← Product Code [MQV](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV) · K190110

# NanoBone SBX PUTTY, NanoBone QD (K190110)

_Artoss GmbH · MQV · Apr 23, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K190110

## Device Facts

- **Applicant:** Artoss GmbH
- **Product Code:** [MQV](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV.md)
- **Decision Date:** Apr 23, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

NanoBone® SBX Putty is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis). These osseous defects are surgically created of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. NanoBone® SBX Putty resorbs and is replaced with bone during the healing process.

## Device Story

NanoBone® SBX Putty/QD is a resorbable bone void filler; consists of non-sintered nanocrystalline hydroxyapatite (HA) embedded in a porous silica gel matrix with a poloxamer carrier. Supplied in sterile applicators or cartridges; used by surgeons to fill osseous defects in extremities, pelvis, or posterolateral spine. Device does not set in-situ; acts as an osteoconductive scaffold; resorbs over time and is replaced by natural bone during healing. Provides a stable environment for bone regeneration; intended for non-load-bearing applications. No complex electronics or software involved.

## Clinical Evidence

Bench testing included chemical/elemental analysis, XRD phase purity, dissolution, biocompatibility, pyrogenicity, sterilization validation, and shelf-life testing. Additional animal testing was conducted to support the posterolateral fusion indication. Results demonstrate the device functions as intended with a safety and effectiveness profile similar to predicate and reference devices.

## Technological Characteristics

Material: Non-sintered nanocrystalline hydroxyapatite (HA) in silica gel matrix; Poloxamer carrier. Structure: Highly porous (nano, micro, macro pores) mimicking human cancellous bone. Complies with ASTM F1185. Sterilization: Radiation. Form factor: Putty supplied in sterile applicators/cartridges (0.25mL to 10mL).

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- NanoBone® SBX Putty ([K161351](/device/K161351.md))

## Reference Devices

- APATECH Actifuse ABX E-Z-Fil Putty Bone Graft Substitute ([K071206](/device/K071206.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

ARTOSS GmbH Walter Gerike, Ph.D. Managing Partner Friedrich-Barnewitz-Strasse 3 18119 Rostock. Germany

Re: K190110

Trade/Device Name: NanoBone® SBX PUTTY, NanoBone® QD Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: January 21, 2019 Received: January 23, 2019

Dear Dr. Gerike:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K190110

Device Name NanoBone® SBX PUTTY

#### Indications for Use (Describe)

NanoBone® SBX Putty is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis). These osseous defects are surgically created of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. NanoBone® SBX Putty resorbs and is replaced with bone during the healing process.

Type of Use (*Select one or both, as applicable*)

| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)                 |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K190110

Device Name NanoBone® QD

Indications for Use (Describe)

NanoBone® QD Putty is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis). These osseous defects are surgically created of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. NanoBone® QD Putty resorbs and is replaced with bone during the healing process.

| Type of Use (Select one or both, as applicable)                                     |                                                                                    |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "NanoBone" in a stylized font. The word "Nano" is in pink, while the word "Bone" is in green. There is a circled R symbol to the right of the word "Bone", indicating that it is a registered trademark.

510(k) Summary [as required by 21 CFR 807.92(c)]

NanoBone® Bone Graft Substitute NanoBone® SBX Putty NanoBone® QD

510(k) 190110

| DATE PREPARED:    | 18 March 2019                                                                                                                                    |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT:        | ARTOSS GmbH<br>Friedrich-Barnewitz-Strasse 3<br>18119 Rostock, Germany<br>Telephone: +49 (0) 381 5 43 45 - 701<br>Fax: +49 (0) 381 5 43 45 - 702 |
| CONTACT:          | Walter Gerike, PhD<br>Managing Partner<br>ARTOSS GmbH<br>gerike@artoss.com                                                                       |
| TRADE NAME:       | NanoBone® bone graft substitutes<br>• NanoBone® SBX Putty<br>• NanoBone® QD                                                                      |
| COMMON NAME:      | Bone Void Filler                                                                                                                                 |
| CLASSIFICATION    | Resorbable calcium salt bone void filler                                                                                                         |
| REGULATION:       | 21 CFR 888.3045, Product Code MQV                                                                                                                |
| DEVICE CLASS:     | 2                                                                                                                                                |
| PANEL CODE:       | Orthopedic / 888                                                                                                                                 |
| PREDICATE DEVICE: | ARTOSS NanoBone® bone graft substitutes<br>NanoBone® SBX Putty (K161351)                                                                         |
| REFERENCE DEVICE: | APATECH Actifuse ABX E-Z-Fil Putty Bone Graft<br>Substitute (K071206)                                                                            |

### INTENDED USE/INDICATIONS FOR USE:

NanoBone® SBX Putty and NanoBone® QD have the same intended use and indications for use:

NanoBone® SBX Putty is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. NanoBone® SBX Putty resorbs and is replaced with bone during the healing process.

{5}------------------------------------------------

NanoBone® OD Putty is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis). These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. NanoBone® OD Putty resorbs and is replaced with bone during the healing process.

## DEVICE DESCRIPTION:

NanoBone® SBX Putty consists of NanoBone® granulate embedded in an aqueous gel. NanoBone granulate consists of phase-pure non-sintered nanocrystalline osteoconductive hydroxyapatite (HA) embedded in a highly porous silica gel matrix. The high porosity of the product includes nano pores, micro pores, and macro pores. The interconnected and open porous structure of the macro pores of the NanoBone® is similar to human cancellous bone. NanoBone® SBX Puttv does not set in-situ following implantation.

NanoBone® SBX material is supplied in two different style applicators. The NanoBone® SBX Putty is supplied in a sterile applicator with an attached plunger. NanoBone® QD is an alternate packaging of the NanoBone® SBX Putty. NanoBone® OD is supplied in a sterile cartridge with a separate sterile plunger.

| Name of<br>Devices | NanoBone® SBX Putty<br>/ NanoBone® QD                     | NanoBone® SBX Putty                                       | Actifuse ABX Putty                                                                   |
|--------------------|-----------------------------------------------------------|-----------------------------------------------------------|--------------------------------------------------------------------------------------|
|                    | Subject                                                   | Predicate                                                 | Reference                                                                            |
| 510(k)             |                                                           | K161351                                                   | K071206                                                                              |
| Intended Use       | Bone void filler                                          | Bone void filler                                          | Bone void filler                                                                     |
| Target Location    | Extremities, pelvis, spine                                | Extremities, pelvis                                       | Extremities, pelvis, spine                                                           |
| Formulation        | Putty                                                     | Putty                                                     | Putty                                                                                |
| Material           | Hydroxyapatite (HA)<br>embedded in a silica gel<br>matrix | Hydroxyapatite (HA)<br>embedded in a silica gel<br>matrix | Silicate substituted<br>calcium phosphate<br>(hydroxyapatite)                        |
| Carrier            | Poloxamer                                                 | Poloxamer                                                 | Poloxamer                                                                            |
| Granule Size       | 0.6mm x 2mm                                               | 0.6mm x 2mm                                               | 1mm-2mm                                                                              |
| How Supplied       | In applicators ranging in<br>size from 0.25mL to 10<br>mL | In applicators ranging in<br>size from 0.25mL to 10<br>mL | Pack size:<br>1.5mL to 20mL<br>E-Z Prep applicator:<br>10 mL<br>MIS cartridge: 7.5mL |
| Sterilization      | Radiation                                                 | Radiation                                                 | Radiation                                                                            |

#### COMPARISON WITH PREDICATE AND REFERENCE DEVICES:

The NanoBone® SBX Putty/NanoBone® OD and the predicate and reference devices have the same intended use, to fill bony voids and gaps that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or those created from traumatic injury to the bone.

The subject device was not been previously submitted for a posterolateral indication. The subject device and the predicate device have identical design and technological characteristics. NanoBone® SBX Putty/NanoBone® OD have the same indications, contraindications, risks, basic technology, equivalent materials, and potential adverse events as the NanoBone® predicate (K161351). The only difference between the predicate ARTOSS 

{6}------------------------------------------------

NanoBone® bone graft substitute and the subject device is the addition of posterolateral fusion indication, and the addition of the alternate applicator.

## NON-CLINICAL TESTING/PERFORMANCE DATA:

Previously submitted bench testing has shown the NanoBone® products meet all relevant requirements for calcium salt bone void filler devices, including ASTM F1185. Testing was performed on the predicate device to characterize and evaluate the performance of the NanoBone® products. The testing included:

- Chemical / elemental analysis ●
- Phase purity / XRD ●
- . Dissolution testing
- Animal testing ●
- Biocompatibility assessment / testing
- Pyrogenicity ●
- Sterilization validation ●
- Shelf life

The in vitro bench tests demonstrated that the NanoBone® bone graft substitutes met all acceptance criteria and performed similarly to the predicate devices.

Additional animal testing was performed to show the performance for use in posterolateral fusion. Performance data demonstrate that the device functions as intended and has a safety and effectiveness profile that is similar to the predicate and reference devices.

# BIOCOMPATIBILITY:

The NanoBone® bone graft substitutes were compared to the predicate and reference devices. Based on the use of identical materials in the subject device and predicates devices, the biocompatibility of the NanoBone® bone graft substitutes was verified to be the same as those of the predicates.

# CONCLUSION:

The NanoBone® bone graft substitutes (NanoBone® SBX Putty and NanoBone® OD) have the same intended use and the same technological characteristics such as components, design, materials, sterilization method, and operating principles as the predicate. Performance data demonstrates that the device functions as intended.

Therefore, the NanoBone® SBX Putty and NanoBone® QD are substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K190110](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K190110)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
