← Product Code [MQP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP) · K062083

# SYNFIX -LR (K062083)

_Synthes Spine Co.Lp · MQP · Apr 3, 2007 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K062083

## Device Facts

- **Applicant:** Synthes Spine Co.Lp
- **Product Code:** [MQP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP.md)
- **Decision Date:** Apr 3, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The SynFix-LR is a stand-alone vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace a diseased vertebral body resected or excised during partial vertebrectomy procedures for the treatment of tumor or trauma (i.e. fracture) to achieve anterior decompression of the spinal cord and neural tissues. The interior of the spacer component of the SynFix-LR can be packed with bone (autograft or allograft). The SynFix-LR is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period.

## Device Story

SynFix-LR is a stand-alone vertebral body replacement (VBR) system for thoracolumbar spine (T1-L5). Device consists of a spacer component and an anteriorly-located plate. Surgeon inserts four screws through the plate into adjacent vertebral bodies; screws lock to plate via tapered-thread mechanism. Spacer interior can be packed with autograft or allograft bone. Device provides structural stability and restores biomechanical integrity of spinal column post-corpectomy. Used in clinical settings by surgeons to treat tumor or trauma-related vertebral disease. Benefits include anterior decompression of neural tissues and spinal column stabilization.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Combination radiolucent and radiopaque vertebral body replacement device. Features a spacer component and an anterior plate with a tapered-thread locking mechanism for screws. Available in various heights and geometries. Materials are not specified by ASTM standard in the provided text.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Synthes Vertebral Spacer ([K011037](/device/K011037.md))
- Interpore Cross GeoTM Structure ([K010530](/device/K010530.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## KOLLOBS

### 4.0 510 (k) Summary

APR - 3 2007

| Name of Firm:                                 | Synthes Spine<br>1302 Wrights Lane East<br>West Chester, PA 19380                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510 (k) Contact:                              | Bonnie Smith, RAC<br>Synthes Spine Regulatory Affairs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Trade Name:                                   | SynFixTM-LR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Common /<br>Classification Name:              | Vertebral Body Replacement                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Product<br>Code and<br>Classification: | MQP<br>21 CFR 888.3060<br>Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Predicates:                                   | K011037 – Synthes Vertebral Spacer<br>K010530 – Interpore Cross GeoTM Structure                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device Description:                           | The Synthes SynfixTM-LR is a combination radiolucent and<br>radiopaque vertebral body replacement device that provides<br>structural stability in skeletally mature individuals following<br>corpectomy. The SynfixTM-LR may be used to accommodate the<br>anatomical requirements of the space created by the corpectomy.<br>Four screws are inserted through the anteriorly-located plate into<br>the adjacent vertebral bodies. The screws lock securely to the plate<br>using a tapered-thread locking mechanism.<br>The SynfixTM-LR is available as assembled components in various<br>heights and geometries to suit individual pathology and anatomical<br>conditions. |
| Intended Use/<br>Indications for Use:         | The SynFix-LR is a stand-alone vertebral body replacement<br>device intended for use in the thoracolumbar spine (T1-L5) to<br>replace a diseased vertebral body resected or excised during<br>partial vertebrectomy procedures for the treatment of tumor or<br>trauma (i.e. fracture) to achieve anterior decompression of the<br>spinal cord and neural tissues. The interior of the spacer<br>component of the SynFix-LR can be packed with bone<br>(autograft or allograft).<br>The SynFix-LR is designed to restore the biomechanical<br>integrity of the anterior, middle, and posterior spinal column<br>even in the absence of fusion for a prolonged period.         |

page 1 of 1

Revised Indications for SynFix™-LR Spacer – K062083
Response to Request for Additional Information

CONFIDENTIAL

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Synthes Spine % Ms. Bonnie Smith Senior Regulatory Affairs Specialist 1302 Wrights Lane East West Chester, Pennsylvania 19380

APR - 3 2007

K062083 Trade/Device Name: SynFix™-LR Spacer Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertabral body fixation orthosis Regulatory Class: Class II Product Code: MQP Dated: February 1, 2007 Received: February 2, 2007

Dear Ms. Smith:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

## Page 2 – Ms. Bonnie Smith

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark A. Millam

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# k062083

#### 5.0 Indications for Use Statement

Device Name: SynFix™-LR

Indications for Use:

The SynFix-LR is a stand-alone vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace a diseased vertebral body resected or excised during partial vertebrectomy procedures for the treatment of tumor or trauma (i.e. fracture) to achieve anterior decompression of the spinal cord and neural tissues. The interior of the spacer component of the SynFix-LR can be packed with bone (autograft or allograft).

The SynFix-LR is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Murk A. Millican

Division of General, Restorative, and Neurological Devices

510(k) Number K062083

Day 1 of 1

Revised Indications for SynFixTM-LR Spacer - K062083 Response to Request for Additional Information

CONFIDENTIAL

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K062083](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K062083)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
