XIA SPINAL SYSTEM, STAINLESS STEEL UNI-PLANAR SCREWS
K080928 · Stryker Spine · NKB · Apr 30, 2008 · Orthopedic
Device Facts
| Record ID | K080928 |
| Device Name | XIA SPINAL SYSTEM, STAINLESS STEEL UNI-PLANAR SCREWS |
| Applicant | Stryker Spine |
| Product Code | NKB · Orthopedic |
| Decision Date | Apr 30, 2008 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 888.3070 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Xia® Spinal System is intended for anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The 6 mm diameter rods from the DIAPASON™ Spinal System and OPUS™ Spinal System are intended to be used with the other components of the Xia® Titanium Spinal System. The Titanium Multi-axial Cross Connectors are intended to be used with the other components of the Xia® Titanium Spinal System.
Device Story
Xia Spinal System line extension adds stainless steel uniplanar screws to existing spinal fixation platform; system components include screws, rods, hooks, blockers, and connectors; used by surgeons in spinal fusion procedures to provide stabilization; device components implanted during surgery to fixate spinal segments; mechanical stability provided by rigid construct; benefits include correction of spinal deformities and stabilization of unstable spinal segments.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Components manufactured from stainless steel; uniplanar screw design; intended for noncervical spinal fixation; mechanical properties verified via bench testing to be comparable to predicate system.
Indications for Use
Indicated for patients requiring noncervical pedicle and non-pedicle spinal fixation for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal curvatures (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
- Stryker Spine Xia® Spinal System (K060361)
- Stryker Spine Xia® Spinal System (K053115)
- Stryker Spine Xia® Spinal System (K013823)
Reference Devices
- DIAPASON™ Spinal System
- OPUS™ Spinal System
Related Devices
- K051973 — MODIFICATION TO: XIA STAINLESS STEEL SYSTEM · Stryker Spine · Aug 18, 2005
- K031090 — MODIFICATION TO XIA STAINLESS STEEL SYSTEM · Howmedica Osteonics Corp. · Apr 24, 2003
- K053115 — XIA SPINAL SYSTEM · Stryker Spine · Dec 6, 2005
- K031893 — XIA SPINAL SYSTEM AND XIA STAINLESS STEEL SYSTEM · Howmedica Osteonics Corp. · Jul 17, 2003
- K052181 — MODIFICATION TO XIA SPINAL SYSTEM · Stryker Spine · Sep 8, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
K080928
Xia Spina System Line Extension - Uniplanar Screws
Special 510(k) Premarket Notification
## Special 510(k) Summary: Line Extension to the Xia® Spinal System
## APR 3 0 2008
### Submission Information
| Name and Address of the Sponsor of the 510(k) Submission: | Stryker Spine<br>2 Pearl Court<br>Allendale, NJ 07401 |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Contact Person: | Curtis Truesdale<br>Regulatory Affairs Project Manager<br>Telephone: (201) 760-8296<br>Fax: (201) 760-8496 |
| Date of Summary Preparation: | March 19, 2008 |
## Device Identification
| Proprietary Name: | Xia® Spinal System |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Spinal Fixation Appliances |
| Classification Name and Reference: | Spinal Interlaminal Fixation Orthosis,<br>21 CFR §888.3050<br>Spinal Intervertebral Body Fixation Orthosis<br>21 CFR §888.3060<br>Pedicle Screw Spinal System<br>21 CFR 888.3070 (b) (1) & (b) (2) |
| Product Codes: | NKB, KWP, KWQ, MNH, MNI |
## Predicate Device Information:
Stryker Spine Xia® Spinal System (K060361, K053115, K013823)
#### Predicate Device Identification
The Stryker Spine Xia Spinal System consists of Monoaxial and Polyaxial Screws, Washer, Hooks, Blocker, Rods, Staples, and Connectors. The components are manufactured from either Stainless Steel or Titanium material (Ti alloy and CP Titanium).
{1}------------------------------------------------
#### Description of Device Modification
This submission is intended to address a line extension to Xia® Spinal System. The line extension includes the addition of Xia Stainless Steel Uni-Planar Screws.
#### Intended Use:
The Xia® Spinal System is intended for anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indicating: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
The 6 mm diameter rods from the DIAPASON™ Spinal System and OPUS™ Spinal System are intended to be used with the other components of the Xia® Titanium Spinal System. The Titanium Multi-axial Cross Connectors are intended to be used with the other components of the Xia® Titanium Spinal System.
### Statement of Technological Comparison:
The subject components share the same intended use, material, and basic design concepts as that of the predicate device: Stryker Spine Xia® Spinal System (K060361, K053115 and K013823). Mechanical testing also demonstrated comparable mechanical properties to the predicate device: Spinc Xia Spinal System (K060361, K053115 and K013823)
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces left. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# APR 3 0 2008
Stryker Spine % Mr. Curtis Truesdale Regulatory Affairs Project Manager 2 Pearl Court Allendale, NJ 07401
Re: K080928 Trade/Device Name: Xia Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle Screw Spinal System Regulatory Class: III Product Code: NKB, MNI, MNH, KWQ, KWP Dated: April 1, 2008 Received: April 2, 2008
Dear Mr. Truesdale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Curtis Trucsdale
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Xia Spinal System Line Extension - Uniplanar Screws
Indications for Use:
The Xia® Spinal System is intended for anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (DDD) {defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
The 6mm diameter rods from the DIAPASONTM Spinal System and OPUSTM Spinal System are intended to be used with the other components of the Xia Titanium Spinal System. The Titanium Multi-Axial Cross Connectors are intended to be used with the other components of the Xia® Titanium Spinal System.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | | В ПРИВАЕТ В ПРОДОЖЕНИЕ ПРОДОМ СЕ ПОДОМ ПОД ПОДОМ ПОД ПОДОМ ПОД ПОДОБИЛА В ПРИЗИРАНИЕ ПРИЗИРАНИЕ ПРИЗНАЧИВАНИЕ НА ПРИЗНАЧИВА ПРИЗНАЧИВА ПРИЗНАЧИВА ПОДЕЛИ ПОДЕЛИ ПОДЕЛИ ПОДЕЛИ | |
|--------------------------------------------------------|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| mi hanni | and the contraction of the comments of the control of | | |
1 4. 1978 Jus Man (Division Sign-Off) Division of General, Restorative, and Neurological Devices
Page 1 of 1
**510(k) Number** K080928