-Off Division of Surgical, Orthopedic, and Restorative Devices

K091797 510(k) Number_

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K091797](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K091797)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
