K073405 · Osteotech, Inc. · MQV · Mar 3, 2008 · Orthopedic
Device Facts
Record ID
K073405
Device Name
PLEXUR M
Applicant
Osteotech, Inc.
Product Code
MQV · Orthopedic
Decision Date
Mar 3, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Intended Use
PLEXUR M is intended for use in filling bony voids or gaps of the skeletal system (i.e., extremities, pelvis) that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. PLEXUR M is resorbed/remodeled and is replaced by host bone during the healing process.
Device Story
PLEXUR M is a bone void filler composed of processed human allograft bone particles mixed with resorbable/biodegradable polymers. The device is provided sterile for single use. At the time of surgery, the material is heated—using an included single-use heater or a water bath—to become moldable, allowing the surgeon to pack or shape it into the defect site. Upon cooling to body temperature, the material hardens; it can be further shaped, cut, or ground by the surgeon using conventional instruments. The device is resorbed and remodeled into host bone during the healing process. It is intended for use by surgeons in clinical settings to address osseous defects not intrinsic to skeletal stability.
Clinical Evidence
No human clinical data provided. Evidence consists of animal studies demonstrating that the device supports bone ingrowth and new bone formation.
Technological Characteristics
Bone void filler containing processed human allograft bone and resorbable/biodegradable polymers. Provided sterile. Thermoplastic properties: moldable when heated, rigid at body temperature. Includes a compact, single-use heater for activation. No specific ASTM standards cited in the provided text.
Indications for Use
Indicated for filling non-structural bony voids or gaps in the skeletal system, including extremities and pelvis, resulting from surgical procedures or traumatic injury.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Related Devices
K081227 — PLEXUR M · Osteotech, Inc. · Jul 22, 2008
K061982 — PLEXUR P · Osteotech, Inc. · Feb 7, 2007
K080511 — PLEXUR-P · Osteotech, Inc. · Apr 21, 2008
K103036 — ALLOFUSE PLUS · Allosource, Inc. · Jan 10, 2011
{0}------------------------------------------------
510(K) Pre Market Application
#### 510(k) Summary 5.0
# MAR - 3 2008
The following information is provided as required by 21 CFR § 807.87 for Osteotech's PLEXUR M 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.
The safety and effectiveness of PLEXUR M is based upon a determination of the substantial equivalence as well as the safety and effectiveness of its predicate device(s) .
| Sponsor/<br>Manufacturer | Osteotech, Inc.<br>51 James Way<br>Eatontown, NJ 07724 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | Chris Talbot<br>Osteotech, Inc.<br>51 James Way<br>Eatontown, NJ 07724<br>Ph: (732) 542-2800<br>Talbot@osteotech.com |
| Date Prepared | November 30, 2007 |
| Proposed Class | II |
| Proprietary Name | PLEXUR M |
| Common Name | Bone void filler |
| Classification Name | Resorbable calcium salt bone void filler device |
| Regulation Number | 21 CFR 888.3045 |
| Product Code | MQV |
| Predicate Device | PLEXUR M is substantially equivalent, for the purpose of this<br>510(k), to other devices that have received 510(k) clearance for similar indications for<br>use. |
PLEXUR M is intended for use in filling bony voids or gaps of the Intended Use skeletal system (i.e., extremities, pelvis) that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects
{1}------------------------------------------------
created from traumatic injury to the bone. PLEXUR M is resorbed/remodeled and is replaced by host bone during the healing process.
## Device Description
PLEXUR M is a bone void filler product that contains, as a key constituent, processed human bone particles that are mixed with resorbable/biodegradeable non-tissue components. It is intended to be heated at the time of use, whereupon it becomes moldable, thus allowing the surgeon to pack it into the implant site or shape it to accommodate variations in the geometry and size of the particular implant site. As it cools down, the PLEXUR M returns to its normal hardened/rigid state and remains this way at body temperature. The surgeon may further shape, cut or grind PLEXUR M in the hardened state using conventional surgical instruments.
PLEXUR M is packaged/provided for single use in a sterile form. PLEXUR M is resorbed/remodeled and is replaced by host bone during the healing process. A compact, single use heater is available as a means to heat the PLEXUR M at the time of use to make it pliable. Osteotech is seeking approval of this single use heater as part of this 510(k) submission. Alternatively, PLEXUR M may be heated in a water bath.
### Performance Data
The results of studies in animals showed that PLEXUR M supports bone ingrowth and new bone formation.
### Viral Inactivation
In the production of PLEXUR M, the allograft bone is subjected to processing steps that have been shown to inactivate viruses, including HIV, hepatitis B and C and CMV.
### Technical Comparison
PLEXUR M is substantially equivalent to one or more of the predicate devices with respect to materials. PLEXUR M contains human allograft bone tissue, as does one or more of the predicate devices. PLEXUR M also contains resorbable polymer of the same
{2}------------------------------------------------
type as those in one or more of the predicate devices. Also, like one or more of the predicate devices, PLEXUR M is provided sterile in various sizes that, upon heating (using a compact, single use sterile heater to be marketed by Osteotech), is made moldable and can be cut or shaped by the user into various shapes or sizes. It may be further shaped by the surgeon in its hardened state using conventional surgical instruments.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Osteotech, Inc. % Mr. Christopher Talbot Director of Regulatory Affairs 51 James Way Eatontown, New Jersey 07724
MAR - 3 2008
Re: K073405
Trade/Device Name: PLEXUR M Bone Void Filler Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler devices Regulatory Class: Class II Product Code: MOV Dated: November 30, 2007 Received: December 4, 2007
Dear Mr. Talbot:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 – Mr. Christopher Talbot
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use Statement 4.0
K073405 510(k) Number:
Device Name: PLEXUR M
PLEXUR M is intended for use in filling bony voids or gaps of the Indications for Use: skeletal system (i.e., extremities, pelvis) that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. PLEXUR M is resorbed/remodeled and is replaced by host bone during the healing process.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE
Mark A. Mckerson
(Division Sign-Of al. Pestorative, Division of Genera and Neurological Dences
510(k) Number
Page 1 of ___1__
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.