← Product Code [MQP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP) · K241468

# Vertiwedge® Intraosseous System (K241468)

_Foundation Surgical Group, Inc. · MQP · Nov 7, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K241468

## Device Facts

- **Applicant:** Foundation Surgical Group, Inc.
- **Product Code:** [MQP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP.md)
- **Decision Date:** Nov 7, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Vertiwedge® Intraosseous System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement of a diseased or damaged vertebral body resected or excised to replace a portion and/or restore the height of a collapsed vertebral body for the treatment of previous trauma (i.e. fracture or tumor) or degenerative spine disease. The system is to be placed within the vertebral body following an osteotomy and supplemented with autograft or allograft and to be used with its contralateral staple; should a physician choose to use fewer than the maximum number of screws, then supplemental fixation must be used to augment stability. The Vertiwedge® Intraosseous System is intended to restore the integrity of the spinal column even in the absence of fusion in patients with advanced stage tumors involving the thoracolumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

## Device Story

The Vertiwedge® Intraosseous System is a motion-preservation partial vertebral replacement device used in Vertebral Body Osteotomy (VBO®) procedures. It is implanted via an anterior to psoas/oblique approach into a single osteotomized vertebral body. The system comprises an intraosseous wedge/spacer, a lateral plate, and a contralateral staple for integrated fixation. It is filled with autogenous or allogeneic bone graft. Protrusions on the superior and inferior surfaces provide expulsion resistance. The device is used by surgeons to restore spinal column integrity and vertebral height. It does not span disc levels. Clinical benefit includes structural support and stability in patients with trauma, degenerative disease, or advanced tumors where fusion is not feasible.

## Clinical Evidence

Bench testing only. Testing included ASTM F2077 static and dynamic axial compression, compression shear, and torsion; ASTM F1877 particulate analysis; ASTM F2267 subsidence testing; expulsion testing; staple and plate pull-apart testing; screw back-out testing; and implantation testing. Results demonstrate mechanical strength sufficient for intended use and equivalence to predicate devices.

## Technological Characteristics

Materials: Ti-6Al-4V ELI (ASTM F3001 or ASTM F136). Design: Non-expandable intraosseous wedge/spacer with lateral plate and contralateral staple. Fixation: Integrated, screw-based. Connectivity: None. Sterilization: Not specified. Software: None.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- PILLAR SA PEEK Spacer System ([K081849](/device/K081849.md))
- VLIFT®-s Vertebral Body Replacement System ([K183071](/device/K183071.md))
- Matrixx System ([K171140](/device/K171140.md))
- Dominion Expandable Corpectomy System ([K233359](/device/K233359.md))
- AccuFit Lateral Plate System ([K220324](/device/K220324.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size, also in blue.

November 7, 2024

Foundation Surgical Group, Inc. % Nathan Wright, MS, RAC Engineer & Regulatory Specialist Empirical Technologies 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K241468

Trade/Device Name: Vertiwedge® Intraosseous System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: MQP Dated: October 3, 2024 Received: October 3, 2024

Dear Mr. Wright:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Brent Showalter -S

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K241468

Device Name Vertiwedge® Intraosseous System

#### Indications for Use (Describe)

The Vertiwedge® Intraosseous System is indicated for use in the thoracolumbar spine (TI-L5) for partial replacement of a diseased or damaged vertebral body resected or excised to replace a portion and/or restore the height of a collapsed vertebral body for the treatment of previous trauma (i.e. fracture or tumor) or degenerative spine disease. The system is to be placed within the vertebral body following an osteotomy and supplemented with autograft or allograft and to be used with its contralateral staple; should a physician choose to use fewer than the maximum number of screws, then supplemental fixation must be used to augment stability.

The Vertiwedge® Intraosseous System is intended to restore the integrity of the spinal column even in the absence of fusion in patients with advanced stage turnors involving the thoracolumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## K241468 - 510(K) SUMMARY

| Submitter's Name:              | Foundation Surgical                                                                                                                   |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address:           | 7327 E Tierra Buena Lane, Suite 101<br>Scottsdale, Arizona 85260                                                                      |
| Submitter's Telephone:         | 727-209-7436                                                                                                                          |
| Contact Person:                | Nathan Wright, MS, RAC<br>Empirical Technologies<br>1-719-351-0248<br>nwright@empiricaltech.com<br>Image: Empirical Technologies Logo |
| Date Summary was Prepared:     | May 23, 2024                                                                                                                          |
| Trade or Proprietary Name:     | Vertiwedge® Intraosseous System                                                                                                       |
| Device Classification Name:    | Spinal Vertebral Body Replacement Device                                                                                              |
| Classification & Regulation #: | Class II per 21 CFR §888.3060                                                                                                         |
| Product Code:                  | MQP                                                                                                                                   |
| Classification Panel:          | Orthopedic - Spinal Devices (DHT6B)                                                                                                   |

## Description of the Device Subject to Premarket Notification:

Vertiwedge® Intraosseous System is an intraosseous device that is used as a motion-preservation partial vertebral replacement device in the thoracolumbar spine in the motion-sparing Vertebral Body Osteotomy (VBO®) procedure. The Vertiwedge® Intraosseous device is used within a single osteotomized vertebral body, and it does not span any disc levels nor involve resection of or interference with the discs. The Vertiwedge® Intraosseous device is implanted through an anterior to psoas/oblique approach following an osteotomy of the vertebral body, in a Vertebral Body Osteotomy (VBO®) procedure The Vertiwedge® Intraosseous System consists of the intraosseous wedge (or spacer) with a lateral plate and a unique staple that provides integrated contralateral fixation. The Vertiwedge® Intraosseous System is available in a variety of heights and geometric options to accommodate variations in pathology and patient anatomy. Protrusions on the superior and inferior surfaces grip the osteotomized vedges to aid in expulsion resistance. The Vertiwedge® Intraosseous System is to be filled with autogenous bone graft and/or allogeneic bone graft material, and is to be used with the titanium alloy bone screws and cover plate to provide fixation within the single impacted vertebral body. All implant components are manufactured from Ti-6Al-4V ELI per ASTM F3001 or ASTM F136.

#### INDICATIONS FOR USE

The Vertiwedge® Intraosseous System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement of a diseased or damaged vertebral body resected or excised to replace a portion and/or restore the height of a collapsed vertebral body for the treatment of previous trauma (i.e, fracture or turnor) or degenerative spine disease. The system is to be placed within the vertebral body following an osteotomy and supplemented with autograft or allograft and to be used with its contralateral staple; should a physician choose to use fewer than the maximum number of screws, then supplemental fixation must be used to augment stability.

The Vertiwedge® Intraosseous System is intended to restore the integrity of the spinal column even in the absence of fusion in patients with advanced stage tumors involving the thoracolumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

{5}------------------------------------------------

## TECHNOLOGICAL CHARACTERISTICS

The subject and predicate devices have similar technological characteristics and the differences do not raise any new issues of safety and effectiveness. The predicates included in this submission were selected based on the best practices described in the FDA Draft Guidance document Best Practices for Selecting a Predicate Device to Support a Premarket Notification [510(k)] Submission. Specifically, the following characteristics are the same between the subject and predicates:

- Indications for Use ●
- Materials of manufacture
- Structural support mechanism
- General Design Features .
	- Non-expandable spacer o
	- Integrated Fixation without additional supplemental fixation O
	- Screw Backout Prevention Plate O
- Size Ranges
- Sterilization Methods ●

The differences between the subject and the predicates, which include 1) lateral integrated fixation without supplemental fixation rather than anterior integrated fixation without supplemental fixation, 2) coronal angulation for the spacer, and 3) the use of a contralateral staple for additional integrated fixation, do not raise any questions for safety and efficacy of the subject device since the subject device met performance testing requirements and because the coronal angulation and staple features are not novel features in FDA-cleared lumbar spacers.

| 510k    | Trade or Proprietary or Model Name         | Manufacturer             | Product Code  | Predicate  |
|---------|--------------------------------------------|--------------------------|---------------|------------|
| Number  |                                            |                          |               | Type       |
| K081849 | PILLAR SA PEEK Spacer System               | Blackstone Medical, Inc. | OVD, MQP      | Primary    |
| K183071 | VLIFT®-s Vertebral Body Replacement System | Stryker Spine            | PLR, MQP      | Additional |
| K171140 | Matrixx System                             | Nexxt Spine LLC          | ODP, MAX, MQP | Additional |
| K233359 | Dominion Expandable Corpectomy System      | Astura Medical           | MQP, PLR      | Additional |
| K220324 | AccuFit Lateral Plate System               | Precision Spine, Inc.    | KWQ           | Additional |

#### Predicate Devices

#### Performance Data

The Vertiwedge® Interosseous System has been tested in the following test modes:

- ASTM F2077 static axial compression ●
- ASTM F2077 static compression shear ●
- ASTM F2077 static torsion ●
- ASTM F2077 dynamic axial compression
- ASTM F2077 dynamic compression shear
- ASTM F2077 dynamic torsion
- ASTM F1877 particulate analysis of F2077 dynamic runouts ●
- ASTM F2267 subsidence testing
- Expulsion testing ●
- Staple pull-apart testing
- . Plate pull-apart testing

{6}------------------------------------------------

- Screw back-out testing ●
- Implantation testing .

The results of this non-clinical testing show that the strength of the Vertiwedge® Interosseous System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

## CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the Vertiwedge® Interosseous System is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K241468](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K241468)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
