← Product Code [MQV](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV) · K233259

# Xerosyn (K233259)

_Xerothera, Inc. · MQV · Jun 21, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K233259

## Device Facts

- **Applicant:** Xerothera, Inc.
- **Product Code:** [MQV](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV.md)
- **Decision Date:** Jun 21, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

XeroSyn bone void filler is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). XeroSyn is used with autograft as a bone graft extender in the extremities and pelvis. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. XeroSyn resorbs and is replaced with bone during the healing process.

## Device Story

XeroSyn is a synthetic, bioactive, osteoconductive bone void filler; supplied as fine white powder of micro-granules (20-30 μm) with 80% porosity. Sol-gel synthesized from Ca, P, O, H, and Si; silane-treated to control granule size. Intended for use by surgeons in clinical settings to fill non-structural bony voids or gaps in extremities and pelvis; mixed with autogenous bone graft. Upon implantation, the porous structure absorbs body fluids; material is bioactive, forming a surface apatite layer (scaffolding) in vivo to facilitate new bone growth. Device resorbs over time and is replaced by host bone. Benefits include providing an osteoconductive scaffold to support bone healing in traumatic or surgical defects.

## Clinical Evidence

No clinical data. Evidence includes bench testing (analytical characterization, chemical composition, physical properties) and an animal functional study using a critical size cancellous bone defect model in New Zealand White Rabbits at 2, 4, 6, and 12 weeks. In vitro bioactivity confirmed via Kokubo test (apatite layer formation).

## Technological Characteristics

Synthetic sol-gel derived bioactive glass (Ca, P, O, H, Si). Micro-granules (20-30 μm); 80% porosity. Osteoconductive; forms surface apatite layer. Sterile (SAL 10^-6), dry heat sterilization. Single-use, non-setting powder. Complies with ISO 10993-1 for biocompatibility.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- AltaPore ([K192363](/device/K192363.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

June 21, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

XeroThera Inc % Mehdi Kazemzadeh-Narbat Director, Regulatory Affairs MCRA 803 7th Street. Floor 3 Washington, District of Columbia 20001

Re: K233259

Trade/Device Name: XeroSyn Bone Void Filler Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: May 25, 2024 Received: May 28, 2024

Dear Mehdi Kazemzadeh-Narbat:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Jesse Muir - S" in a large font. To the right of the name, there is a digital signature that reads "Digitally signed by Jesse Muir - S". The date of the signature is also included, which is "2024.06.21 12:12:11 -04'00'".

Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K233259

Device Name XeroSyn™ Bone Void Filler

#### Indications for Use (Describe)

XeroSyn bone void filler is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). XeroSyn is used with autograft as a bone graft extenties and pelvis. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. XeroSyn resorbs and is replaced with bone during the healing process.

| Type of Use (Select one or both, as applicable)       |                                               |
|-------------------------------------------------------|-----------------------------------------------|
| <b>X</b> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K233259

# 510(K) SUMMARY

| Device Trade Name: | XeroSyn Bone Void Filler                                                                                                                                                                     |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:      | XeroThera, Inc.<br>5 W Gay Street<br>West Chester, PA 19380, USA<br>Mobile: (908) 625-7734                                                                                                   |
| Contact:           | Noel Rolon, MA, CT<br>VP Regulatory and Clinical Affairs<br>Mobile: (908) 625-7734<br>Email: rolon@xerotherainc.com                                                                          |
| Prepared by:       | Mehdi Kazemzadeh-Narbat, PhD, PMP, CQA<br>Director, Regulatory Affairs,<br>MCRA, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20001<br>Office: 202.552.6011<br>mkazemzadeh@mcra.com |
| Date Prepared:     | June 18, 2024                                                                                                                                                                                |
| Classifications:   | 21 CFR §880.3045, Resorbable calcium salt bone void filler device                                                                                                                            |
| Class:             | II                                                                                                                                                                                           |
| Product Codes:     | MQV                                                                                                                                                                                          |
| Primary Predicate: | AltaPore (K192363)                                                                                                                                                                           |

#### Indications For Use:

XeroSyn bone void filler is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities and pelvis). XeroSyn is used with autograft as a bone graft extender in the extremities and pelvis. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. XeroSyn resorbs and is replaced with bone during the healing process.

#### Device Description:

XeroSyn bone void filler is a synthetic bone graft substitute designed for use as an autograft extender in the extremities and the pelvis. XeroSyn bone void filler is a sol-gel synthesized

{4}------------------------------------------------

bioactive, osteoconductive bone void filler. The chemical composition of XeroSyn is very similar to the predicate, Altapore, containing elements that exist naturally in bone (Ca, P, O, H, Si), albeit at different ratios. The sol-gel synthesis process produces porous microparticles which are silane treated to enhance the particle growth to form the desired granule size. The interconnected highly porous structure of XeroSyn readily absorbs body fluids and facilitates quick and easy mixing with autogenous bone.

XeroSyn bone void filler is supplied in a sterile glass vial within a sterile pouch and contains 1 gram of XeroSyn micro-granules sized 20-30 um, 80% total porosity, and appears as a fine white powder. XeroSyn does not set in situ following implantation.

XeroSyn is bioactive as supported by the Kokubo test. This in vitro test shows the formation of a surface apatite layer when the synthetic graft material is immersed in a standard simulated body fluid with the ion concentration of human blood plasma. This apatite layer provides scaffolding onto which the patient's new bone will grow.

### Predicate Device:

XeroThera Inc. submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, XeroSyn is substantially equivalent in intended use, indications, design principles, and performance to the following predicate device, which has been determined by FDA to be substantially equivalent to pre-amendment devices:

Predicate: AltaPore (K192363)

## Performance Testing Summary:

Bench testing including analytical characterization, chemical composition, physical properties, and animal functional studies were conducted for the subject XeroSyn and predicate device. The results of the studies demonstrate substantial equivalence to the predicate device.

#### Substantial Equivalence:

Each of the following test results met the acceptance criteria and support the substantial equivalence of the proposed device as compared to the predicate for its intended use.

#### Bench Testing

XeroSyn is bioactive based on in vitro studies with the Kokubo test that show it forms a surface apatite layer when submerged in simulated body fluid that contains the same ion concentrations as human blood plasma. The SEM and FTIR spectral analyses reveal the formation of crystalline carbonated apatite (c-AP) on the surface of XeroSyn after 2 days of immersion. This apatite layer provides scaffolding onto which the patient's new bone will grow. Similarly, the predicate device forms an apatite layer on its surface after immersion in simulated body fluids. These results demonstrate similar bioactivity for XeroSyn when compared to the predicate device.

Bench testing including chemical composition, and physical properties, were conducted for the subject XeroSyn device. The results of the studies demonstrate substantial equivalence to the predicate device.

#### Biocompatibility Testing

{5}------------------------------------------------

XeroSyn has been assessed in accordance with the requirements of ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The full complement of testing included the following battery of tests which all confirmed the biocompatibility of the device:

- Cytotoxicity .
- . Sensitization
- Irritation
- . Acute Systemic Toxicity
- Subacute/Subchronic Toxicity
- . Material-Mediated Pyrogenicity
- . Genotoxicity

## Animal Testing

The in vivo response to XeroSyn and the AltaPore predicate device were evaluated and compared at 2, 4, 6 and 12 weeks after implantation using a standard and reproducible critical size cancellous bone defect model in skeletally mature female New Zealand (NZ) White Rabbits. The results of this study were used to determine safety and effectiveness for use in extremities when mixed with autogenous iliac crest bone graft.

The subject device and the predicate device have the same intended uses, the same product classification and product code (MQV) and have similar "Indications for Use" statements in the extremities and pelvis. The subject device and the predicate devices are bone void fillers and bone graft extenders that are intended for bony voids or gaps that are not intrinsic to the stability of the bony structure. The subject device and the primary predicate incorporate the same basic design, similar materials, and are both provided sterile for single-patient and single-use.

No clinical data were required to support the substantial equivalence.

| Attribute                        | XeroSyn<br>(Subject Device)                                                                                                                                                                                                                                                                          | AltaPore<br>(K192363)                                                                                                                                                                                                                                                                                   |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                     | XeroSyn bone void filler is an<br>implant intended to fill bony voids<br>or gaps of the skeletal system (i.e.,<br>extremities and pelvis).                                                                                                                                                           | AltaPore is an implant intended to<br>fill bony voids or gaps of the<br>skeletal system (i.e., extremities,<br>posterolateral spine and pelvis).                                                                                                                                                        |
| Indications for Use<br>Statement | XeroSyn bone void filler is an<br>implant intended to fill bony voids<br>or gaps of the skeletal system (i.e.,<br>extremities and pelvis). XeroSyn<br>is used with autograft as a bone<br>graft extender in the extremities<br>and pelvis. These osseous defects<br>may be surgically created or the | AltaPore is an implant intended to<br>fill bony voids or gaps of the<br>skeletal system (i.e., extremities,<br>posterolateral spine and pelvis).<br>AltaPore can be used by itself, with<br>autograft as a bone graft extender or<br>with autogenous bone marrow<br>aspirate. These osseous defects may |

Table 1. Substantial Equivalence of the Subject and Predicate Device.

{6}------------------------------------------------

|                         | result of traumatic injury to the<br>bone and are not intrinsic to the<br>stability of the bony structure.<br>XeroSyn resorbs and is replaced<br>with bone during the healing<br>process. | be surgically created or the result of<br>traumatic injury to the bone and are<br>not intrinsic to the stability of the<br>bony structure. AltaPore resorbs and<br>is replaced with bone during the<br>healing process. |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Chemical<br>Composition | Silicon dioxide glass containing<br>calcium oxide and phosphorous<br>pentoxide composed solely of<br>elements that exist naturally in<br>normal bone (Ca, P, O, H, Si).                   | Calcium Phosphate Salt:<br>Silicate-substituted calcium<br>phosphate composed solely of<br>elements that exist naturally in<br>normal bone (Ca, P, O, H, Si).                                                           |
| Design                  | microgranules,<br>sized 20-30 μm in fine powder<br>form                                                                                                                                   | microgranules,<br>sized 1-2 mm                                                                                                                                                                                          |
| Porosity                | $80 \pm 2%$                                                                                                                                                                               | $82.5 \pm 2.5%$                                                                                                                                                                                                         |
| Osteoconductivity       | Osteoconductive                                                                                                                                                                           | Osteoconductive                                                                                                                                                                                                         |
| Bioactivity             | Bioactive, forms surface apatite<br>layer after implantation                                                                                                                              | Bioactive, forms surface apatite<br>layer after implantation                                                                                                                                                            |
| Resorption              | Device degrades and resorbs over<br>time                                                                                                                                                  | Device degrades and resorbs over<br>time                                                                                                                                                                                |
| Biocompatibility        | Established, ISO 10993                                                                                                                                                                    | Established, ISO 10993                                                                                                                                                                                                  |
| Sterility               | SAL of 10-6, Dry Heat                                                                                                                                                                     | SAL of 10-6, Irradiation                                                                                                                                                                                                |
| Packaging               | Sterile, single use                                                                                                                                                                       | Sterile, single use                                                                                                                                                                                                     |
| Duration of<br>Implant  | About 6 weeks                                                                                                                                                                             | More than 12 weeks                                                                                                                                                                                                      |

## Conclusion:

The subject device and the predicate devices have intended use, have similar technological characteristics, and are made of similar materials. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above. XeroSyn bone void filler is as safe, as effective, and performs as well as, or better, than the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K233259](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K233259)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
