← Product Code [MQV](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV) · K202639

# si-Mochi (K202639)

_Biostone, Ltd. · MQV · Apr 21, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K202639

## Device Facts

- **Applicant:** Biostone, Ltd.
- **Product Code:** [MQV](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV.md)
- **Decision Date:** Apr 21, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

si-Mochi is an implant intended to fill bony voids or gaps of the skeletal system (i.e. extremities, pelvis). These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. si-Mochi resorbs and is replaced with bone during the healing process.

## Device Story

si-Mochi is a bone void filler consisting of multi-porous bi-phase calcium phosphate ceramic granules (80% hydroxylapatite, 20% beta-tricalcium phosphate) with trace silicate, suspended in an aqueous Kolliphor polymer carrier gel. The granules feature interconnected macropores (100µm–1mm), midipores (10–100µm), and microspaces (1–10µm). Supplied in sterile applicators, the device is used by surgeons in an operating theater to fill osseous defects. Upon implantation, the porous structure acts as a scaffold, encouraging host bone formation and capillary ingrowth; the material undergoes physiologically-mediated resorption and is replaced by natural bone. The 33% Kolliphor gel provides necessary viscosity for handling at room temperature. The device does not set in-situ and contains no antibiotics.

## Clinical Evidence

No human clinical data provided. Evidence consists of non-clinical bench testing (biocompatibility, pyrogenicity, endotoxin, shelf-life, packaging) and an animal study. The animal study used a rabbit critical size defect in the distal femora to compare si-Mochi against Actifuse ABX. Results indicated no statistically significant difference in the percentage of new bone formation between the subject and predicate devices at all post-implantation time points.

## Technological Characteristics

Materials: Bi-phase calcium phosphate ceramic (80% hydroxylapatite, 20% beta-tricalcium phosphate) with trace silicate; Kolliphor polymer carrier gel. Structure: Multi-porous granules (1-2mm) with interconnected macropores, midipores, and microspaces. Form factor: Granules in aqueous gel, supplied in sterile applicators. Sterilization: Sterile. Energy source: None (passive implant).

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- TriPore® ([K070132](/device/K070132.md))
- Actifuse ABX™ ([K082575](/device/K082575.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

April 21, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Biostone Limited Wei-Jen Lo, Ph.D. Chief Scientific Officer BioCity, Pennyfoot Street Nottingham, Notts NG1 1GF United Kingdom

## Re: K202639

Trade/Device Name: si-Mochi Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: March 16, 2021 Received: March 29, 2021

Dear Dr. Lo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

#### 510(k) Number (if known) K202639

Device Name

si-Mochi

Indications for Use (Describe)

si-Mochi is an implant intended to fill bony voids or gaps of the skeletal system (i.e.extremities, pelvis). These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. si-Mochi resorbs and is replaced with bone during the healing process.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| Submitter<br>Telephone<br>Facsimile<br>Contact Person              | Biostone Limited<br>BioCity<br>Pennyfoot Street<br>Nottingham<br>NG1 1GF<br>United Kingdom<br>011 44 7906 001281<br>Wei-Jen Lo PhD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|--------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared<br>Trade Name                                        | 21/ July/ 2019<br>si-Mochi                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Common Name                                                        | Resorbable calcium salt bone void filler device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Classification                                                     | Resorbable calcium salt bone void filler devices have<br>been classified by the Orthopedics Device Panel as<br>Class II Special Controls per 21 CFR 888.3045.<br>Product code: MQV                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Predicate Devices                                                  | TriPore® K070132, Actifuse ABX™ K082575                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device Description                                                 | si-Mochi contains a multi porous bi-phase Calaium<br>Phosphate ceramic granules, 1~2mm, suspended in an<br>aqueous polymer carrier gel. The chemical composition of<br>the multi-porous ceramic granules is trace silicate induced<br>bi-phase calcium phosphate ceramic which has 80%<br>hydroxylapatite $Ca_5(PO_4)_3(OH)$ and 20% $β$ -tricalcium<br>phosphate $Ca_3(PO_4)_2$ , similar levels to those identified in<br>naturally-growing bone.<br>Its porous structure comprising three types of porosities<br>which are interconnected: macropores (100µm~1mm),<br>midipores (10~100 µm) and microspaces (1~10 µm).<br>Calcium phosphate bone graft substitutes have been<br>the topic of extensive clinical studies for several decades.<br>Biocompatibility is addressed in the non-clinical testing<br>section below. The interconnected macro-, midi- and<br>micro- porous structure encourages the rapid formation of<br>host bone and the growth of capillary blood vessels<br>throughout the network of interconnecting pores. After<br>implantation, si-Mochi undergoes physiologically-<br>mediated resorption and is replaced by natural bone.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                    | The resorption of the Ca/P porous ceramic granules<br>were controlled by the host nature bone remodelling<br>process due to the proliferated osteocytes formation<br>within the microporous structure of the ceramic<br>granules. The resorption is not controlled by its chemical<br>composition. si-Mochi is supplied in three different types<br>of sterile applicator. si-Mochi does not set in-situ<br>following implantation. si-Mochi does not contain<br>antibiotics.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Intended Use                                                       | si-Mochi is an implant intended to fill bony voids or<br>gaps of the skeletal system (i.e. extremities, pelvis).<br>These osseous defects may be surgically created or<br>the result of traumatic injury to the bone and are not<br>intrinsic to the stability of the bony structure. si-Mochi<br>resorbs and is replaced with bone during the healing<br>process.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Technical Characteristics<br>and Substantial<br>Equivalence        | si-Mochi and the predicate device, Actifuse ABX share<br>similar characteristics in that they are all calcium salt<br>bone void fillers covered by 'Class II Special Controls<br>Guidance Document: Resorbable Calcium Salt Bone<br>Void Filler Device" (FDA Guidance Document 855,<br>dated June 2, 2003).<br>Both si-Mochi and Actifuse ABX are packed in different<br>type of applicator for different clinic application. Both<br>have the multi-porous Ca/P ceramic granules with<br>added trace silicate, 1~2mm, suspended in the same<br>chemical composition of synthetic aqueous binding gel.<br>The multi-porous ceramic granules in si-Mochi are near<br>identical to TriPore SBG. They share the same chemical<br>composition but the multiparous Ca/P ceramic granules<br>had added additional trace silicate. They have the same<br>multi-porous structure and their manufacturing process<br>and annealing profile are exactly the same.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Determination of<br>substantial equivalence                        | Biostone has determined that si-Mochi is substantially<br>equivalent to the predicate devices on the basis of                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| (non-clinical data)                                                | chemical composition tests on all three devices as prescribed in the<br>'Class II Special Controls Guidance Document' referenced above.<br>Secondly, si-Mochi itself complies with the requirements of the Special<br>Controls Document referred to above. The non-clinical data also included<br>biocompatibility, pyrogenicity testing and endotoxin monitoring, shelf-life a<br>packaging validation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Determination of<br>substantial<br>equivalence<br>(animal studies) | Animal study, rabbit critical size defect in the distal femora model,<br>making direct comparison against the predicate device, Actifuse ABX.<br>However, the percentage of new bone formations were not statistically<br>significant difference in animal model at all time points post-implantation<br>between si-Mochi and the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Technical<br>Characteristics<br>Difference                         | si-Mochi and Actifuse ABX<br>1) The first is the chemical composition, where the calcium phosphate<br>granules in Actifuse ABX is pure hydroxylapatite with 0.8% Silicate. T<br>calcium phosphate granules in si-Mochi comes as bi-phase calcium<br>phosphate comprised of 20% beta tri-calcium phosphate and 80%<br>hydroxylapatite with 0.8% silicate. The differences in the chemical<br>composition of the porous calcium phosphate ceramic granules do not<br>affect the safety or effectiveness of si-Mochi, since all materials are<br>resorbable and provide the same function.<br>2) The second is the viscosity of the Kolliphor gel in both si-Mochi and<br>Actifuse ABX. The composition of Kolliphor gel in the Actifuse ABX is a<br>20%, and the viscosity will increase when the temperature raised to<br>human body temperature at 37°C. However, it was found that the low<br>viscosity gel is not sufficient to hold the granules together during the<br>lower operation theater temperature. Therefore, the composition of the<br>Kolliphor gel was increased to 33% to construct si-Mochi, for si-Mochi<br>function normally in operation theater condition. The Kolliphor gel is<br>bioinert and will not impede the function of the porous calcium<br>phosphate ceramic granules as the bone ingrowth scaffold.<br>si-Mochi and Tripore<br>The only difference is that the porous Ca/P ceramic granules contain<br>additional 0.8% of Silica in its chemical composition, the same as the othe<br>predicate device Actifuse ABX. However, the added Silica shall not affect<br>clinical performance. Since the osteointegration thesis of the porous cerar<br>granules and TriPore® are precisely the same.<br>These differences do not raise new issues of safety. |
| Conclusions                                                        | Therefore, Biostone concludes that the non clinical and animal studios                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

{4}------------------------------------------------

{5}------------------------------------------------

| Conclusions | Therefore, Biostone concludes that the non-clinical and animal studies<br>discussed above demonstrate that si-Mochi performs as well the predicate<br>device. |
|-------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K202639](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K202639)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
