OPEN RADIOTRANSPARENT INTERSOMATIC IMPLANT (ROI)

K043349 · Ldr Spine USA · MQP · Jun 29, 2005 · Orthopedic

Device Facts

Record IDK043349
Device NameOPEN RADIOTRANSPARENT INTERSOMATIC IMPLANT (ROI)
ApplicantLdr Spine USA
Product CodeMQP · Orthopedic
Decision DateJun 29, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ROI is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a partial vertebral body replacement in the thoracolumbar spine (from T1 to L5) and is intended for use with supplemental internal fixation. The ROI may be implanted singularly or in pairs. Supplemental internal fixation is required to properly utilize this system.

Device Story

Radiotransparent Open Implant (ROI) is a modular vertebral body replacement system; consists of hollow, rectangular, flat or wedge-shaped PEEK blocks with surface teeth for stability. Tantalum wire markers provide radiographic visualization. Used by surgeons in the thoracolumbar spine to replace resected/excised vertebral bodies due to tumor or trauma. Implanted singularly or in pairs; requires supplemental internal fixation. Radiolucent PEEK material facilitates post-operative assessment of fusion. Supplied sterile via gamma radiation; reusable instruments require steam sterilization.

Clinical Evidence

Bench testing only. Fatigue and static testing performed according to accepted engineering and scientific principles. Results demonstrate performance equivalent to predicate devices.

Technological Characteristics

Materials: PEEK (ASTM F-2026) with Tantalum markers (ASTM F-560). Modular cage design; hollow chamber; rectangular block with surface teeth. Supplied sterile (gamma radiation). Reusable instruments require steam autoclave sterilization (134°C, 18 mins).

Indications for Use

Indicated for partial vertebral body replacement in thoracolumbar spine (T1-L5) following resection or excision due to tumor or trauma/fracture. Requires supplemental internal fixation. May be implanted singularly or in pairs.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | IDD Car | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | 510(k)<br>Summarv | Spine USA | | | JUNE 9, 2005 | | Premarket Notification, Section 510(k) | | Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92 #### 1. Device Name: Radiotransparent Open Implant (ROI) - Partial Vertebral Body Trade Name: Replacement (PVBR) Common Vertebral body replacement Name(s): Classification Name(s): Vertebral body replacement (MQP) #### Establishment Name & Registration Number: 2. | Name: | LDR Spine USA | |---------|---------------| | Number: | 3004903783 | #### 3. Classification(s): 21 CFR § Sec. 888.3060 Spinal intervertebral body fixation orthosis. (a) Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of vertebral bodies. An eve-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct "sway back," scoliosis (lateral curvature of the spine), or other conditions. Device Class: Class II for the requested indications Classification Panel: Orthopaedic and Rehabilitation Devices Panel Product Code(s): MQP # 4. Equivalent Predicate Device: LDR Spine USA proposes that the ROI is substantially equivalent to the following: - -EBI CAS Spine Spacer System- K042268 - -Signus PEEK Tetris™ Spinal Implant - K031757 - Signus Curved PEEK Tetris™ Spinal Implant K041888 - Equivalence is demonstrated in the design, material composition, surgical technique and intended use. #### ર્ડ Device Description: The ROI vertebral implants consists of a series of flat and wedge shaped implants. The device is used singly or in pairs. The device is offered in nine different configurations to better approximate the anatomical variation observed in different vertebral levels and/or patient anatomy. The LDR Spine USA, ROI Partial Vertebral Body Replacement System is comprised of a variety of components fabricated and manufactured from Polyetheretherketone (PEEK) as described by ASTM F-2026. This material is utilized due to its radiolucent properties, which aid the surgeon in determining if fusion in the operative site has occurred. The LDR Spine ROI Partial Vertebral Body Replacement System is a two piece, hollow chamber, rectangular block. The superior and inferior surfaces of the construct have a pattern of teeth to provide increased stability and to help prevent movement of the device. Tantalum wire markers (ASTM F-560) are inserted into components to give surgeons a visual aid in determining the location of the implant, both inter and post-operatively. The device comes in various sizes and is offered in both straight and tapered styles. ROI.doc {1}------------------------------------------------ Materials: all implants are made from implant grade PolyEtherEther-Ketone polymer (PEEK) with tantalum alloy position markers as indicated in the table below: | PEEK<br>Optima LT | USP Class VI<br>ASTM F-2026 | ISO 10993 | |-------------------|-----------------------------|------------| | Tantalum | ASTM F-560 | ISO 5832-3 | Indications for Use. The ROI is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a partial vertebral body replacement in the thoracolumbar spine (from TI to LS) and is intended for use with supplemental internal fixation. The ROI may be implanted singularly or in pairs. Supplemental internal fixation is required to properly utilize this system. Testing Summary. Fatigue and static testing is complete. Samples were tested according to accepted engineering and scientific principals. Test results demonstrate that the system can be expected to perform in a manner equivalent to the predicate devices. #### Applicant Name & Address: 6. LDR Spine USA 4030 W. Braker Ln., Ste. 360 Austin, TX. 78759 Office: (512) 344-3333 Fax: (512) 344-3350 # 7. Company Contact: Mr. Edward E. Newton Dir. Reg. & Clinical Affairs LDR Spine USA 4030 W. Braker Ln., Ste. 360 Austin, TX. 78759 Office: (512) 344-3316 Fax: (512) 344-3350 # 8. Submission Correspondent: Mr. Brian Burkinshaw Dir. Innovation & Technology LDR Spine USA 4030 W. Braker Ln., Ste. 360 Austin, TX. 78759 Office: (512) 344-3304 Fax: (512) 344-3350 # 9. Performance Standards: United States Food and Drug Administration mandated performance standards for this device do not exist. Various voluntary performance standards are utilized. Voluntary standards utilized include ASTNA LIDR Spine, USA Standard Operating Procedures, vendor & process certification and qualification procedures, Quality Systems Regulations, ISO materials standards and ISO 13485 series quality regulations. LDR Spine USA also meets appropriate general controls authorized under Sections 501, 502, 510, 516, 518, 519, and 520 of the Food, Drug, and Cosmetic Act. ROl.doc {2}------------------------------------------------ # Storage, Packaging & Sterilization Information: 10. The implantable portions of the ROI are supplied "STERILE". The sterilization process is radiation and the selected protocol has been validated. The minimum Sterility Assurance Level (SAL) of at least 10 . The instruments are supplied non-sterile and must be cleaned and sterilized prior to first use and each The recommended sterilization process for the instruments is high temperature steam subsequent use. autoclave sterilization. The referenced sterilization cycle produces a Sterility Assurance Level (SAL) of at least 10-6. # The validated cycle is: Method: Steam Cycle: Gravity Temperature: 270°F (134°C) Exposure Time: 18 minutes All packages containing implants or instruments should be intact upon receipt. Damaged packaging may indicate the presence of unsafe product. If the package or product is damaged, the product should not be used and should be returned. Product must be handled, stored and opened in such a way that it is protected from inadvertent damage or contamination. When used, the product must be placed into use following cleaning, sterilization and accepted surgical sterile technique. | FEATURE | ROI | CAS Spine Spacer System | PEEK and Curved PEEK<br>Tetris™ Spinal Implant | SE? | |-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----| | Indications for<br>Use: | The ROI is indicated for use to<br>replace a vertebral body that has<br>been resected or excised due to<br>tumor or trauma/fracture. The<br>device is intended for use as a<br>partial vertebral body replacement<br>in the thoraco-lumbar spine (from<br>TI to L5) and is intended for use<br>with supplemental internal fixation.<br>The ROI may be implanted<br>sinularly or in pairs. Supplemental<br>fixation is required to properly<br>utilize this system | Same as ROI | Indicated for use to replace a vertebral<br>body that has been resected or excised<br>due to tumor or trauma/fracture. The<br>device is intended for use as a<br>vertebral body replacement in the<br>thoracolumbar spine (from TI to L5)<br>and is intended for use with<br>supplemental internal fixation. | YES | | Design: | Modular Cage | Cage | Same as CAS SSS | YES | | Supplied<br>Sterile: | Yes - Gamma radiation | Yes | Yes - Gamma radiation | YES | | Material: | PEEK w/ tantalum markers | Titanium | PEEK w/ titanium markers | YES | | Instruments: | Specialized reusable instruments are<br>required. Instruments must be<br>cleaned and sterilized prior to 1st use<br>and each subsequent use. | Same | Same | YES | | How used: | Singularly or in pairs | Same | Singularly | YES | | K Number: | K043349 (under review) | K042268 | K031757 - K041888 | YES | | Manufacturer: | LDR Medical | EBI | Signus Medizintechnik, Gmbh | YES | | Product Code: | MOP | MOP | MOP | YES | - Summary Comparison Table: 11. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with text around the perimeter. Inside the circle is a stylized eagle with three lines representing its wings or feathers. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Brian Burkinshaw Director, Innovation and Technology Solutions LDR Spine USA Incorporated 4030 West Braker Lane, Suite 360 Austin, Texas 78759 Re: K043349 Trade/Device Name: Radiotransparent Open Implant (ROI)- Partial Vertebral Body Replacement (PVBR) Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: MQP Dated: June 9, 2005 Received: June 10, 2005 Dear Mr. Burkinshaw: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. JUN 2 9 2005 {4}------------------------------------------------ Page 2 – Mr. Brian Burkinshaw This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Stupt Rhodes Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 643349 Page 1 of 1 K043349 510(k) Number: Radiotransparent Open Implant (ROI) – Partial Vertebral Body Device Name(s): Replacement (PVBR) # Indications for Use: The ROI is indicated for use to replace a vertebral body that has been resected or The KOr is included for as to replace is intended for use as a excised due to tunnor of thatha inche thoracolumbar spine (from TV to LS) and partial vertebrar oddy replacemental internal fixation. The ROI may be implanted singularly or in pairs. Supplemental internal fixation is required to properly utilize this system. Over-The-Counter Use OR × Prescription Use_ PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IN NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE) Stupt Rhodes (Division Sign-Off) Division of General, Restorative, and Neurological Devices KOY3349 510(k) Number_ (Per 21 CFR 801.109) (Optional format 1-2-96) . P. 1 ot 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...