EBI SPINELINK - II SPINAL FIXATION SYSTEM

K031355 · Ebi, L.P. · NKB · Jul 22, 2003 · Orthopedic

Device Facts

Record IDK031355
Device NameEBI SPINELINK - II SPINAL FIXATION SYSTEM
ApplicantEbi, L.P.
Product CodeNKB · Orthopedic
Decision DateJul 22, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

The EBI® SpineLink® II Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and or lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion. The transverse connector is not intended for anterior/anterolateral fixation.

Device Story

Spinal fixation system comprising interconnecting links, screws (fixed/polydirectional), transverse connectors, hooks, endcaps, and locknuts. Used by surgeons in clinical settings to stabilize the spine via pedicle screw, posterior hook, sacral/iliac screw, or anterolateral fixation. Provides mechanical stabilization to facilitate fusion in patients with degenerative, traumatic, or deformity-related spinal conditions. Benefits include structural support for spinal segments.

Clinical Evidence

Bench testing only; no clinical data presented.

Technological Characteristics

Components manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM F136. System includes interconnecting links, screws, transverse connectors, hooks, endcaps, and locknuts. Mechanical fixation device.

Indications for Use

Indicated for skeletally mature patients requiring non-cervical spinal fixation for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), deformities (scoliosis, kyphosis, lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 031355 NGL 1/2 # JUL 2 2 2 2003 ### 510(k) Summary This 510(k) Summary for the EBI® SpineLink® - II Spinal Fixation System is provided as required per Section 513(3) of the Food, Drug and Cosmetic Act. 1. Submitter: EBI, L.P. 100 Interpace Parkway Parsippany, NJ 07054 Contact Person: Frederic Testa, RAC Telephone: (973) 299-9300, ext.2208 Date prepared: June 30, 2003 | 2. Proprietary Name: | EBI® SpineLink® - II Spinal Fixation System | |-----------------------|----------------------------------------------| | Common Name: | Spinal Fixation Device | | Classification Names: | Spinal Intervertebral Body Fixation Orthosis | #### 3. Predicate or legally marketed devices that are substantially equivalent: - Spinal Concepts BacFix Spinal Fixation System (K013887) - Sofamor Danek CD Spinal System (K964159) - Sofamor Danek Townley Pedicle Screw Plating System (K983706) . - 4. Description of the device: The EBT® SpineLink® II Spinal Fixation System is a spinal fixation device that utilizes interconnecting links, screws (fixed and polydirectional designs), transverse connectors, hooks, endcaps, locknuts, and link ties. - 5. Intended Use: The EBI® SpineLink® II Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. {1}------------------------------------------------ i70-55 2/2 Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and or lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion. The transverse connector is not intended for anterior/anterolateral fixation. - 6. Materials: The components of the System are manufactured from titanium alloy (Ti-6A1-4V ELI) per ASTM F136. - 7. Comparison of the technological characteristics of the device to predicate devices: There are no significant differences between the EBI® SpineLink® - II Spinal Fixation System and other legally marketed spinal systems. It is substantially equivalent* to the predicate devices in regards to intended use, materials, and function. Mechanical testing comparing the System to predicate systems demonstrates that the device complies with applicable standards and meets all of its functional requirements. * Any statement made in conjunction with this submission regarding a determination of substantial equivalence to any other product is intended only to relate to whether the product can be lawfully marketed without pre-market approval or reclassification and is not interpreted as an admission or any other type of evidence in patent infringement litigation. [Establishment Registration and Premarket Notification Procedures, Final Regulation, Preamble, August 23, 1977, FR 42520 (Docket No. 76N-0355.)] {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the department's name around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms. The logo is printed in black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## JUL 2 2 2003 Mr. Jon Caparotta, RAC Manager. Regulatory Affairs EBI, L.P. 100 Interpace Parkway Parsippany, New Jersey 07054 K031355 Re: Trade Name: EBI® SpineLink™ II Spinal Fixation System Regulation Number: 21 CFR 888.3070, 888.3060, 888.3050 Regulation Name: Pedicle screw spinal system, Spinal intervertebral body fixation orthosis, Spinal interlaminal fixation orthosis Regulatory Class: III Product Code: NKB, MNI, MNH, KWQ, KWP Dated: April 28, 2003 Received: April 30, 2003 Dear Mr. Caparotta: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ #### Page 2 - Mr. Jon Caparotta, RAC forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Mark N. Melkerson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Statement of Indications for Use Page_1_of_____________________________________________________________________________________________________________________________________________________________________ K031355 ----------------------------------------------------------------------------------------------------------------------------------------------------------------------510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Device Name: EBI® SpineLink® - II Spinal Fixation System Indications For Use: The EBI® SpineLink® II Spinal Fixation System is a non-cervical fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and or lordosis), tumor, stenosis, pseudoarthrosis, and failed previous fusion. The transverse connector is not intended for anterior/anterolateral fixation. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) Mark N. Melkerson Division Sign-O Division of General. Restorative. and Neurological Devices 510(k) Number K031355
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...