WRIGHT PLASTER OF PARIS PELLETS (SUBJECT TO REVISION)

K960978 · Wrightmedicaltechnologyinc · MQV · Jun 21, 1996 · Orthopedic

Device Facts

Record IDK960978
Device NameWRIGHT PLASTER OF PARIS PELLETS (SUBJECT TO REVISION)
ApplicantWrightmedicaltechnologyinc
Product CodeMQV · Orthopedic
Decision DateJun 21, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3045
Device ClassClass 2
AttributesTherapeutic

Intended Use

Wright Plaster of Paris Pellets are intended to be gently packed into non-loadbearing long bone voids. These bone voids may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The Wright Plaster of Paris Pellets provide a bone filler that resorbs and is replaced with bone during the healing process.

Device Story

Biodegradable, radiopaque calcium sulfate dihydrate pellets; used as bone void filler; packed into non-loadbearing long bone defects; resorbed in 30-60 days; replaced by bone during healing; sterile, single-patient use; 0.1g per pellet; supplied in 100-count pouches.

Clinical Evidence

Bench testing only; dissolution studies demonstrate equivalent dissolution rates to predicate device.

Technological Characteristics

Composition: 98% medical grade calcium dihydrate (CaSO4 • 2H2O), 2% stearic acid; form factor: 0.1g pellets; sterile; biodegradable; radiopaque.

Indications for Use

Indicated for filling non-loadbearing long bone voids resulting from surgical procedures or traumatic injury.

Regulatory Classification

Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} JUN 21 1996 K960978 WRIGHT MEDICAL TECHNOLOGY, INC. 5677 AIRLINE ROAD ARLINGTON, TN 38002 901-867-9971 Contact Person: Robert Churinetz Date Prepared: June 6, 1996 510(k) Summary of Safety and Efficacy Trade/Proprietary Name: Wright Plaster of Paris Pellets Common Name: Calcium Sulfate Dihydrate Product Classification: Unclassified Predicate Devices: Plaster of Paris Pellets (Ethicon) This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. Description/Intended Use Wright Plaster of Paris Pellets are provided sterile for single patient use. The biodegradable, radiopaque pellets are resorbed in 30 to 60 days when used according to labeling. Wright Plaster of Paris Pellets are 0.1g each and are supplied in pouches of 100 pellets. Wright Plaster of Paris Pellets are intended to be gently packed into non-loadbearing long bone voids. These bone voids may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The Wright Plaster of Paris Pellets provide a bone filler that resorbs and is replaced with bone during the healing process. Material Wright Plaster of Paris Pellets are made of 98% medical grade calcium dihydrate (CaSO₄ • 2H₂O) and 2% stearic acid. Testing Summary Dissolution studies demonstrate Wright Plaster of Paris Pellets to possess equivalent dissolution rates to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...