CALLOS PROMODEL BONE VOID FILLER, SKAFFOLD MP (MACRO POROUS) BONE VOID FILLER
Applicant
Skeletal Kinetics, LLC
Product Code
MQV · Orthopedic
Decision Date
Feb 7, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Callos ProModel is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, posterolateral spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos ProModel is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced with bone during the healing process
Device Story
Callos ProModel is a moldable, biocompatible calcium phosphate/calcium sulphate composite bone void filler. The device is supplied as a single-use kit containing powder, dilute sodium silicate liquid, a mixing system, and a cannula. The clinician mixes the powder and liquid to form a paste, which is applied directly to the bone void. The material acts as a scaffold that resorbs and is replaced by natural bone during the healing process. It is intended for use in non-load-bearing (non-structural) defects. The device is used by surgeons in clinical settings to treat osseous defects.
Clinical Evidence
No clinical data. Evidence consists of bench testing and a large animal critical size defect model comparing the device to predicate devices, demonstrating similar biologic response.
Technological Characteristics
Composite of calcium phosphate/calcium sulfate powder and sodium silicate solution. Forms a moldable paste. Sterilized via radiation (ANSI/AAMI/ISO 11137-2:2006). Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for filling non-structural bony voids or gaps in the skeletal system, including extremities, posterolateral spine, and pelvis, resulting from surgical procedures or traumatic injury.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
K212721 — Genex Bone Graft Substitute · Biocomposites, Ltd. · Feb 4, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
KI12383 PAGE 2 OF 2
FEB - 7 2012
Image /page/0/Picture/2 description: The image shows the logo for Skeletal Kinetics. The logo consists of a circular graphic to the left of the text "Skeletal Kinetics". The graphic is a circle with a complex network of lines inside, resembling a geodesic dome or a stylized representation of a cell structure. The text "Skeletal Kinetics" is in a bold, sans-serif font.
# A COLSON ASSOCIATE
# 510(k) Summary
# General Information as required by 21 CFR 807.92 (a) (1)
| Submitters Name/address: | Skeletal Kinetics® LLC<br>10201 Bubb Road<br>Cupertino, CA 95014, USA |
|--------------------------|----------------------------------------------------------------------------------------------------------------------|
| Date: | February 2, 2012 |
| Contact Person: | Christine Kuo,<br>Director, Regulatory Affairs and Quality Assurance<br>Phone: (408) 350-5842<br>Fax: (408) 366-1077 |
# Device Name as required by 21 CFR 807.92 (a) (2)
| Trade Name: | Callos ProModel Bone Void Filler |
|--------------------|-------------------------------------------------|
| Regulation Number: | 21 CFR 888.3045 |
| Regulation Name: | Resorbable calcium salt bone void filler device |
| Regulatory Class: | II |
| Product Code: | MQV |
# Predicate Devices as required by 21 CFR 807.92 (a) (3)
The subject device is substantially equivalent to the following legally marketed devices (predicates):MIIG® SR Bone Void Filler (K060011), PRO-DENSE™ Bone Graft Substitute (K070437) and Callos® Bone Void Filler (K051123).
# Device Description as required by 21 CFR 807.92 (a) (4)
Callos ProModel Bone Void Filler is a moldable and biocompatible calcium phosphate/calcium sulphate composite bone void filler. The single-use Callos ProModel kit contains the necessary components for mixing of the bone void filler. The Callos ProModel sterile kit contains: Calcium Phosphate/Calcium Sulfate Powder, Dilute Sodium Silicate Liquid, a Mixing System (Mixer and Syringe) and a Cannula.
{1}------------------------------------------------
# Intended Use as required by 21 CFR 807.92 (a) (5)
Callos ProModel is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, posterolateral spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos ProModel is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced with bone during the healing process
# Summary of Technological Characteristics as required by 21 CFR 807.92 (a) (6)
Callos ProModel consists of calcium phosphate/calcium sulfate powder and sodium silicate solution; when mixed together it forms a paste that can be applied directly to the injured bone.
The intended use, operating principles, design features, and materials of Callos ProModel is substantially equivalent to predicate devices.
# Summary of Non-clinical Tests as required by 21 CFR 807.92 (b) (1)
The results of biocompatibility testing demonstrated that Callos ProModel met the standards set forth in ISO 10993-1. Biological Evaluation of Medical Devices, and the radiation sterilization validation is in compliance to ANSI/A AMI/ISO 11137-2:2006.
Bench testing demonstrated that Callos ProModel is substantially equivalent to the marketed predicate devices. The results indicated that the device met performance criteria outlined for its intended use.
The animal study provided data that Callos ProModel elicited similar biologic response in a large animal critical size defect model as compared to predicate devices for its intended use.
# Summary of Clinical Tests as required by 21 CFR 807.92 (b) (2)
Callos ProModel does not meet the criteria to require clinical testing.
# Conclusion as required by 21 CFR 807.92 (3)
In vitro and in vivo performance demonstrate that Callos ProModel is substantially equivalent to legally marketed predicate devices.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with its wings spread, facing right. The eagle is composed of three thick, curved lines. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Skeletal Kinectics® LLC % Ms. Christine Kuo Director, Regulatory Affairs and Quality Assurance 10201 Bubb Road Cupertino, California 95014
Re: K112383
Trade/Device Name: Callos® ProModel Bone Void Filler Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV · Dated: January 17, 2012 Received: January 18, 2012
Dear Ms. Kuo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
- 7 2012 FEB
{3}------------------------------------------------
## Page 2 - Ms. Christine Kuo
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you active spective ad 100 boutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part nove the regulation unities, the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Molkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Section 4: Indications for Use
510(K) Number (if Known):
K112383
Device Name:
Callos® ProModel Bone Void Filler
#### Indications for Use:
Callos® ProModel Bone Void Filler is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, posterolateral spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos ProModel is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
Prescription Use (Per 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21CFR 801 Subpart C
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
x
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
Division Slon-Off Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K112383
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.