K112429 · Eden Spine, LLC · MQP · Jan 26, 2012 · Orthopedic
Device Facts
Record ID
K112429
Device Name
GIZA(R) VERTEBRAL BODY REPLACEMENT
Applicant
Eden Spine, LLC
Product Code
MQP · Orthopedic
Decision Date
Jan 26, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Giza® Vertebral Body Replacement is intended for use during open surgical procedures in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (e.g. fracture). The Giza® Vertebral Body Replacement is intended to be used with supplemental internal spinal fixation systems that have been labeled for use in the thoracic and lumbar spine (i.e. posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). The use of allograft with the Giza® Vertebral Body Replacement device is optional.
Device Story
Giza® is a thoracolumbar vertebral body replacement (VBR) implant system. Designed for open surgical procedures to replace collapsed, damaged, or unstable vertebral bodies caused by tumor or trauma. Implantable components are manufactured from medical grade titanium alloy (Ti6Al4V-ELI). Available in three diameters (14mm, 21mm, 27mm), heights ranging from 18mm to 90mm, and adjustable for lordosis (2-12 degrees). Used in conjunction with supplemental internal spinal fixation systems (e.g., posterior pedicle screw/rod or anterior plate/screw systems). Optional use with allograft. Provides mechanical stability to the spine. Operated by surgeons in an OR setting.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included static and dynamic axial compression and torsion (ASTM F2077-03), subsidence (ASTM F2267-04), and expulsion testing.
Indicated for patients requiring vertebral body replacement in the thoracolumbar spine (T1-L5) due to tumor or trauma-related collapse, damage, or instability. Must be used with supplemental internal spinal fixation.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
JAN 2 6 2012
# 510(k) SUMMARY Eden Spine's Giza® Vertebral Body Replacement
Date: Contact: August 19, 2011 Guillaume Viallaneix, CEO Eden Spine, LLC. 377 Maitland Ave. Suite 1015 Altamonte Springs, FL 32701 407-900-9986
| Product Class: | Class II |
|-----------------|------------------|
| Classification: | 21 CFR §888.3060 |
| Product Codes: | MQP |
| Panel Code: | 87 |
### Name of Device and Name/Address of Sponsor
Giza® Vertebral Body Replacement Eden Spine Co, LLC. 377 Maitland Ave. Suite 1015 Altamonte Springs, FL 32701 407-900-9986
### Common or Usual Name
Spinal intervertebral body fixation orthosis
#### Device Description
The Giza® is a thoracolumbar vertebral body replacement device. The Giza is comprised of a variety of implant sizes to accommodate various patients' anatomy and pathology, and associated instrumentation. The devices are available in three diameters, 14mm, 21mm and 27mm. The Giza® device can accommodate lordosis from 2-12 degrees and from18 mm to 90mm in height. . All implantable components are manufactured from medical grade titanium alloy (Ti6Al4V-ELI).
#### Predicate Devices
The Giza® was shown to be substantially equivalent to legally marketed predicate devices. These include the VBR by Osteotech (K012254) and the Obelisc VBR by Ulrich GmbH (K060416).
#### Indications for Use
The Giza® Vertebral Body Replacement is intended for use during open surgical procedures in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (e.g. fracture). The Giza® Vertebral Body Replacement is intended to be used with supplemental internal spinal fixation systems that have been labeled for use in the thoracic and lumbar spine (i.e. posterior pedicle screw and rod systems, anterior plate systems,
{1}------------------------------------------------
and anterior screw and rod systems). The use of allograft with the Giza® Vertebral. Body Replacement device is optional.
## Technological Characteristics Performance Data
Static and dynamic axial compression and static torsion were completed following ASTM F2077-03. Static and dynamic torsion was completed per ASTM F2077-03. Subsidence was tested following ASTM F2267-04. Expulsion testing was conducted following a recognized protocol to allow comparison evaluation of intervertebral body fusion device assemblies, and characterize their resistance to expulsion. The above pre-clinical testing performed on the Giza® indicated that the Giza® is substantially equivalent to the predicate devices and is adequate for the intended use. In addition, the Giza® VBR is provided packaged according to ISO 11607 and sterile.
## Summary:
The Giza® and predicate devices have the same intended use, to provide mechanical stability by replacing a collapsed, damaged, or unstable vertebral body due to tumor or trauma (e.g. fracture). The indications for use of the Giza® are exactly the same as the predicate devices. Moreover, the device is very similar in sizes to the predicate devices. The materials used are also the same as in some predicate devices. There are no significant differences in technological characteristics compared to the predicates, and the minor differences that do exist do not raise any new types of safety or efficacy issues. Furthermore, bench testing demonstrates that these differences do not adversely impact device performance. Eden Spine concludes that the Giza® device is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three heads, representing the department's focus on health, human services, and science. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 2 6 2012
Eden Spine, LLC % Silver Pine Consulting, LLC Rich Jansen, Pharm.D. 13540 Guild Avenue Apple Valley, Minnesota 55124
Re: K112429
Trade/Device Name: Giza_Vertebral Body Replacement Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: MQP Dated: December 21, 2011 Received: December 22, 2011
Dear Dr. Jansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
## Page 2 - Rich Jansen, Pharm.D.
or any Federal statutes and regulations administered by other Federal agencies. You must or uny I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the qualify bytvelle (<=>>> (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire operitie ad 170 boutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement
#### 510(k) Number (if known): K112429
Device Name: Giza Vertebral Body Replacement
Indications for Use:
The Giza® Vertebral Body Replacement is intended for use during open surgical procedures in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (e.g. fracture). The Giza® Vertebral Body Replacement is intended to be used with supplemental internal spinal fixation systems that have been labeled for use in the thoracic and lumbar spine (i.e. posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). The use of allograft with the Giza® Vertebral Body Replacement device is optional.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Onthopedie, and Restorative Devices
KII2429 510(k) Number_