← Product Code [MQP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP) · K061401

# INTERPLATE VBR SYSTEM (K061401)

_Rsb Spine, LLC · MQP · Aug 18, 2006 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K061401

## Device Facts

- **Applicant:** Rsb Spine, LLC
- **Product Code:** [MQP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP.md)
- **Decision Date:** Aug 18, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The InterPlate™ VBR device is indicated for the replacement of a complete or partial vertebrectomy, when used with a bone graft to facilitate fusion. It is designed to restore biomechanical integrity of the thoracic and lumber spine, from T1 to L5, which has been damaged due to a collapsed or unstable vertebral body resulting from a tumor or traumatic injury. The InterPlate™ VBR device is intended to be used with a supplemental internal fixation system.

## Device Story

InterPlate™ VBR System is a spinal intervertebral body fixation orthosis; consists of plates, bone screws, and screw covers. Used to replace complete or partial vertebrectomy; restores biomechanical integrity of thoracic and lumbar spine (T1-L5) damaged by tumor or trauma. Requires bone graft for fusion; must be used with supplemental internal fixation system. Operated by surgeons in clinical/OR settings. Provides structural support to vertebral column; facilitates spinal fusion; benefits patients by stabilizing spine after vertebral body loss.

## Clinical Evidence

No clinical data provided; substantial equivalence based on design and material characteristics.

## Technological Characteristics

Components manufactured from Ti-6Al-4V titanium alloy (ASTM F136). System includes plates, bone screws, and screw covers. Mechanical fixation device; no software or electronic components.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- MC+ Partial VBR (LDR Spine USA)
- SynFix™ - LR (Synthes USA)
- Stalif™ TT (Surgicraft, LTD)
- Telescopic Plate Spacer (Interpore Cross International)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

AUG 1 8 2006

## 510(k) Summary

| Sponsor:                    | RSB Spine, LLC<br>3030 Superior Ave., Suite 703<br>Cleveland, OH 44114                                                                                                                                                                                                                                                                                                                                                                                                                   |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact<br>Person:          | James M. Moran, D. Eng.<br>Vice President of Engineering and Chief Technical Officer                                                                                                                                                                                                                                                                                                                                                                                                     |
| Proprietary<br>Trade Name:  | InterPlate™ VBR System                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification<br>Name      | 888.3060 - Spinal Intervertebral Body Fixation Orthosis                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Product<br>Code:     | MQP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device<br>Description:      | The InterPlate™ VBR consists of plates, bone screws and screw covers. Various<br>plate sizes are available to accommodate individual patient anatomy and graft<br>material size. Screw covers are individually matched to the plate size.                                                                                                                                                                                                                                                |
| Intended Use:               | The InterPlate™ VBR device is indicated for the replacement of a complete or<br>partial vertebrectomy, when used with a bone graft to facilitate fusion. It is<br>designed to restore biomechanical integrity of the thoracic and lumber spine, from<br>T1 to L5, which has been damaged due to a collapsed or unstable vertebral body<br>resulting from a tumor or traumatic injury. The InterPlate™ VBR device is<br>intended to be used with a supplemental internal fixation system. |
| Materials:                  | The InterPlate™ VBR System components are manufactured from Ti-6Al-4V<br>titanium alloy (ASTM F136).                                                                                                                                                                                                                                                                                                                                                                                     |
| Substantial<br>Equivalence: | Documentation was provided which demonstrated the InterPlate™ VBR<br>System to be substantially equivalent to previously cleared devices including<br>the MC+ Partial VBR (LDR Spine USA), the SynFix™ - LR (Synthes USA),<br>the Stalif™ TT (Surgicraft, LTD) and the Telescopic Plate Spacer (Interpore<br>Cross International). The substantial equivalence is based upon equivalence in<br>intended use, indications, anatomic sites, performance and material of<br>manufacture.    |

·

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a caduceus-like symbol with three parallel lines representing the branches of government, and a stylized eagle with its wings spread, symbolizing protection and service.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## AUG 1 8 2006

RSB Spine, LLC % Karen E. Warden, Ph.D. Representative/Consultant 8202 Sherman Road Chesterland, Ohio 44026

Re: K061401

Trade/Device Name: InterPlate™ VBR System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: MQP Dated: July 26, 2006 Received: July 31, 2006

Dear Dr. Warden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Karen E. Warden, Ph.D.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Barbara Buck MD

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number: K061 401

## Device Name: InterPlate ™ Vertebral Body Replacement (VBR) System

Indications for Use:

The InterPlate™ VBR device is indicated for the replacement of a complete or partial vertebrectomy, when used with a bone graft to facilitate fusion. It is designed to restore biomechanical integrity of the thoracic and lumbar spine, from T1 to L5, which has been damaged due to a collapsed or unstable vertebral body resulting from a tumor or traumatic injury. The InterPlate™ VBR device is intended to be used with a supplemental internal fixation system.

Prescription Use X (Per 21 CFR 801.109) OR

Over-the-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Barbara Vreeland

vision of General, nes and Neurological Devices

**510(k) Number** K061401

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K061401](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K061401)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
