EquivaBone Osteoinductive Bone Graft Substitute (EquivaBone )

K243609 · Etex Corporation · MQV · Dec 18, 2024 · Orthopedic

Device Facts

Record IDK243609
Device NameEquivaBone Osteoinductive Bone Graft Substitute (EquivaBone )
ApplicantEtex Corporation
Product CodeMQV · Orthopedic
Decision DateDec 18, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3045
Device ClassClass 2
AttributesTherapeutic

Intended Use

EquivaBone is a bone graft substitute that combines synthetic calcium phosphate and demineralized bone. It is resorbed and replaced with new bone during the healing process. It is intended for use to fill bony voids or gaps of the skeletal system of the extremities, spine (i.e. posterolateral spine) and pelvis that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone,

Device Story

EquivaBone is a bone graft substitute composed of calcium phosphate, carboxymethylcellulose, and demineralized bone matrix (DBM). Supplied as a sterile kit containing powders and hydration components; mixed in the operating room by clinical staff to create a flowable putty. Implanted manually or via syringe into bony voids or gaps in the extremities, spine, or pelvis. Material hardens at body temperature post-implantation; resorbs and remodels into new bone during the healing process. Benefits patient by providing a scaffold for bone regeneration in non-load-bearing defects.

Clinical Evidence

No clinical data provided. Bench testing included injectability, washout resistance, compressive strength, dimensional stability, paste hardness, setting pH/temperature, mass-to-volume ratio, porosity, and user testing. Osteoinductive potential validated via in vitro C2C12 assay, demonstrating correlation with predicate's in vivo athymic rat model (R2 ≥ 0.8).

Technological Characteristics

Composition: Calcium Phosphate, Carboxymethylcellulose, Demineralized Bone Matrix. Form: Flowable putty (mixed at point of use). Hardening: Temperature-dependent setting. Sterilization: Sterile kit components.

Indications for Use

Indicated for patients requiring bone void filling in the skeletal system, specifically extremities, posterolateral spine, and pelvis, for defects not intrinsic to structural stability, including surgically created or traumatic osseous defects.

Regulatory Classification

Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ December 18, 2024 Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. ETEX Corporation Priti Dhabekar Regulatory Affairs Sr Specialist 55 Messina Drive Braintree, Massachusetts 02184 Re: K243609 Trade/Device Name: EquivaBone Osteoinductive Bone Graft Substitute (EquivaBone) Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV, MBP Dated: November 21, 2024 Received: November 22, 2024 Dear Priti Dhabekar: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, Digitally signed by JESSE JESSE MUIR -S MUIR -S MUIR-S Date: 2024.12.18 14:34:2024.12.18 14:34:20 -05'00' Jesse Muir, PhD Assistant Director DHT6C: Division of Restorative, Repair & Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Submission Number (if known) K243609 Device Name EquivaBone Osteoinductive Bone Graft Substitute (EquivaBone ) Indications for Use (Describe) EquivaBone is a bone graft substitute that combines synthetic calcium phosphate and demineralized bone. It is resorbed and replaced with new bone during the healing process. It is intended for use to fill bony voids or gaps of the skeletal system of the extremities, spine (i.e. posterolateral spine) and pelvis that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone, Type of Use (Select one or both, as applicable) > | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ | 510(k) #: K243609 | | 510(k) Summary | | Prepared on: 2024-12-13 | |------------------------------------|-----------------------------------|---------------------------------------------------------------|--|-------------------------| | Contact Details | | | | 21 CFR 807.92(a)(1) | | Applicant Name | | ETEX Corporation | | | | Applicant Address | | 55 Messina Drive Braintree MA 02184 United States | | | | Applicant Contact Telephone | | 8625919661 | | | | Applicant Contact | | Ms. Priti Dhabekar | | | | Applicant Contact Email | | priti.dhabekar@zimmerbiomet.com | | | | Device Name | | | | 21 CFR 807.92(a)(2) | | Device Trade Name | | EquivaBone Osteoinductive Bone Graft Substitute (EquivaBone ) | | | | Common Name | | Resorbable calcium salt bone void filler device | | | | Classification Name | | Filler, Bone Void, Calcium Compound | | | | Regulation Number | | 888.3045 | | | | Product Code(s) | | MQV, MBP | | | | Legally Marketed Predicate Devices | | | | 21 CFR 807.92(a)(3) | | Predicate # | | Predicate Trade Name (Primary Predicate is listed first) | | Product Code | | K101557 | | EquivaBone Osteoinductive Bone Graft Substitute | | MQV, MBP | | K082793 | InterGro DBM (Putty, Paste, Plus) | | | MQV, MBP, G | | Device Description Summary | | | | 21 CFR 807.92(a)(4) | EquivaBone is a biocompatible bone graft substitute material consisting of Calcium Phosphate (CaP3), Carboxymethylcellulose (CMC) and Demineralized Bone Matrix (DBM). EquivaBone is supplied in a single use kit as sterile powders and hydration that are mixed together at the time of use in the operating room to form flowable putty which is implanted manually or can be extruded through a syringe. After implantation the product hardens at body temperature and remodels during the healing process. # Intended Use/Indications for Use EquivaBone is a bone graft substitute that combines synthetic calcium phosphate and demineralized bone. It is resorbed and replaced with new bone during the healing process. It is intended for use to fill bony voids or gaps of the skremities, spine (i.e. posterolateral spine) and pelvis that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. # Indications for Use Comparison The indications for use of the subject device are identical to the primary predicate device. Technological Comparison 21 CFR 807.92(a)(6) 21 CFR 807.92(a)(5) 21 CFR 807.92(a)(5) <P {5}------------------------------------------------ The proposed device has similar technological characteristics as the primary predicate system. Demineralized Bone Matrix (DBM) Source: The proposed modification will transition DBM sourcing from AlloSource to LifeLink. The comparison of the DBM from both sources is based on the chemical composition, physical properties and performance characteristics of the DBM. In each comparison, we have determined that the DBM sourced from LifeLink Tissue Bank is equivalent to the DBM sourced from AlloSource. #### Osteoinductive Potential: The proposed alternative assay test method C2C12 will be employed for determining osteoinductivity. A correlation study was conducted to validate the C2C12 assay results against the current in vivo testing in place for the predicate device. This validation established a pass/fail threshold that corrent pass/fail threshold for the predicate device, that of an in vivo osteoinduction (Ol) score of 1. The in vitro C2C12 Ol assay test results correlate with the athymic rat implantation linear correlation (m = 0.0997) with correlation coefficient (R2 = 0.9204), meeting the required acceptance criteria for the method (R2 ≥ 0.8). #### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Non-Clinical Tests/Justifications: - Full Conversion - · Injectability - Washout Resistance - · Compressive Strength and Dimensional Stability - Relative Paste Hardness - Setting pH and Temperature - Mass-to-Volume Ratio - User Testing - Porosity Testing · Osteoinductive Potential- C2C12 Assay Animal and Clinical Tests: • None provided
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...