NUVASIVE COROENT SYSTEM

K043405 · Nuvasive, Inc. · MQP · Feb 7, 2005 · Orthopedic

Device Facts

Record IDK043405
Device NameNUVASIVE COROENT SYSTEM
ApplicantNuvasive, Inc.
Product CodeMQP · Orthopedic
Decision DateFeb 7, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Intended Use

The NuVasive CoRoent System is a partial vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased vertebral body resected or excised for the treatment of a collapsed, damaged, or unstable vertebral body(s) due to tumor or trauma and to achieve decompression of the spinal cord and neural tissues. The System is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft material may be used at the surgeon's discretion.

Device Story

The NuVasive CoRoent System is an implantable PEEK vertebral body replacement device; used in the thoracolumbar spine (T1-L5) to replace diseased, collapsed, or unstable vertebral bodies resulting from tumor or trauma. The device provides structural support to achieve decompression of spinal cord and neural tissues. It is available in various shapes and sizes to accommodate patient-specific anatomy and pathology. The system is intended for use in conjunction with FDA-cleared supplemental internal spinal fixation systems; surgeons may optionally pack the device with autograft or allograft material. The device is implanted by a surgeon in a clinical/surgical setting.

Clinical Evidence

No clinical data. Bench testing only.

Technological Characteristics

Implantable vertebral body replacement device constructed from PEEK. Available in various shapes and sizes. Designed for use with supplemental internal spinal fixation systems.

Indications for Use

Indicated for patients requiring vertebral body replacement in the thoracolumbar spine (T1-L5) due to tumor or trauma resulting in collapsed, damaged, or unstable vertebral bodies; requires use with supplemental internal spinal fixation.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## VII. 510(k) Summary K043405 Page 1 of 2 In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title of the Code of n decorantee with the 807 (21 CFR §807), and in particular §807.92, the following summary of safety and effectiveness information is provided: ### A. Submitted by Laetitia Cousin Director of Regulatory Affairs and Quality Assurance NuVasive, Incorporated 4545 Towne Centre Court San Diego, CA 92121 Telephone: (858) 909-1868 Date Prepared: February 1, 2005. ### B. Device Name | Trade or Proprietary Name: | NuVasive CoRoent System | |----------------------------|-----------------------------------| | Common or Usual Name: | Vertebral Body Replacement Device | | Classification Name: | Vertebral Body Replacement Device | #### C. Predicate Devices The subject device is substantially equivalent to similar previously cleared devices. ## D. Device Description The NuVasive CoRoent System is an implantable PEEK vertebral body replacement device indicated for use in the thoracic and lumbar spine (i.e., T1 to L5) to replace a diseased vertebral body resected or excised for the treatment of a collapsed, damaged, or unstable vertebral body(s) due to tumor or trauma and to achieve decompression of the spinal cord and neural tissues. The device is available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient. {1}------------------------------------------------ | | | | | - BA-t-e- | I<br>----------- | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | == =<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 1 | 1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | -A-B-A-FLA-FIND-A<br>1 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | 3<br>Ma | 1 | | | | œ<br>1 | The NuVasive CoRoent System is a partial vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased vertebral body resected or excised for the treatment of a collapsed, damaged, or unstable vertebral body(s) due to tumor or trauma and to achieve decompression of the spinal cord and neural tissues. The System is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft material may be used at the surgeon's discretion. # F. Comparison to Predicate Devices As was established in this submission, the subject device is substantially equivalent to the following predicate devices: | K033517 | Spinal Concepts Cadence™ | Spinal concepts, Inc. | |---------|--------------------------|--------------------------| | K011037 | Vertebral Spacer | Synthes | | K032476 | NuVasive Mesh | NuVasive, Inc. | | K041939 | Blackstone VBR System | Blackstone Medical, Inc. | Engineering drawings, labeling, and mechanical testing have demonstrated that the subject device is substantially equivalent, if not identical, to its predicate devices in terms of design, materials of composition, indications for use, and such other characteristics as may be associated with the manufacture of any medical device. # G. Summary of Non-Clinical Tests Mechanical testing was presented. # H. Summary of Clinical Tests (Not Applicable). {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure with three wing-like shapes. FEB - 7 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Laetitia Cousin Director of Regulatory Affairs and Quality Assurance NuVasive, Inc. 4545 Towne Centre Court San Diego, California 92121 Re: K043405 Trade/Device Name: NuVasive CoRoent System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: MQP Dated: December 9, 2004 Received: December 10, 2004 Dear Ms. Cousin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Laetitia Cousin This letter will allow you to begin marketing your device as described in your Section 510(k) I mb lotel will and in your a FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rr you abon't the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark McMullen Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## V. Draft Labeling ### A. Indications for Use 510(k) Number (if known): __ K0 4 3405 NuVasive CoRoent System Device Name: Indications for Use: The NuVasive CoRoent System is a partial vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased vertebral body resected or excised for the treatment of a collapsed, damaged, or unstable vertebral body(s) due to tumor or trauma and to achieve decompression of the spinal cord and neural tissues. The System is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft material may be used at the surgeon's discretion. Mark N. Milken al. Resterative, and Neurological Devices **510(k) Number** K043405 Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use __ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...