← Product Code [MQP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP) · K052950

# THE ZIMMER TMT TRABECULAR METAL VERTEBRAL BODY REPLACEMENT SYSTEM (K052950)

_Zimmer Trabecular · MQP · Nov 16, 2005 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K052950

## Device Facts

- **Applicant:** Zimmer Trabecular
- **Product Code:** [MQP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP.md)
- **Decision Date:** Nov 16, 2005
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Trabecular Metal Vertebral Body Replacement System is a vertebral body replacement device intended for use in the thoracolumbar spine (T1 - L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The Trabecular Metal Vertebral Body Replacement is intended for use with supplemental internal spinal fixation systems. The Trabecular Metal Vertebral Body Replacement may be used with bone graft.

## Device Story

Trabecular Metal Vertebral Body Replacement (VBR) System is a porous tantalum implant designed for anterior spinal surgery. It replaces a single diseased or damaged vertebral body and adjacent disc in the thoracic and lumbar spine. The device features a rippled surface pattern on superior and inferior faces to enhance stability. It is used by surgeons in an operating room setting to restore spinal height and tension. The device requires supplemental internal spinal fixation to function. It provides structural support to the spinal column, potentially benefiting patients by stabilizing the spine following tumor resection or traumatic fracture.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included static and dynamic compression, static and dynamic torsion, expulsion, and abrasion testing. Biocompatibility data supported the use of the porous tantalum material.

## Technological Characteristics

Material: Porous Tantalum (Trabecular Metal). Geometry: Various cross-sections and heights with rippled superior/inferior surfaces. Principle: Mechanical vertebral body replacement. Connectivity: None. Sterilization: Not specified.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Trabecular Metal Vertebral Body Replacement System ([K010378](/device/K010378.md))
- Trabecular Metal Vertebral Body Replacement System ([K051196](/device/K051196.md))
- Trabecular Metal Vertebral Body Replacement System ([K051978](/device/K051978.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K052950

## 510(k) SUMMARY Trabecular Metal Vertebral Body Replacement System

| Trabecular Metal Vertebral Body Replacement System Page 1 of 2 |                                                                                                                                                                                                                                              |
|----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name<br>And Address:                                 | Zimmer Trabecular Metal Technology, Inc.<br>80 Commerce Drive<br>Allendale, New Jersey 07401-1600                                                                                                                                            |
| Contact Person:                                                | Marci Halevi                                                                                                                                                                                                                                 |
| Phone Number:                                                  | (201) 818-1800 ext. 507                                                                                                                                                                                                                      |
| Fax Number:                                                    | (973) 829-0825                                                                                                                                                                                                                               |
| Date Prepared:                                                 | October 17, 2005                                                                                                                                                                                                                             |
| Device Trade Name:                                             | Trabecular Metal Vertebral Body Replacement System                                                                                                                                                                                           |
| Device Common Name:                                            | Vertebral Body Replacement Device                                                                                                                                                                                                            |
| Classification Number<br>and Name:                             | 21 CFR § 888.3060.<br>Spinal Intervertebral Body Fixation Orthosis                                                                                                                                                                           |
| Device<br>Description:                                         | The Trabecular Metal Vertebral Body Replacement System is<br>designed to be used as a replacement for a single diseased or<br>damaged vertebral body and the adjacent disc when spinal surgery<br>through an anterior approach is indicated. |

The Trabecular Metal Vertebral Body Replacement (VBR) System is comprised wholly of Trabecular Metal Porous Tantalum. The VBR is available to accommodate replacement of a vertebral body in the thoracic and lumbar region of the spine. The device is available in a variety of cross sections and heights to properly tension the spine.

The superior and inferior surfaces of the device have a pattern of ripples to provide increased stability.

- The Zimmer TMT Trabecular Metal Vertebral Body Replacement Indications for Use: System is comprised of vertebral body replacement devices intended for use in the thoracolumbar spine (T1 - L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The Trabecular Metal Vertebral Body Replacement is intended for use with supplemental internal spinal fixation systems. The Trabecular Metal Vertebral Body Replacement may be used with bone graft.
Device Technoloqical The device is unique in comparison to predicates for this Characteristics and indication with regard to its geometry. The material has been used in cited predicates for other applications. The Zimmer TMT Comparison to Predicate Device: Vertebral Body Replacement System is substantially equivalent to previously cleared components of this same system (K010378, K051196 and K051978).

{1}------------------------------------------------

- Predicate devices that are a part of the Trabecular Metal Performance Data: Vertebral Body Replacement System were tested per applicable standards (reference K010378 and current submission). Biocompatibility data was provided to support the material's use. Performance testing was provided to support equivalent mechanical behavior to the predicate devices. The results demonstrated that the device will perform as intended and is equivalent to the cited predicate devices. Test data was provided regarding:
	- Expulsion, .
	- Static compression, .
	- Dynamic compression, .
	- Static torsion, .
	- Dynamic torsion, and
	- . Abrasion.

Conclusion:

TMT Trabecular Metal The Zimmer Vertebral Body Replacement System is substantially equivalent to the predicate devices identified in this premarket notification.

> K052950 ge 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name in a circular arrangement around the perimeter. In the center of the seal is a stylized representation of a human figure, composed of three curved lines. The figure is meant to symbolize the department's mission of protecting the health of all Americans and providing essential human services.

NOV 1 6 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Marci Halevi Manager of Regulatory Affairs Zimmer Trabecular Metal Technology, Inc. 80 Commerce Drive Allendale, New Jersey 07401

Re: K052950

Trade/Device Name: Trabecular Metal Vertebral Body Replacement System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: MQP Dated: October 19, 2005 Received: October 20, 2005

Dear Ms. Halevi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Marci Halevi

This letter will allow you to begin marketing your device as described in your Section 510(k) I ms lotter will and w yourse FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you dome office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general information on Jon Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

signature

Mark N. Melkerson પ્રેમ Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): _ K05 2950

Device Name: The Zimmer TMT Trabecular Metal Vertebral Body Replacement System

Indications for Use:

The Trabecular Metal Vertebral Body Replacement System is a vertebral body replacement device intended for use in the thoracolumbar spine (T1 - L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The Trabecular Metal Vertebral Body Replacement is intended for use with supplemental internal spinal fixation systems. The Trabecular Metal Vertebral Body Replacement may be used with bone graft.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

(Division Sign-Off Division of General, Rest rative. and Neurological Devices

510(k) Number_KOSQ950

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K052950](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MQP/K052950)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
