The intended use of the Malibu Spinal System iliac monoaxial screws, monoaxial screws, fixed locking caps, rods, hooks, and components, is substantially equivalent to the intended use of the predicate devices. The intended use of the Malibu Spinal System is as a temporary or permanent posterior, non-cervical implant to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusions to occur. The intended use of the Malibu Spinal System, when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System, is to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The indications for use are as follows: degenerative disc disease (DDD) as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies, severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusions by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion (indication for use only when used as a Spondylolisthesis Spinal Fixation Device System), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), spinal tumor, pseudoarthrosis, and/or failed previous fusion. The intended use and indications of the Malibu Spinal System, when used as a Spinal Interlaminal Fixation Orthosis or Hook Spinal System, are limited to T1-L5 and are as follows: degenerative disc disease (DDD) as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies, spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudarthosis, and/or failed previous fusion.
Device Story
Malibu Spinal System is a posterior, non-cervical spinal implant system; includes iliac monoaxial screws, monoaxial screws, fixed locking caps, rods, and hooks. Components are titanium alloy; instruments are stainless steel, Radel, and Pomalux. System provides immobilization and stabilization of spinal segments to facilitate fusion. Used by surgeons in clinical settings; compatible with SeaSpine UCR Spinal System. Supplied non-sterile; requires steam sterilization before use. Output is mechanical stabilization of the spine; clinical benefit is correction of spinal disorders and support of biological fusion process.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Titanium alloy implants; stainless steel, Radel, and Pomalux instruments. Mechanical fixation via screws, hooks, rods, and locking caps. Non-sterile; steam sterilization required. Standalone system compatible with UCR Spinal System components.
Indications for Use
Indicated for skeletally mature patients requiring spinal stabilization/immobilization as adjunct to fusion for DDD (discogenic back pain), spondylolisthesis (including severe L5-S1 grades 3-4), trauma (fracture/dislocation), spinal stenosis, deformities (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion. Pedicle screw system use: thoracic, lumbar, sacral spine. Hook/interlaminal system use: T1-L5.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}------------------------------------------------
# JUL 1 2 2005
CONFIDENTIAL
### Malibu Spinal System
### 510(K) SUMMARY
Pursuant to 510(i) of the Federal Food, Drug, and Cosmetic Act, as amended, and in accordance with 21 CFR § 807.92.
| Submitter Information: | SeaSpine, Inc.<br>Contact: Diana Smith<br>2302 La Mirada Drive<br>Vista, CA 92081-7862<br>Phone: 760-727-8399 Fax: 760-727-8809 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Registration Number: | 2032593 |
| Submission Correspondent: | SeaSpine, Inc.<br>Contact: Diana Smith, Manager of<br>Regulatory Affairs and Quality Assurance<br>2302 La Mirada Drive<br>Vista, CA 92081-7862<br>Phone: 760-727-8399 Fax: 760-727-8809 |
| Date Summary Prepared: | June 15, 2005 |
| Classification Name: | Pedicle Screw Spinal System<br>MNH (Class II) - 888.3070(b)(1)<br>MNI (Class II) - 888.3070(b)(1)<br>NKB (Class III) -- 888.3070(b)(2)<br>Spinal Interlaminal Fixation Orthosis<br>KWP (Class II) – 888.3050 |
| Common/Usual Name: | Iliac Monoaxial Screws, Monoaxial Screws,<br>Fixed locking caps, Rods, Hooks,<br>Components, and Instruments |
| Device Trade Name: | Malibu Spinal System |
The devices used for comparison in this summary are Synthes Spine Company's Synthes Click'X Monoaxial Screw System (K031175), DePuy Acromed's Moss Miami System (K002607 and K992168) and SeaSpine Inc.'s UCR Spinal System (K993503, K021623, K031381, K032739/S1 and K043232).
(The statements of intended use are identical.) 1. Intended Use:
The intended use of the Malibu Spinal System iliac monoaxial screws, monoaxial screws, fixed locking caps, rods, hooks, and components, is substantially equivalent to the intended use of the predicate devices. The intended use of the Malibu Spinal
20
ge 1 of 4
{1}------------------------------------------------
#### Malibu Spinal System
System is as a temporary or permanent posterior, non-cervical implant to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusions to occur.
The intended use of the Malibu Spinal System, when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System, is to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The indications for use are as follows:
- degenerative disc disease (DDD) as defined by back pain of discogenic origin . with degeneration of the disc confirmed by patient history and radiographic studies,
- severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally . mature patients receiving fusions by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion (indication for use only when used as a Spondylolisthesis Spinal Fixation Device System),
- spondylolisthesis, .
- trauma (i.e., fracture or dislocation), .
- spinal stenosis. ●
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis) .
- . spinal tumor,
- pseudoarthrosis, and/or .
- failed previous fusion. .
The intended use and indications of the Malibu Spinal System, when used as a Spinal Interlaminal Fixation Orthosis or Hook Spinal System, are limited to T1-L5 and are as follows:
- degenerative disc disease (DDD) as defined by back pain of discogenic origin . with degeneration of the disc confirmed by patient history and radiographic studies,
- spondylolisthesis .
- trauma (i.e., fracture or dislocation), ◆
- ◆ spinal stenosis,
- . deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
- tumor, .
- pseudarthosis, and/or .
- failed previous fusion. .
### 2. Description:
The Malibu Spinal System includes titanium alloy iliac monaxial and monaxial screws. Both screw types will be offered in numerous diameters and lengths. The screw is comprised of an integrated body and housing. Both screws are designed to
21
{2}------------------------------------------------
## Malibu Spinal System
be part of the Malibu system, but are compatible with and work in conjunction with the components in the current UCR Spinal System. The product is supplied "NON-STERILE" and must be sterilized prior to use.
The Malibu Spinal System will also include titanium alloy caps, rods, and hooks. The caps will come in one size, the rods in two diameters, and the four types of hooks in a variety of sizes. All of these devices are designed to be a part of the stand alone Malibu Spinal System, but are compatible with and work in conjunction with the components in the current UCR Spinal System. The product is supplied "NON-STERILE" and must be sterilized prior to use.
The Malibu Spinal System will offer a wide variety of instruments that range from rod forks to hook pushers. These various instruments will be made primarily from various grades of stainless steel with a few components made from titanium alloy, Radel, and medical grade Pomalux. These items are supplied "NON-STERILE" and must be sterilized prior to use.
### 3. Technological Characteristics:
The iliac monoaxial screws, monoaxial screws, fixed locking caps, rods, hooks, components, and instruments in this notification are components of a new stand alone system called Malibu Spinal System, but are designed to be compatible with and work in conjunction with the components in the current SeaSpine UCR Spinal System. The devices in this submission have substantially equivalent technological characteristics to the predicate devices. Refer to Table 1 in the following section, entitled Comparison Analysis, for a summation of technological characteristics such as design, dimensional specifications, and material.
#### 4. Comparison Analysis:
The overall designs of the Malibu Spinal System iliac monoaxial screws, monoaxial screws, fixed locking caps, rods, hooks, and components are substantially equivalent to the predicate devices. See Table 1 on the following page for a comparison of the Malibu Spinal System iliac monoaxial screws, monoaxial screws, fixed locking caps, rods, hooks, and components to the predicate devices.
22
3 of 4
{3}------------------------------------------------
# Malibu Spinal System
| Feature | Malibu Spinal System | Synthes<br>Click'X<br>Monoaxial<br>System | Moss Miami<br>Spinal System | SeaSpine UCR<br>Spinal System | Substantially<br>Equivalent |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|-----------------------------|-------------------------------|-----------------------------|
| Intended Use | The intended use of the<br>Malibu Spinal System is as<br>a temporary or permanent<br>posterior, non-cervical<br>implant to correct spinal<br>disorders and provide<br>stabilization of the spine to<br>permit the biological<br>process of spinal fusions to<br>occur | Similar | Similar | Similar | Yes |
| Indications for<br>Use | • Degenerative disc disease<br>(DDD) as defined by back<br>pain of discogenic origin<br>with degeneration of the<br>disc confirmed by patient<br>history and radiographic<br>studies,<br>• spondylolisthesis,<br>• trauma (i.e., fracture or<br>dislocation),<br>• spinal stenosis,<br>• deformities or curvatures<br>(i.e., scoliosis, kyphosis,<br>and/or lordosis),<br>• tumor,<br>• pseudarthosis, and/or<br>• failed previous fusion. | Similar | Similar | Similar | Yes |
| Design | Iliac monoaxial screws,<br>monoaxial screws, fixed<br>locking caps, rods, hooks,<br>and components | Similar | Similar | Similar | Yes |
| lliac<br>Monoaxial<br>Screws | See prints | Similar | Similar | Similar | Yes |
| Monoaxial<br>Screws | See prints | Similar | Similar | Similar | Yes |
| Caps | See prints | Similar | Similar | Similar | Yes |
| Laminar<br>Hooks | See prints | Similar | Similar | Similar | Yes |
| Pedicle Hooks | See prints | Similar | Similar | Similar | Yes |
| Offset Hooks | See prints | Similar | Similar | Similar | Yes |
| Material | Titanium alloy | Similar | Similar | Similar | Yes |
| Sterile | Non-sterile | Similar | Similar | Similar | Yes |
| Method of<br>Sterilization | High-temperature steam | Similar | Similar | Similar | Yes |
| Mechanical<br>Strength | See test results | Similar | Similar | Similar | Yes |
Table 1: Summary of Design Comparison
િહ્મુદ
પ્
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three heads, representing health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
JUL 1 2 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Diana Smith Manager of Regulatory Affairs and Quality Assurance SeaSpine, Inc. 2302 La Mirada Drive Vista, California 92081-7862
Re: K051663
Trade/Device Name: Malibu Spinal System Regulation Number: 21 CFR 888.3070, 21 CFR 888.3050 Regulation Name: Pedicle screw spinal system, Spinal interlaminal fixation orthosis Regulatory Class: III Product Code: MNH, MNI, KWP, and NKB Dated: June 15, 2005 Received: June 24, 2005
Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
### Page 2 – Ms. Diana Smith
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will and in your le FDA finding of substantial equivalence of your device to a legally prematics notificate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Atyt Clurdes
Miriam Provost, Ph.D.
Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
2
# Malibu Spinal System
# Indications for Use Statement
510(k) Number (if known): K051663
Device Name: Malibu Spinal System
Indications for Use:
The intended use of the Malibu Spinal System, iliac monoaxial screws, monoaxial I he intended use of the Manbu Dpinal Upinial Upinianta, is substantially equivalent screws, fixed lockling Caps, fous, nooks, and vehipensial - 7 the Malibu Spinal
to the intended use of the predicate device. The intended used immlem to correct to the intended use of the products corrector, non-cervice implant to correct System is as a temporary of permatelli posteries, and the biological process of spinal fusions to occur.
The intended use of the Malibu Spinal System, when used as a Pedice Screw The intended use of the Manbu Spinal System Device System, is to provide Spillal System or Spondylonstiteers Enal segments in skeletally mature patients as immountzanon and staombation of the following acute and chronic instablities an adjulier to fusion in the treatments of and sacral spine. The indications for use are as follows:
- degenerative disc disease (DDD) as defined by back pain of discogenic origin . degeneration of the discuse (222) nfirmed by patient history and radiographic studies,
- studios,
severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally ● severe spondylonomons ( Liaions by autogenous bone graft having implants mature patients Tooering Tastener of ine (L3 to sa crum) with removal of the implants after the attainment of a solid fusion (indication for use only when used as a Spondylolisthesis Spinal Fixation Device System),
- spondylolisthesis, ●
- trauma (i.e., fracture or dislocation), .
- spinal stenosis, .
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis) .
- spinal tumor, .
| Prescription Use | X |
|-----------------------------|------------------------|
| (Part 21 CFR 801 Subpart D) | |
| | AND/OR |
| | Over-The-Counter-Use |
| | (21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED).
(Division Sign Off) Office of Device Evaluation (ODE)
Division of General, Restorative, and Neurological Devices
| 510(k) Number | K051663 |
|---------------|---------|
|---------------|---------|
27
{7}------------------------------------------------
D
# Indications for Use Statement continued
- pseudoarthrosis, and/or .
- failed previous fusion. .
The intended use and indications of the Malibu Spinal System, when used as a The Interlaminal Fixation Orthosis or Hook Spinal System, are limited to Ti-L5 and are as follows:
- rate as follows:
degenerative disc disease (DDD) as defined by back pain of discogenic origin . degenerative discuss (2) ===================================================================================================================================================== studies,
- spondylolisthesis .
- trauma (i.e., fracture or dislocation), .
- spinal stenosis, .
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), .
- . tumor,
- pseudarthosis, and/or .
- failed previous fusion. ●
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X | AND/OR | Over-The-Counter-Use<br>(21 CFR 807 Subpart C) | |
|-------------------------------------------------|---|--------|------------------------------------------------|--|
|-------------------------------------------------|---|--------|------------------------------------------------|--|
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED).
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| 510(k) Number | K051663 |
|---------------|---------|
|---------------|---------|
28
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.