← Product Code [MQV](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV) · K200064

# OsteoFlo NanoPutty-Quadphasic Synthetic Bone Graft (K200064)

_SurGenTec, LLC · MQV · Aug 14, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K200064

## Device Facts

- **Applicant:** SurGenTec, LLC
- **Product Code:** [MQV](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV.md)
- **Decision Date:** Aug 14, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

OsteoFlo® NanoPutty® - Quadphasic Synthetic Bone Graft is indicated to fill bony voids or gaps of the skeletal system (i.e. the extremities and pelvis) that are not intrinsic to the stability of the bony structure. These osseous defects may be surgically created or the result of traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process.

## Device Story

OsteoFlo NanoPutty is an osteoconductive, non-hardening bone void filler; composed of macroporous quadphasic calcium phosphate particulates (HA, α-TCP, β-TCP, bioactive glass) and micro-sized HA particles in a bioresorbable polymer binder. Device is supplied sterile, ready-to-use, and requires no mixing. Applied by clinicians into osseous defects; binder resorbs post-implantation; particles act as scaffold for new bone formation; particles gradually resorb. Benefits include ease of packing and adherence to bone surfaces; facilitates bone regeneration in non-structural voids.

## Clinical Evidence

Bench testing only. Included biocompatibility (ISO 10993-1), sterilization validation (ISO 11137), packaging validation (ISO 11607), shelf-life (ASTM 1980), and material characterization. In vivo evaluation in a critical-size rabbit femoral defect model compared performance to NovaBone Putty, showing equivalent bone formation and no adverse events.

## Technological Characteristics

Quadphasic calcium phosphate (HA, α-TCP, β-TCP, bioactive glass) and micro-sized HA particles in bioresorbable polymer binder. Sterile (E-beam, SAL 10^-6), single-use. Dimensions: 1.0 ml – 10 ml. Standards: ISO 10993-1, ISO 11137-1/2, ISO 11607-1/2, ASTM 1980-16.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- NovaBone Putty - Bioactive Synthetic Bone Graft ([K110368](/device/K110368.md), [K101860](/device/K101860.md), [K082672](/device/K082672.md))

## Reference Devices

- Bioactive Bone Graft Putty ([K132071](/device/K132071.md))
- NovaBone MacroPor-Si+ ([K110925](/device/K110925.md))
- NanOss BVF-E ([K081558](/device/K081558.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 14, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

SurGenTec, LLC Travis Greenhalgh CEO 911 Clint Moore Rd Boca Raton, Florida 33487

Re: K200064

Trade/Device Name: OsteoFlo® NanoPutty® - Quadphasic Synthetic Bone Graft Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: July 23, 2020 Received: July 24, 2020

Dear Mr. Greenhalgh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Laura C. Rose, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K200064

Device Name

OsteoFlo® NanoPutty®- Quadphasic Synthetic Bone Graft

Indications for Use (Describe)

OsteoFlo® NanoPuty - Quadphasic Synthetic Bone Graft is indicated to fill bony voids or gaps of the skeletal system (i.e. the extremities and pelvis) that are not intrinsic to the stability of the bony structure. These osseous defects may be surgically created or the result of traumatic injury to the bone. The device resorbs and is replaced with bonedus ing the healing process.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other spect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Surgentec logo. The logo consists of a blue and gray circular design on the left, followed by the word "SURGENTEC" in bold, black letters. A registered trademark symbol is located in the upper right corner of the word.

# Premarket Notification 510(k) Summary OsteoFlo NanoPutty Bone Void Filler - K200064

#### Submitter Information: 1.

| Owner:          | SurGenTec, LLC                             |
|-----------------|--------------------------------------------|
| Address:        | 911 Clint Moore Rd<br>Boca Raton, FL 33487 |
| Contact:        | Travis Greenhalgh, CEO                     |
| Telephone:      | (916) 759-7999                             |
| Email:          | travis@SurGenTec.com                       |
| Date Submitted: | December 9, 2019                           |

#### 2. Name of Device:

| Trade Name:          | OsteoFlo® NanoPutty® - Quadphasic Synthetic Bone Graft |
|----------------------|--------------------------------------------------------|
| Common Name:         | Bone Void Filler                                       |
| Classification Name: | Resorbable Calcium Salt Bone Void Filler               |
| Device Regulation:   | 21 CFR 888.3045                                        |
| Regulatory Class:    | Class II                                               |
| Product Code:        | MQV                                                    |

#### 3. Legally Marketed Predicate Device:

Predicate: NovaBone Putty - Bioactive Synthetic Bone Graft NovaBone Products, LLC K110368, K101860, K082672

The following devices also are cited as reference devices within the submission:

- Bioactive Bone Graft Putty: K132071; Biostructures, LLC
- NovaBone MacroPor-Si+: K110925; NovaBone Products, LLC
- NanOss BVF-E: K081558; Pioneer Surgical Technology

#### Device Description 4.

The OsteoFlo® NanoPutty® - Quadphasic Synthetic Bone Graft is an osteoconductive, non-hardening bone void filler. The device is comprised of macroporous calcium phosphate particulates in a bioresorbable polymer binder. The quadphasic particles are composed of HA, a-TCP, B-TCP and bioactive glass. The device also contains micro-sized HA particles. On implantation, the binder is resorbed and bone forms on and between the porous particles as they are gradually resorbed.

{4}------------------------------------------------

OsteoFlo® NanoPutty can be easily packed into osseous defects and adheres to bone surfaces. The device is ready to use, requiring no mixing before application. The single-use device is supplied sterile in multiple package formats.

#### 5. Indications for Use

The OsteoFlo® NanoPutty® - Quadphasic Synthetic Bone Graft is indicated to fill bony voids or gaps of the skeletal system (i.e. the extremities and pelvis) that are not intrinsic to the stability of the bony structure. These osseous defects may be surgically created or the result of traumatic injury to the bone. The device resorbs and is replaced with bone during the healing process.

#### Technological Characteristics and Substantial Equivalence 6.

OsteoFlo® NanoPutty - Ouadphasic Synthetic Bone Graft is substantially equivalent to the predicate NovaBone Putty Bioactive Synthetic Graft. The devices have the same intended use and basic method of operation. The table below compares the indications and key technological attributes of the OsteoFlo NanoPutty device to the predicate.

| Comparator                   | New Device                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                              | OsteoFlo NanoPutty                                                                                                                                                                                                                                                                                                                                                                                                                | NovaBone Putty [K110368]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Product Code;<br>Common Name | Resorbable Calcium Salt Bone Void Filler Device<br>(21 CFR 888.3045)                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Indications for<br>Use       | OsteoFlo® NanoPutty –<br>Quadphasic Synthetic Bone<br>Graft is indicated to fill bony<br>voids or gaps of the skeletal<br>system (i.e. the extremities and<br>pelvis) that are not intrinsic to<br>the stability of the bony<br>structure. These osseous defects<br>may be surgically created or the<br>result of traumatic injury to the<br>bone. The device resorbs and is<br>replaced with bone during the<br>healing process. | NovaBone Putty is indicated for bony<br>voids or gaps that are not intrinsic to<br>the stability of the bony structure.<br>NovaBone Putty is indicated to be<br>gently packed into bony voids or gaps<br>of the skeletal system (i.e. the<br>extremities, posterolateral spine, and<br>pelvis). These defects may be<br>surgically created osseous defects or<br>osseous defects created from traumatic<br>injury to the bone. The product<br>provides a bone void filler that resorbs<br>and is replaced with bone during the<br>healing process. |
| Intended Use                 | Both devices are intended for use as bone void fillers for repair of defects<br>in the skeletal system. The devices are intended to be resorbed during<br>defect healing and replaced with new bone tissue.                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Mechanism of<br>Operation    | The device is applied as a pre-<br>mixed putty. The synthetic<br>binder is resorbed to expose<br>quadphasic particles. The<br>particles act as osteoconductive<br>scaffolds for new bone<br>formation as they are slowly<br>resorbed                                                                                                                                                                                              | The device is applied as a pre-mixed<br>putty. The synthetic binder is resorbed<br>to expose bioactive glass particles.<br>The particles act as osteoconductive<br>scaffolds for new bone formation as<br>they are slowly resorbed                                                                                                                                                                                                                                                                                                                 |
| Material<br>Composition      | Quadphasic macroporous<br>calcium phosphate particulates<br>in a bioresorbable polymer                                                                                                                                                                                                                                                                                                                                            | Bioactive glass particles, 90-710 μm,<br>in a bioresorbable polymer binder                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

{5}------------------------------------------------

| Comparator   | New Device<br>OsteoFlo NanoPutty                   | Predicate<br>NovaBone Putty [K110368]         |
|--------------|----------------------------------------------------|-----------------------------------------------|
|              | binder                                             |                                               |
| Sterility    | Sterile; SAL 10-6;<br>E-beam radiation; Single-Use | Sterile, SAL 10-6;<br>γ radiation; Single-Use |
| Product Size | 1.0 ml – 10 ml                                     | 2.5 ml – 15 ml                                |

#### 7. Performance Data

The following non-clinical performance data were provided to demonstrate substantial equivalence of the subject device to the predicate.

- Biocompatibility per ISO 10993-1:2018
- . Sterilization validation per ISO 11137-1:2006 and ISO 11137-2:2013
- Packaging validation per ISO 11607-1:2009 and ISO 11607-2:2006 ●
- Shelf-life testing per ASTM 1980-16 ●
- Material characterization, including x-ray diffraction, particle size, and ● particle porosity and surface area
- . In vivo evaluation in a critical-size rabbit femoral defect model

The in vivo evaluation compared the performance of the device to the NovaBone Putty predicate. Evaluation of local biological effects and bone formation demonstrated performance substantially equivalent to the predicate, with no evidence of adverse events or device related failures.

#### 8. Conclusion

The design characteristics of the OsteoFlo NanoPutty device do not raise different questions of safety and effectiveness. Non-clinical study data supports that the device is safe and effective. These data support that the OsteoFlo NanoPutty device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K200064](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K200064)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
