← Product Code [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB) · K082797

# BLACKSTONE PEDICLE SCREW SYSTEM (K082797)

_Blackstone Medical, Inc. · NKB · Oct 17, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K082797

## Device Facts

- **Applicant:** Blackstone Medical, Inc.
- **Product Code:** [NKB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB.md)
- **Decision Date:** Oct 17, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Blackstone Pedicle Screw System is intended for posterior, non-cervical pedicle fixation. Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for all of the following indications regardless of the intended use: - 1) degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies) - 2) spondylolisthesis. - 3) trauma (i.e., fracture or dislocation). - 4) spinal stenosis. - 5) deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), - 6) tumor, - 7) pseudoarthrosis, and - 8) failed previous fusion The Blackstone Pedicle Screw System components are used with certain components of the Blackstone SFS system, including rods, rod connectors and cross-connectors.

## Device Story

Blackstone Pedicle Screw System (4.0 mm diameter screws) provides posterior, non-cervical spinal fixation; used as adjunct to fusion. System comprises multiaxial and monoaxial pedicle screws; integrates with Blackstone SFS system components (rods, connectors, cross-connectors). Implanted by surgeons in clinical/OR settings to stabilize spinal segments; facilitates fusion in patients with degenerative disc disease, trauma, deformities, or other spinal pathologies. Mechanical performance verified via bench testing to ensure structural integrity equivalent to predicate systems.

## Clinical Evidence

Bench testing only; no clinical data provided.

## Technological Characteristics

Multiaxial and monoaxial pedicle screws; 4.0 mm diameter; non-cervical spinal fixation; metallic construction; manual surgical instrumentation.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Blackstone Medical, Inc. Pedicle Screw System ([K081684](/device/K081684.md))
- Synthes Pangea Spine System ([K052123](/device/K052123.md))
- Synthes Pangea Spine System ([K052151](/device/K052151.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# OCT 1 7 2008

Premarket Notification 510(k) Blackstone Medical, Inc. Blackstone Pedicle Screw System Modification Confidential

## 510(k) SUMMARY

# Blackstone Medical, Inc. Pedicle Screw System – 4.0 mm Diameter Screws

| Sponsor:                  | Blackstone Medical, Inc.<br>1211 Hamburg Turnpike<br>Suite 300<br>Wayne, NJ 07470            |
|---------------------------|----------------------------------------------------------------------------------------------|
| Registration Number:      | 3004606875                                                                                   |
| Contact Person:           | Whitney G. Törning, Senior Director of Regulatory Affairs<br>& Quality Assurance             |
| Telephone Number:         | 973.406.2838                                                                                 |
| Fax Number:               | 973.406.2938                                                                                 |
| Email:                    | wtorning@blackstonemedical.com                                                               |
| Submitter:                | Martin G. Sprunck<br>Regulatory Affairs Manager                                              |
| Manufacturer:             | Blackstone Medical, Inc.<br>90 Brookdale Drive<br>Springfield, MA 01104                      |
| Registration Number:      | 1225457                                                                                      |
| Contract Manufacturer:    | Pulse Technologies<br>2000 AM Drive<br>Quakertown, PA 18951                                  |
| System Name:              | Blackstone Pedicle Screw System                                                              |
| Trade Name:               | 4.0 mm Diameter Pedicle Screws                                                               |
| Common Name:              | Posterior Thoracolumbar System                                                               |
| Product Code:             | NKB - Orthosis, Spinal Pedicle Fixation, for Degenerative<br>Disc Disease                    |
| Subsequent Product Codes: | MNI - Orthosis, Spinal Pedicle Fixation<br>MNH - Orthosis, Spondylolisthesis Spinal Fixation |

{1}------------------------------------------------

## 14082797

Premarket Notification 510(k) Blackstone Medical, Inc. Blackstone Pedicle Screw System Modification Confidential

| Regulatory Classifications: | Class III Preamendment Device, 888.3070 - Pedicle Screw<br>Spinal System - *Class III Summary and Certification<br>Required<br>Class II - 888.3070 - Pedicle Screw Spinal System |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Review Panel:               | Orthopedic Device Panel                                                                                                                                                          |
| Predicate Devices:          | Blackstone Medical, Inc. Pedicle Screw System (K081684<br>SE 9/15/08)                                                                                                            |
|                             | Synthes Pangea Spine System (K052123 SE 9/23/05 &<br>K052151 SE 12-7-05)                                                                                                         |

### Intended Use / Indications for Use

The Blackstone Pedicle Screw System is intended for posterior, non-cervical pedicle fixation. Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for all of the following indications regardless of the intended use:

- 1) degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)
- 2) spondylolisthesis.
- 3) trauma (i.e., fracture or dislocation).
- 4) spinal stenosis.
- 5) deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
- 6) tumor,
- 7) pseudoarthrosis, and
- 8) failed previous fusion

The Blackstone Pedicle Screw System components are used with certain components of the Blackstone SFS system, including rods, rod connectors and cross-connectors.

#### Technological Characteristics

The Blackstone Pedicle Screw System 4.0 mm Diameter Screws consists of an assortment of multiaxial and monoaxial pedicle screws.

#### Performance Data

Mechanical testing of the Blackstone Pedicle Screw System 4.0 mm Diameter Screws was conducted which demonstrates that the system is substantially equivalent to predicate devices that have the same intended use, similar indications, technological characteristics and principles of operation.

{2}------------------------------------------------

# K082797

Premarket Notification 510(k) Blackstone Medical, Inc. Blackstone Pedicle Screw System Modification Confidential

#### Basis of Substantial Equivalence

Mechanical testing was conducted to demonstrate that the additional 4.0 mm diameter screw components are substantially equivalent to the current Blackstone Pedicle Screw System, (K081684 SE 9/15/08), which has been cleared by FDA for the purpose of building a spinal implant construct in the non-cervical spine.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/3/Picture/12 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus-like symbol, with three diagonal lines representing the branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Blackstone Medical, Inc. % Ms. Whitney G. Törning Senior Director of Regulatory Affairs & Quality Assurance 1211 Hamburg Turnpike, Suite 300 Wayne, New Jersey 07470

OCT 1 7 2008

Re: K082797

Trade/Device Name: Blackstone Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Names: Pedicle screw spinal system. Regulatory Class: III Product Code: NKB, MNI, MNH Dated: September 22, 2008 Received: September 23, 2008

Dear Ms. Törning:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

#### Page 2 - Ms. Whitney G. Törning

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Mark M. Milliman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### Indications for Use Statement

510(k) Number (if known):_ 1408 2797

System Name: Blackstone Pedicle Screw System Device Name: 4.0 mm Diameter Pedicle Screws

Indications for Use:

The Blackstone Pedicle Screw System is intended for posterior, non-cervical pedicle fixation. Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for all of the following indications regardless of the intended use:

- 1) degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)
- 2) spondylolisthesis,
- 3) trauma (i.e., fracture or dislocation),
- 4) spinal stenosis,
- 5) deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
- 6) tumor.
- 7) pseudoarthrosis, and
- 8) failed previous fusion

The Blackstone Pedicle Screw System components are used with certain components of the Blackstone SFS system, including rods, rod connectors and cross-connectors.

Prescription Use X AND/OR Over-The-Counter Use (Part 21 C.F.R. 801 Subpart D) (21 C.F.R. 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

NithRPoshe for nxn

Page 1 of 1

Division Sign-Off Division of General. Restorative. and Neurological Devices

510(k) Number KO 82797

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K082797](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/NKB/K082797)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
