CREASPINE SUPSTANCE VERTEBRAL BODY REPLACEMENT SYSTEM LINE EXTENSION

K081564 · Creaspine · MQP · Jun 30, 2008 · Orthopedic

Device Facts

Record IDK081564
Device NameCREASPINE SUPSTANCE VERTEBRAL BODY REPLACEMENT SYSTEM LINE EXTENSION
ApplicantCreaspine
Product CodeMQP · Orthopedic
Decision DateJun 30, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SupStance VBR System, Line Extension is a vertebral body replacement system intended to replace a vertebral body. The SupStance VBR System is designed for use in the thoraco-lumbar spine (T1- L5) to replace a collapsed, damaged, or unstable vertebral body during tumor or trauma (i.e., fracture) management procedures. The SupStance VBR System is intended to be used with supplemental internal fixation systems. Anterior thoracolumbar plates and screws or pedicle screw and rod systems are among the options for the surgeon to use. The use of allograft and/or autograft with the SupStance VBR System is optional.

Device Story

The Creaspine SupStance VBR System is a spinal implant system designed to replace collapsed, damaged, or unstable vertebral bodies in the thoraco-lumbar spine (T1-L5). The device acts as a structural spacer and is intended for use in conjunction with supplemental internal fixation systems, such as anterior thoracolumbar plates and screws or pedicle screw and rod systems. Surgeons may optionally pack the device with autograft or allograft bone material. This line extension introduces additional implant cage dimensions (lengths and diameters) to the existing product line. The device is implanted by a surgeon during spinal reconstruction procedures to restore vertebral height and stability, potentially reducing pain and improving patient outcomes following trauma or tumor-related vertebral destruction.

Clinical Evidence

Bench testing only. Mechanical testing performed per FDA guidance for spinal systems, including static and dynamic compression, static and dynamic torsion, and expulsion testing.

Technological Characteristics

Spinal vertebral body replacement cage. Materials and design concepts are identical to the predicate K072537. The system is a mechanical implant used with supplemental fixation. No software, energy source, or electronic components.

Indications for Use

Indicated for patients requiring vertebral body replacement in the thoraco-lumbar spine (T1-L5) due to collapsed, damaged, or unstable vertebral bodies resulting from tumor or trauma (fracture).

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K081564 (pg. 1 of 2) # JUN 3 0 2008 ## SPECIAL 510(K) SUMMARY ## Creaspine SupStance Vertebral Body Replacement System, Line Extension | Proprietary Name: | Creaspine SupStance Vertebral Body Replacement System | |----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Spinal Vertebral Body Replacement System | | Proposed Regulatory Class: | Class II<br>Spinal intervertebral body fixation orthosis<br>21 CFR 888.3060 | | Device Product Code: | 87 MQP, Spinal Vertebral Body Replacement Device | | For Information Contact: | Marc Bernard<br>Regulatory Affairs Director | | | PTIB Xavier Arnozan<br>Avenue du Haut-Lévèque<br>33600 Pessac - France<br>Telephone +33 557 102 854<br>Fax: +33 557 102 886<br>Email: marc.bernard@creaspine.com | | Date Summary Prepared: | June 3, 2008 | ## Predicate Devices The proposed Creaspine SupStance Vertebral Body Replacement System (SupStance VBR System), Line Extension is substantially equivalent to the Creaspine SupStance VBR System subject of K072537 and the following legally marketed spinal vertebral body replacement devices: - Surgical Titanium MeshIM System (DePuy-Acromed™, Inc., K003043) . - Surgical Dynamics Mesh Cage System (United States Surgical, K003709) � ## Device Description The proposed Creaspine SupStance VBR System, Line Extension, is a modification of the original Creaspine SupStance VBR System that was the subject of K072537. This device modification has been submitted as a Special 510(k) Premarket Notification because the proposed SupStance VBR System, Line Extension, is identical in intended use and fundamental scientific technology to the parent SupStance VBR System originally described in K072537. {1}------------------------------------------------ $$\begin{pmatrix} \mathfrak{a}\mathfrak{f}^{\mathfrak{e}} \ \mathfrak{a}\mathfrak{f} \end{pmatrix} \quad \mathfrak{h}\mathfrak{f}\mathfrak{h}\mathfrak{g}\mathfrak{h}$$ The modifications made to the parent device to produce the proposed SupStance VBR System, Line Extension, are limited to the expansion of the implant (cage) dimensions available. The line extension includes the implants with the length and diameter combinations bolded in the following table: | Diameter | Ø19 | Ø22 | Ø25 | Ø28 | |------------|------|------|------|------| | Angulation | 0° | 5° | 5° | 5° | | Length | 30 | 30 | 30 | 30 | | | 32.5 | 32.5 | 32.5 | 32.5 | | | 35 | 35 | 35 | 35 | | | 37.5 | 37.5 | 37.5 | 37.5 | | | 40 | 40 | 40 | 40 | | | 42.5 | 42.5 | 42.5 | 42.5 | | | 45 | 45 | 45 | 45 | | | 47.5 | 47.5 | 47.5 | 47.5 | | | 50 | 50 | 50 | 50 | | | | | 52.5 | 52.5 | | | | | 55 | 55 | #### Intended Use The SupStance VBR System, Line Extension is a vertebral body replacement system intended to replace a vertebral body. The SupStance VBR System is designed for use in the thoraco-lumbar spine (T1- L5) to replace a collapsed, damaged, or unstable vertebral body during tumor or trauma (i.e., fracture) management procedures. The SupStance VBR System is intended to be used with supplemental internal fixation systems. Anterior thoracolumbar plates and screws or pedicle screw and rod systems are among the options for the surgeon to use. The use of allograft and/or autograft with the SupStance VBR System is optional. ### Statement of Technological Comparison The subject device shares the same intended use, basic design concepts, and materials as that of the predicate devices. Performance testing included the types of mechanical testing recommended for vertebral body replacement systems (static and dynamic compression testing, static and dynamic torsion testing, and expulsion testing) in the "FDA Guidance for Industry and FDA Staff, Spinal System 510(k)s" (issued May 3, 2004). The data collected confirms that the mechanical properties of the proposed line extension to the SupStance VBR System product line are comparable to those of the predicate devices. The information and data collected support a claim of substantial equivalence for the proposed SupStance VBR System, Line Extension to the specified predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol on the right, which is a staff entwined with snakes, representing medicine and health. To the left of the caduceus is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 PTIB Xavier Arnozan % Mr. Marc Bernard Regulatory Affairs Director Avenue du Haut-Lévèque 33600 Pessac-France JUN 3 0 2008 Re: K081564 Trade/Device Name: Creaspine SupStance Vertebral Body Replacement System Regulation Number: 21 CFR 888.3060 Regulation Names: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: MQP Dated: June 3, 2008 Received: June 4, 2008 Dear Mr. Bernard: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. Marc Bernard This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark N. Mullean Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## INDICATIONS FOR USE ## 510(k) Number (if known): K081564 #### Device Name: Creaspine SupStance Vertebral Body Replacement System, Line Extension Indications for Use: The Creaspine SupStance Vertebral Body Replacement System, Line Extension, (SupStance VBR System) is a vertebral body replacement system intended to replace a vertebral body. The SupStance VBR System is designed for use in the thoraco-lumbar spine (T1- L5) to replace a collapsed, damaged, or unstable vertebral body during tumor or trauma (i.e., fracture) management procedures. The SupStance VBR System is intended to be used with supplemental internal fixation systems. Anterior thoracolumbar plates and screws or pedicle screw and rod systems are among the options for the surgeon to use. The use of allograft and/or autograft with the SupStance VBR System is optional. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) **(Division Sign-Off)** Division of General, Restorative, and Neurological, Restorative, and Neurological Deveral, June 3, 2008 age 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...