K221256 · Bone Solutions, Inc. · MQV · Jun 28, 2022 · Orthopedic
Device Facts
Record ID
K221256
Device Name
Mg OSTEOINJECT
Applicant
Bone Solutions, Inc.
Product Code
MQV · Orthopedic
Decision Date
Jun 28, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Mg OSTEOINJECT™ is intended for bony voids or defects of the extremities and pelvis that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created by traumatic injury to the bone. Mg OSTEOINJECT™ can be used as an adjunct to conventional rigid hardware fixation by supporting the bone fragments during the surgical procedure only in the extremities and pelvis. Once the material has set, it acts as a temporary support medium and is not intended to provide structural support during the healing process. Mg OSTEOINJECT™ is intended to be placed into bony voids either before or after final fixation. Mg OSTEOINJECT™ is resorbed and replaced with bone during the healing process. Mg OSTEOINJECT™ is not intended to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
Device Story
Mg OSTEOINJECT™ is a magnesium-based synthetic bone void filler; supplied as a kit containing powder and buffered saline mixing solution. Intra-operative mixing initiates an acid-base reaction, creating a cohesive, drillable, radiopaque, and osteoconductive paste. Applied by surgeons in clinical settings to bony voids in extremities and pelvis; adheres to adjacent bone during curing. Acts as a temporary support medium during surgical procedures; resorbed and replaced by bone during healing. Not intended for structural support during the healing process. Benefits patients by filling voids and supporting bone fragments as an adjunct to rigid fixation.
Clinical Evidence
No clinical data included. Substantial equivalence supported by non-clinical bench testing referenced from K212991, including chemical composition, physical properties, sterilization, shelf life, biocompatibility, and bacterial endotoxin testing (USP <85>).
Technological Characteristics
Magnesium-based synthetic bone void filler. Composition: β-tricalcium phosphate (8%), magnesium oxide (41%), monopotassium phosphate (44%), monosodium phosphate (3%). Form: Powder and liquid kit (3 cc and 5 cc). Principle: Acid-base reaction forms cohesive, drillable, radiopaque, osteoconductive paste. Sterilization: Gamma irradiation. Single-use.
Indications for Use
Indicated for bony voids or defects of the extremities and pelvis not intrinsic to structural stability, including surgically created or traumatic osseous defects. Used as an adjunct to rigid hardware fixation. Not for large defects unlikely to heal spontaneously.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{1}------------------------------------------------
June 28, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Bone Solutions, Inc. % Kevin A. Thomas, Ph.D. Vice President & Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K221256
Trade/Device Name: Mg OSTEOINJECT™ Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV, OIS Dated: May 2, 2022 Received: May 2, 2022
Dear Dr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name
Mg OSTEOINJECTTM
#### Indications for Use (Describe)
Mg OSTEOINJECT™ is intended for bony voids or defects of the extremities and pelvis that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects created by traumatic injury to the bone. Mg OSTEOINJECT™ can be used as an adjunct to conventional rigid hardware fixation by supporting the bone fragments during the surgical procedure only in the extremities and pelvis. Once the material has set, it acts as a temporary support medium and is not intended to provide structural support during process. Mg OSTEOINJECT™ is intended to be placed into bony voids either before or after final fixation. Mg OSTEOINJECT™ is resorbed and replaced with bone during the healing process. Mg OSTEOINJECT™ is not intended to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K221256 Mg OSTEOINJECT™ Bone Solutions, Inc.
June 28, 2022
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Bone Solutions, Inc.<br>5712 Colleyville Boulevard, Suite 210<br>Colleyville, Texas 76034<br>Telephone +1 817-809-8850 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Drew Diaz, CEO |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 |
| San Diego, CA 92130 | |
|---------------------|--------------------|
| Telephone | +1 858-792-1235 |
| Fax | +1 858-792-1236 |
| Email | kthomas@paxmed.com |
| | flarson@paxmed.com |
## DEVICE NAME AND CLASSIFICATION
| Trade/Device Name | Mg OSTEOINJECT™ |
|------------------------|-------------------------------------------------------------------------------|
| Common Name | Filler, bone void, calcium compound |
| Regulation Number | 21 CFR 888.3045 |
| Regulation Name | Resorbable calcium salt bone void filler device |
| Regulatory Class | Class II |
| Product Code | MQV |
| Secondary Product Code | OIS |
| Classification Panel | Orthopedic |
| Reviewing Office | Office of Health Technology 6 (Orthopedic Devices) |
| Reviewing Division | Division of Health Technology 6 C<br>(Restorative, Repair and Trauma Devices) |
## PREDICATE DEVICE INFORMATION
Primary Predicate Device K212991, OSTEOREVIVE™, Bone Solutions, Inc.
Reference Device K140375, MASTERGRAFT® Strip; MASTERGRAFT® Putty, Medtronic Sofamor Danek USA, Inc.
{5}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
Mg OSTEOINJECT™ is intended for bony voids or defects of the extremities and pelvis that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created by traumatic injury to the bone. Mg OSTEOINJECT™ can be used as an adjunct to conventional rigid hardware fixation by supporting the bone fragments during the surgical procedure only in the extremities and pelvis. Once the material has set. it acts as a temporary support medium and is not intended to provide structural support during the healing process. Mg OSTEOINJECT™ is intended to be placed into bony voids either before or after final fixation. Mg OSTEOINJECT™ is resorbed and replaced with bone during the healing process. Mg OSTEOINJECT™ is not intended to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
#### SUBJECT DEVICE DESCRIPTION
Mg OSTEOINJECT™ is a magnesium-based synthetic bone void filler that is drillable, resorbable, radiopaque, and osteoconductive. The Mg OSTEOINJECT™ Kit contains powder (magnesium-based compound) and a mixing solution (buffered saline). Once the components are mixed intra-operatively prior to implantation, an acid-base reaction happens to form a cohesive paste. Once the product is placed into the bony void, the paste will adhere to the adjacent bone during the curing process. The device is provided sterile to the end user for single-use only in two sizes, 3 cc and 5 cc.
#### PERFORMANCE DATA
Non-clinical testing data according to the guidance documents Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device (issued June 2003) and Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile (issued January 2016) were referenced from K212991. The non-clinical testing data referenced to demonstrate substantial equivalence included: chemical composition, physical properties, sterilization, sterile barrier shelf life, product shelf life, and biocompatibility. Performance testing data also referenced from K212991 demonstrated that the subject device is drillable, and may be used as an adjunct to conventional rigid hardware during the surgical procedure (only when used in the extremities and pelvis).
Bacterial endotoxin testing has been performed to ensure the device meets pyrogen limit specifications. The Limulus amebocyte lysate (LAL) test, kinetic turbidimetric method, was performed according to USP <85> Bacterial Endotoxins Test. The LAL testing met the limit acceptance criterion of ≤ 20 EU/device, based upon the recommendations for implanted devices in the FDA guidance document Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile, issued January 21, 2016 (Section V, A, 4).
No clinical data were included in this submission.
{6}------------------------------------------------
#### EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in indications and design principles to the primary predicate device K212991 and the reference device K140375. Provided at the end of this summary is a table comparing the Indications for Use Statements and the technological characteristics of the subject device and the primary predicate device.
The primary predicate device is K212991, for the identical formulation, packaging, and sterilization as the subject device. The subject device differs from K212991 in terms of the indications for use; the subject device is indicated only for use in the extremities and pelvis.
The reference device K140375 is for support of substantial equivalence of the subject device provided in a kit with a volume of 3 cc. The reference device K140375 includes the same indications for use in the extremities and pelvis and includes a range of volumes from 0.75 cc to 9 cc.
The subject device and the primary predicate device K21299 have the same intended use, the same product classification and product codes (MQV and OIS), and have similar Indications for Use statements. The subject device, the primary predicate device, and the reference device are indicated for use as bone void fillers in the extremities and pelvis. Although the subject device, the primary predicate device, and the reference device have slightly different Indications for Use language, these differences in language do not change the intended use as a bone void filler.
The subject device, the primary predicate device, and the reference device all incorporate calcium phosphate materials that are mixed to form a solid after setting (subject device and primary predicate device) or are within a polymeric binder or scaffold (reference device). The subject device and primary predicate device have identical materials and formulation and are mixed intra-operatively. The subject device, the primary predicate device, and the reference device are all provided in comparable volumes, sterilized by irradiation, and are for single-patient, single-use.
Differences among the subject device, the primary predicate device, and the reference device include the exact indications for use (including use in the posterolateral spine for the primary predicate and reference device), the mineral components, and the scaffold or binder material (bovine collagen in the reference device K 140375). The minor differences among the subject device, the primary predicate device, and the reference device do not raise new issues of safety or effectiveness, and therefore, do not impact substantial equivalence.
#### CONCLUSION
The subject device, the primary predicate device, and the reference device have the same intended use, have similar technological characteristics, and are made of similar materials. The subject device and the primary predicate device are manufactured from identical materials. The subject device, the primary predicate, and the reference device encompass the same range of physical dimensions (volumes), are packaged in similar materials and are sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate device and the reference device listed above.
{7}------------------------------------------------
## Table of Substantial Equivalence
| Features / Comparisons | Subject Device | Primary Predicate Device |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Mg OSTEOINJECTTM<br>Bone Solutions Inc. | K212991<br>OSTEOREVIVETM<br>Bone Solutions Inc. |
| Indications for Use Statement | Mg OSTEOINJECTTM is intended for bony voids<br>or defects of the extremities and pelvis that are not<br>intrinsic to the stability of the bony structure.<br>These defects may be surgically created osseous<br>defects or osseous defects created by traumatic<br>injury to the bone.<br>Mg OSTEOINJECTTM can be used as an adjunct to<br>conventional rigid hardware fixation by supporting<br>the bone fragments during the surgical procedure.<br>Once the material has set, it acts as a temporary<br>support medium and is not intended to provide<br>structural support during the healing process.<br>Mg OSTEOINJECTTM is intended to be placed into<br>bony voids either before or after final fixation.<br>Mg OSTEOINJECTTM is resorbed and replaced<br>with bone during the healing process. Mg<br>OSTEOINJECTTM is not intended to treat large<br>defects that in the surgeon's opinion would fail to<br>heal spontaneously | OSTEOREVIVETM is intended for bony voids or<br>defects of the extremities, posterolateral spine, and<br>pelvis that are not intrinsic to the stability of the<br>bony structure.<br>These defects may be surgically created osseous<br>defects or osseous defects created by traumatic<br>injury to the bone.<br>OSTEOREVIVETM can be used as an adjunct to<br>conventional rigid hardware fixation by supporting<br>the bone fragments during the surgical procedure.<br>Once the material has set, it acts as a temporary<br>support medium and is not intended to provide<br>structural support during the healing process.<br>OSTEOREVIVETM is intended to be placed into<br>bony voids either before or after final fixation.<br>OSTEOREVIVETM is resorbed and replaced with<br>bone during the healing process.<br>OSTEOREVIVETM must be used with morselized<br>autograft bone at a ratio of 1:1 by volume in the<br>posterolateral spine.<br>OSTEOREVIVETM is not intended to treat large<br>defects that in the surgeon's opinion would fail to<br>heal spontaneously. |
| Reason for Predicate/Reference<br>Device | Not applicable - Subject Device | Identical or similar indications for use, formulation,<br>packaging, and sterilization |
| Product Codes | MQV, OIS | MQV, OIS |
| Intended Use | Bone void filler for skeletal system;<br>extremities and pelvis | Bone void filler for skeletal system;<br>extremities, pelvis, and posterolateral spine |
| Design | | |
| Form | Powder and liquid components;<br>after mixing device is injectable and sets in the<br>surgical site | Powder and liquid components;<br>after mixing device is injectable and sets in the<br>surgical site |
| Granule Size | Not applicable, non-porous solid after mixing and<br>setting | Not applicable, non-porous solid after mixing and<br>setting |
| Porosity | Not applicable, non-porous solid after mixing and<br>setting | Not applicable, non-porous solid after mixing and<br>setting |
| Materials | | |
| Mineral component<br>Calcium/other salts | β-tricalcium phosphate (8%)<br>Magnesium oxide (41%)<br>Monopotassium phosphate (44%)<br>Monosodium phosphate (3%) | β-tricalcium phosphate (8%)<br>Magnesium oxide (41%)<br>Monopotassium phosphate (44%)<br>Monosodium phosphate (3%) |
| Scaffold/Binder | None | None |
| Indicted for Use in<br>Extremities and Pelvis | Yes | Yes |
| Mix with autograft bone prior<br>to use | No<br>Not indicated for mixing with autograft in<br>extremities or pelvis | No<br>Not indicated for mixing with autograft in<br>extremities or pelvis |
| Indicated for Use in<br>Posterolateral Spine | No | Yes |
| Mix with autograft bone prior<br>to use | Not applicable – not indicated for use in<br>posterolateral spine | Required to mix with autograft 1:1 by volume for<br>use in the posterolateral spine |
| Features / Comparisons | Subject Device | Primary Predicate Device |
| | Mg OSTEOINJECTTM<br>Bone Solutions Inc. | K212991<br>OSTEOREVIVETM<br>Bone Solutions Inc. |
| How Provided | | |
| Sizes | Kits of powder and liquid components for volumes of:<br>3 cc and 5 cc | Kits of powder and liquid components for volumes of:<br>5 cc, 10 cc, and 15 cc |
| Sterility | Provided sterile to end user | Provided sterile to end user |
| Sterilization | Gamma irradiation | Gamma irradiation |
| Usage | Single-patient, single-use | Single-patient, single-use |
{8}------------------------------------------------
## Table of Substantial Equivalence
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.