← Product Code [MQV](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV) · K231968

# Tactoset® Injectable Bone Substitute (K231968)

_Anika Therapeutics, Inc. · MQV · Dec 20, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K231968

## Device Facts

- **Applicant:** Anika Therapeutics, Inc.
- **Product Code:** [MQV](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV.md)
- **Decision Date:** Dec 20, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Tactoset® Injectable Bone Substitute is a synthetic, biocompatible material that hardens and converts to a poorly crystalline hydroxyapatite at body temperature. It is indicated for filling bone voids or defects of the skeletal system (i.e. extremities and pelvis) that are not intrinsic to the stability of bony structure. These defects may be surgicallycreated osseous defects or defects created from traumatic injury to the bone. The device provides an injectable, selfsetting, osteoconductive bone graft substitute that resorbs and is replaced by the growth of new bone during the healing process. Tactoset® Injectable Bone Substitute can augment hardware and support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process.

## Device Story

Tactoset is an injectable, settable, osteoconductive calcium phosphate bone graft substitute. Provided as two components (dry powder and aqueous solution) in pre-loaded syringes; mixed intra-operatively using a supplied system to form a cohesive paste. Used in clinical settings by surgeons to fill skeletal bone voids or defects (extremities and pelvis) and augment hardware. The paste hardens in situ at body temperature, converting to poorly crystalline hydroxyapatite; resorbs over time and is replaced by new bone. Acts as temporary support media; not intended for structural load-bearing. Benefits include osteoconductivity and radiopacity (via Iohexol) to assist surgical visualization.

## Clinical Evidence

No clinical data. Bench testing included analytical characterization, chemical composition, physical properties, and mechanical testing (pull-out, compressive strength, bone alignment). Animal functional study (rabbit critical-sized femoral defect model) compared Tactoset with Iohexol to predicate (without Iohexol). At 26 weeks, new bone formation was ~17% for both groups; implant material remaining was ~61% (subject) vs ~59% (predicate).

## Technological Characteristics

Injectable calcium phosphate bone graft substitute. Powder: alpha-tricalcium phosphate, calcium carbonate, monocalcium phosphate. Liquid: sodium phosphate dibasic, citric acid, hyaluronic acid, Iohexol, water. Setting principle: chemical hardening at body temperature to poorly crystalline hydroxyapatite. Sterile, single-use. Powder sterilized via gamma irradiation; liquid via autoclave and VHP.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- Tactoset ([K212083](/device/K212083.md))
- Tactoset ([K223915](/device/K223915.md))
- Cerament Bone Void Filler ([K201535](/device/K201535.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 20, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Anika Therapeutics, Inc. % Mehdi Kazemzadeh-Narbat, PhD, PMP, CQA Associate Director, Regulatory Affairs MCRA, LLC 1050 K Street NW. Suite 1000 Washington, District of Columbia 20001

Re: K231968

Trade/Device Name: Tactoset® Injectable Bone Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: December 1, 2023 Received: December 1, 2023

Dear Dr. Kazemzadeh-Narbat:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Sara S. Thompson -S

For

Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices

{2}------------------------------------------------

Enclosure

Office of Product Evaluation and Quality Center for Devices and Radiological Health

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K231968

Device Name Tactoset® Injectable Bone Substitute

#### Indications for Use (Describe)

Tactoset® Injectable Bone Substitute is a synthetic, biocompatible material that hardens and converts to a poorly crystalline hydroxyapatite at body temperature. It is indicated for filling bone voids or defects of the skeletal system (i.e. extremities and pelvis) that are not intrinsic to the stability of bony structure. These defects may be surgicallycreated osseous defects or defects created from traumatic injury to the bone. The device provides an injectable, selfsetting, osteoconductive bone graft substitute that resorbs and is replaced by the growth of new bone during the healing process.

Tactoset® Injectable Bone Substitute can augment hardware and support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process.

Type of Use (Select one or both, as applicable)

| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)    </span> </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </span>                 |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) SUMMARY

| Device Trade Name:     | Tactoset                                                                                                                                                                                               |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:          | Anika Therapeutics, Inc.<br>32 Wiggins Avenue<br>Bedford, MA 01730                                                                                                                                     |
| Contact:               | Wei Zhao<br>Executive Director, Regulatory Affairs<br>Mobile: (978)888-5948<br>E-Mail: wzhao@anika.com                                                                                                 |
| Prepared by:           | Mehdi Kazemzadeh-Narbat, PhD, PMP, CQA<br>Associate Director, Regulatory Affairs,<br>MCRA, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20001<br>Office: 202.552.6011<br>mkazemzadeh@mcra.com |
| Date Prepared:         | December 19, 2023                                                                                                                                                                                      |
| Classifications:       | 21 CFR §880.3045, Resorbable calcium salt bone void filler device                                                                                                                                      |
| Class:                 | II                                                                                                                                                                                                     |
| Product Codes:         | MQV                                                                                                                                                                                                    |
| Primary Predicate:     | Tactoset (K212083)                                                                                                                                                                                     |
| Additional Predicates: | Tactoset (K223915)<br>Cerament Bone Void Filler (K201535)                                                                                                                                              |

#### Indications For Use:

Tactoset® Injectable Bone Substitute is a synthetic, biocompatible bone graft substitute material that hardens and converts to a poorly crystalline hydroxyapatite at body temperature. It is indicated for filling bone voids or defects of the skeletal system (i.e. extremities and pelvis) that are not intrinsic to the stability of bony structure. These defects may be surgically-created osseous defects or defects created from traumatic injury to the bone. The device provides an injectable, self-setting, osteoconductive bone graft substitute that resorbs and is replaced by the growth of new bone during the healing process. Tactoset Injectable Bone Substitute can augment hardware and support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process.

{5}------------------------------------------------

## Device Description:

Tactoset® Injectable Bone Substitute is an injectable, settable osteoconductive calcium phosphate bone graft substitute material. It is provided to the end-user as two components (a dry powder and an aqueous solution in separate pre-loaded syringes) that must be mixed intra-operatively prior to implantation using the supplied mixing system to form a cohesive paste. The dry powder component is composed of the alpha phase of tricalcium phosphate [Ca3(PO4)2], calcium carbonate [CaCO3], and monocalcium phosphate [Ca(H2PO4)2]. The liquid component is composed of sodium phosphate dibasic [Na2HPO4], citric acid [C6H8O7], hyaluronic acid (HA), Iohexol (C19H26J3N309) and water for injection. Tactoset is provided sterile for single use.

#### Predicate Device:

Anika Therapeutics, Inc. submits the following in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, Tactoset is substantially equivalent in indications, design principles, and performance to the following predicate devices, which have been determined by FDA to be substantially equivalent to preamendment devices:

| Primary Predicate:     | Tactoset (K212083)                                        |
|------------------------|-----------------------------------------------------------|
| Additional Predicates: | Tactoset (K223915)<br>Cerament Bone Void Filler (K201535) |

## Performance Testing Summary:

Bench testing data including analytical characterization, chemical composition, physical properties, mechanical testing (Pull-Out. Compressive Strength, Bone Alignment) and animal functional study were conducted for the subject Tactoset and the predicate devices. The results of the study demonstrate substantial equivalence to the predicate devices. The performance of Tactoset containing iohexol radiopacifier was compared to the predicate device, Tactoset (without iohexol), in a rabbit critical sized femoral defect model. At the 26-week timepoint, animal study data demonstrated an average new bone formation of approximately 17% in the Tactoset with iohexol group, and 17% in the Tactoset (predicate) group. Animal study data at 26 weeks demonstrated that approximately 61% of implant material remained in the Tactoset with iohexol group, and 59% remained in the Tactoset (predicate) group, with respect to the geometry of the defect filled with implant material at the time of surgery. At 12 weeks the empty defect contained approximately 6% bone. Clinical performance has not been evaluated.

## Substantial Equivalence:

The Subject Device is substantially equivalent to the primary predicate Tactoset (K212083) in terms of material composition, manufacturing process, sterilization, and packaging and both are manufactured in the same facility. The Subject Device and the predicate devices (primary and additional) have the same intended uses, the same product classification and product code (MOV) and have similar "Indications for Use" statements in the extremities and pelvis. The Subject Device and the predicate devices are bone void fillers that are intended for bony voids or gaps that are not intrinsic to the stability of the bony structure. The Subject Device and the primary predicate incorporate the same basic design, incorporates similar materials, have identical manufacturing process, identical packaging and are both provided sterile for singlepatient, single-use in the same device volume.

{6}------------------------------------------------

The Subject Device and additional predicate Cerament Bone Void Filler (K201535) uses an identical contrast agent/radiopacifier (Iohexol) in its formulation.

No clinical data were required to support the substantial equivalence.

| Information                      | Subject Device<br>(Tactoset)                                                                                                                                                                                                                     | Primary<br>Predicate<br>(Tactoset)                                                                                                                                                                                                               | Additional<br>Predicate<br>(Tactoset)                                                                                                                                                                                                            | Additional<br>Predicate<br>(Cerament)                                                  |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Manufacturer                     | Anika Therapeutics,<br>Inc.                                                                                                                                                                                                                      | Anika Therapeutics,<br>Inc.                                                                                                                                                                                                                      | Anika Therapeutics,<br>Inc.                                                                                                                                                                                                                      | Bone Support AB                                                                        |
| Trade Name                       | Tactoset                                                                                                                                                                                                                                         | Tactoset                                                                                                                                                                                                                                         | Tactoset                                                                                                                                                                                                                                         | Cerament Bone Void<br>Filler                                                           |
| 510(k)<br>Number                 | TBD                                                                                                                                                                                                                                              | K212083                                                                                                                                                                                                                                          | K223915                                                                                                                                                                                                                                          | K201535                                                                                |
| Product Code                     | MQV                                                                                                                                                                                                                                              | MQV                                                                                                                                                                                                                                              | MQV                                                                                                                                                                                                                                              | MQV                                                                                    |
| Device<br>Classification<br>Name | Resorbable calcium<br>salt bone void filler<br>device                                                                                                                                                                                            | Resorbable calcium<br>salt bone void filler<br>device                                                                                                                                                                                            | Resorbable calcium<br>salt bone void filler<br>device                                                                                                                                                                                            | Resorbable calcium<br>salt bone void filler<br>device                                  |
| Physical Form                    | Injectable<br>Paste/Putty                                                                                                                                                                                                                        | Injectable Paste/Putty                                                                                                                                                                                                                           | Injectable Paste/Putty                                                                                                                                                                                                                           | Injectable Paste                                                                       |
| Device<br>Volume (CC)            | 4                                                                                                                                                                                                                                                | 4                                                                                                                                                                                                                                                | 4                                                                                                                                                                                                                                                | 5, 10, 18                                                                              |
| Mixing Time<br>(Min)             | 1                                                                                                                                                                                                                                                | 1                                                                                                                                                                                                                                                | 1                                                                                                                                                                                                                                                | 0.5                                                                                    |
| Working<br>Time (Minute)         | 7-18                                                                                                                                                                                                                                             | 7-18                                                                                                                                                                                                                                             | 7-18                                                                                                                                                                                                                                             | 3 to 5                                                                                 |
| Setting Time<br>(Minute)         | 10                                                                                                                                                                                                                                               | 10                                                                                                                                                                                                                                               | 10                                                                                                                                                                                                                                               | Not available                                                                          |
| Sterility<br>Method              | Single Use, Sterile,<br>SAL 10-6<br>Setting Solution:<br>Autoclave and VHP<br>Powder Component:<br>Gamma Irradiation                                                                                                                             | Single Use, Sterile,<br>SAL 10-6<br>Setting Solution:<br>Autoclave and VHP<br>Powder Component:<br>Gamma Irradiation                                                                                                                             | Single Use, Sterile,<br>SAL 10-6<br>Setting Solution:<br>Autoclave and VHP<br>Powder Component:<br>Gamma Irradiation                                                                                                                             | Single Use, Sterile,<br>SAL 10-6<br>Gamma irradiation;<br>Steam; and Ethylene<br>oxide |
| Pyrogenicity                     | LAL < 20 EU/device                                                                                                                                                                                                                               | LAL < 20 EU/device                                                                                                                                                                                                                               | LAL < 20 EU/device                                                                                                                                                                                                                               | LAL < 20 EU/device                                                                     |
| Packaging                        | The liquid syringe is<br>double packaged in<br>heat-sealed Tyvek<br>pouches. The cement<br>powder is provided<br>in a mixing syringe<br>packaged in a foil<br>pouch. The foil<br>pouch is then<br>packaged in a plastic<br>tray with a Tyvek lid | The liquid syringe is<br>double packaged in<br>heat-sealed Tyvek<br>pouches. The cement<br>powder is provided in<br>a mixing syringe<br>packaged in a foil<br>pouch. The foil<br>pouch is then<br>packaged in a plastic<br>tray with a Tyvek lid | The liquid syringe is<br>double packaged in<br>heat-sealed Tyvek<br>pouches. The cement<br>powder is provided in<br>a mixing syringe<br>packaged in a foil<br>pouch. The foil<br>pouch is then<br>packaged in a plastic<br>tray with a Tyvek lid | Blister tray packed in<br>a Tyvek-PET/PE<br>pouch                                      |

Table 6-1. Substantial Equivalence of the Subject and Predicate Devices.

{7}------------------------------------------------

## Conclusion:

The Subject Device and the predicate devices have intended use, have similar technological characteristics, and are made of similar materials. The subject and predicate devices are packaged in identical materials and are sterilized using identical methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above. Tactoset is as safe, as effective, and performs as well as, or better, than the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K231968](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K231968)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
