← Product Code [MQV](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV) · K040755

# EXACTECH RESORBABLE ROOM TEMPERATURE BONE PASTE (K040755)

_Exactech, Inc. · MQV · Jun 17, 2004 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K040755

## Device Facts

- **Applicant:** Exactech, Inc.
- **Product Code:** [MQV](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV.md)
- **Decision Date:** Jun 17, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Exactech Resorbable Room Temperature Bone Paste is intended for use as a bone graft extender (extremities, spine, pelvis) and as a bone void filler for bony voids or gaps of the extremities and pelvis that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Exactech Resorbable Room Temperature Bone Paste may be used with rigid fixation systems.

## Device Story

Exactech Resorbable Room Temperature Bone Paste is an implantable bone void filler composed of demineralized bone matrix (DBM) in a bioinert, bioabsorbable polyethylene glycol (PEG) polymer carrier. Provided as an aseptic, single-use, ready-to-use paste derived from a single donor. Intended for surgical application by clinicians to fill non-structural osseous defects in extremities, spine, and pelvis. The paste gradually resorbs and is replaced by new bone during the healing process. It may be used in conjunction with rigid fixation systems. The device provides no structural support; therefore, it is contraindicated in skeletal sites where structural integrity is required during healing. Safety is managed through donor screening, aseptic manufacturing, and viral inactivation validation.

## Clinical Evidence

No clinical data provided. Evidence consists of bench testing, including microbiologic sterility testing per 21CFR610.12, biocompatibility testing per ISO 10993, and an osteoinductive potential assay in an athymic nude mouse model (ASTM Draft Standard). Viral inactivation validation was performed using four virus models (HAV, HIV-1, PPV, PrV) to demonstrate the demineralization process reduces viral titers.

## Technological Characteristics

Composition: Demineralized bone matrix (DBM) in a polyethylene glycol (PEG) polymer carrier. Form: Ready-to-use paste. Sterilization: Aseptic manufacture. Biocompatibility: Tested per ISO 10993. Osteoinduction: Assessed via athymic mouse model. Viral inactivation: Validated via four-virus model study.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- Exactech Resorbable Bone Paste ([K020078](/device/K020078.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K04-0755

## JUN 1 7 2004

## Special 510(k) Summary of Safety and Effectiveness

This 510(k) Summary for Exactech Resorbable Room Temperature Bone Paste is provided as required per Section 513(1)(3) of the Food, Drug and Cosmetic Act.

| 1. | Submitter:                                                       | Exactech Inc.<br>2320 NW 66th Court<br>Gainesville, Florida 32653<br>Telephone 352-377-1140<br>Fax 352-378-2617                              |
|----|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
|    | Contact person:                                                  | Steve Lin, D.Sc<br>Vice President, Advanced Technology & Business Development<br>Exactech Inc.<br>Telephone 352-377-1140<br>Fax 352-378-2617 |
|    | Date of original submission:                                     | 03-19-2004                                                                                                                                   |
|    | FDA Establishment Number                                         | 1038671                                                                                                                                      |
| 2. | Proprietary Name:                                                | Exactech Resorbable Room Temperature Bone<br>Paste                                                                                           |
|    | Common Name:                                                     | Bone void filler                                                                                                                             |
|    | Product Code:                                                    | MQV, MBP                                                                                                                                     |
|    | Device Class:                                                    | Class II                                                                                                                                     |
|    | Classification Panel:                                            | Orthopaedic                                                                                                                                  |
| 3. | Legally Marketed Devices for Substantial Equivalence Comparison: |                                                                                                                                              |

## ਦ

| Product Code | Manufacturer  | 510(k) Number | Product                       |
|--------------|---------------|---------------|-------------------------------|
| MQV, MBP     | Exactech Inc. | K020078       | Exactech Resorbable Bone Past |

#### Comparison to the Predicate Device(s): 4.

Exactech Resorbable Room Temperature Bone Paste shares the same components, function and intended use as the predicate device, Exactech Resorbable Bone Paste. Modifications to this device do not affect safety and efficacy of the device and is therefore substantially equivalent to predicate device.

{1}------------------------------------------------

## Special 510(k) Summary of Safety and Effectiveness

#### Device Description: 5.

Exactech Resorbable Room Temperature Bone Paste is a mixture of demineralized bone matrix (DBM) in a bioinert and bioabsorbable polyethylene divcol (PEG) based polymer and is provided as an aseptic manufacture single use, ready to use implantable device derived from a single donor. Exactech Resorbable Room Temperature Bone Paste gradually resorbs and is replaced with new bone during the healing process.

#### 6. Indications for Use

Exactech Resorbable Room Temperature Bone Paste is intended for use as a bone graft extender (extremities, spine, pelvis) and as a bone void filler for bony voids or gaps of the extremities and pelvis that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.

Exactech Resorbable Room Temperature Bone Paste may be used with rigid fixation systems.

#### 7. Contraindications:

Exactech Resorbable Room Temperature Bone Paste is not intended to provide structural support during the healing process; therefore. Exactech Resorbable Room Temperature Bone Paste is contraindicated in cases where structural support of the skeletal system is required by the graft during healing.

This allograft should not be implanted into sites with an active infection.

Polymixin Sulfate B and Bacytracin are used in processing this graft and trace amounts remain. Since it is impossible to quantify the levels at which any individual may have an allergic response, this product is contraindicated in patients with known sensitivity.

#### 8. Safety and Effectiveness Information:

Exactech Resorbable Room Temperature Bone Paste is aseptically manufactured for single patient use. Each lot of Exactech Resorbable Room Temperature Bone Paste undergoes microbiologic testing for sterility in conformance with 21CFR610.12.

Human demineralized bone used in Exactech Resorbable Room Temperature Bone Paste is known to be biocompatible based on its long history of use without adverse reactions or complications.

The polymer carrier used in Exactech Resorbable Room Temperature Bone Paste was demonstrated to be biocompatible in accordance with ISO 10993 biocompatibility testing and other in vivo testing.

Rev 06-14-2004

{2}------------------------------------------------

### Special 510(k) Summary of Safety and Effectiveness

#### Osteoinductive Potential a.

Test samples from each lot of Exactech Resorbable Room Temperature Bone Paste are implanted bilaterally into the gastrocnemius muscle of 2 athymic nude mice in accordance with the ASTM Draft Standard Guide for the Assessment of Bone Inductive Material. A minimum of two implanted samples must demonstrate osteoinductive potential as assessed by histological evidence of endochondral bone formation, including the presence of cartilage or chondrocytes, active osteoblasts, osteoid, newly formed and mineralized bone and/or marrow and associated fat cells. Osteoinduction assay results in the athymic mouse model should not be interpreted to predict clinical performance in human subjects.

#### Viral Inactivation Validation b.

A viral reduction study was conducted by a CLIA certified testing laboratory using four virus models representing RNA, DNA, envelope and non-envelope virus, which included: 1) Hepatitis A Virus (HAV), non-enveloped, RNA-containing picornavirus 2) Human immunodeficiency virus type 1 (HIV-1) enveloped, RNA-containing retrovirus, 3) Porcine parvovirus (PPV) non-enveloped, DNA-containing parvovirus, which serves as a model for other parvovirus such as human parvovirus B19 and 4) Pseudorabies virus (PrV), enveloped, DNA-containing virus belonging to the Herpesviridae family and serves as a model for other herpesviruses such as Cytomegalovirus (CMV). This study demonstrates the demineralization process used on donor bone contained in Exactech Resorbable Room Temperature Bone Paste significantly diminishes these model viruses and can reasonably be anticipated to diminish the titers of other viruses.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an emblem of an eagle with its wings spread, clutching a staff with a snake wrapped around it, which is a symbol of medicine and healing.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 1 7 2004

Ms. Rebecca S. Roberts CTBS Regulatory Affairs Representative - Biologics Exactech Inc. 2320 NW 66th Court Gainesville, Florida 32653

Re: K040755

Exactech Resorbable Room Temperature Bone Paste Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler devices Regulatory Class: Class II Product Code: MQV, MBP Dated: May 17, 2004 Received: May 19, 2004

Dear Ms. Roberts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if

{4}------------------------------------------------

# Page 2 - Ms. Rebecca S. Roberts CTBS

applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) This iction will anow you to organ made of substantial equivalence of your device to a premarket notification. - The PDF intilation for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), if you desire specific advice to: Jour are at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,
O. Merle A. Melkersen

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Exactech Resorbable Room Temperature Bone Paste

# Indications for Use

510(k) Number (if known):

K040155

Exactech Resorbable Room Temperature Bone Paste

Indications for Use:

Device Name:

Exactech Resorbable Bone Paste is intended for use as a bone graft extender (extremities, spine, Exacted Resorbable Bone it asie is intended for acc a benefit es and pelvis that are not pelvis) and as a bone vold finer for borry volus of gape of the may be surgically created osseous infirmsic to the stabliky of the won't from traumatic injury to the bone.

Exactech Resorbable Bone Paste may be used with rigid fixation systems.

| Prescription Use X<br>(Part 21 CFR 801 Subpart D)        | AND/OR | Over-The-Counter Use<br>(21 CFR 807 Subpart C) |  |  |  |  |
|----------------------------------------------------------|--------|------------------------------------------------|--|--|--|--|
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER |        |                                                |  |  |  |  |
| PAGE OF NEEDED)                                          |        |                                                |  |  |  |  |

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of General, Restorative, and Neurological Devices

| 510(k) Number | K040753 |
|---------------|---------|
|---------------|---------|

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K040755](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQV/K040755)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
