K032476 · Nuvasive, Inc. · MQP · Nov 13, 2003 · Orthopedic
Device Facts
Record ID
K032476
Device Name
NUVASIVE MESH
Applicant
Nuvasive, Inc.
Product Code
MQP · Orthopedic
Decision Date
Nov 13, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The NuVasive Mesh is a vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased or damaged vertebral body caused by tumor or fracture, to restore height of a collapsed vertebral body, and to achieve decompression of the spinal cord and neural tissues. The NuVasive® Mesh is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft material may be used at the surgeon's discretion.
Device Story
Implantable titanium vertebral body replacement device; used in thoracic and lumbar spine (T1-L5). Features hollow core for bone graft packing; rhombic pattern wall holes; integrated teeth/spikes on ends for endplate gripping and expulsion resistance. Available in various shapes/sizes. Implanted by surgeons during spinal procedures to replace diseased/damaged vertebrae, restore height, and decompress neural tissues. Requires use with supplemental internal spinal fixation systems. Benefits include structural support and fusion promotion.
Clinical Evidence
No clinical data. Mechanical testing only.
Technological Characteristics
Titanium vertebral body replacement device; hollow core for bone graft; rhombic wall pattern; integrated endplate teeth. Standalone implant; non-powered; non-software.
Indications for Use
Indicated for patients requiring vertebral body replacement in the thoracolumbar spine (T1-L5) due to tumor or fracture; intended for use with supplemental internal spinal fixation systems.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
K061380 — RETAIN RADIOLUCENT SPACER · Globus Medical, Inc. · Jun 8, 2006
K050449 — QUANTUM VERTEBRAL BODY REPLACEMENT · Quantum Orthopedics, Inc. · Jul 7, 2005
K060665 — XPAND RADIOLUCENT CORPECTOMY SPACER · Globus Medical, Inc. · May 10, 2006
K032812 — ENDOSKELETON TA VERTEBRAL BODY REPLACEMENT · Orthovita, Inc. · Dec 8, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
## VII. 510(k) Summary
032476 ge I of 2
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular §807.92, the following summary of information is provided:
#### Submitted by: A.
Laetitia Bernard Manager of Regulatory Affairs and Quality Assurance NuVasive, Incorporated 10065 Old Grove Road San Diego, California 92131 Telephone: (858) 527-1918 Telefacsimile: (858) 271-7101
#### B. Device Name
Trade Name: NuVasive Mesh Common or Usual Name: Vertebral Body Replacement Device Vertebral Body Replacement Device Classification Name:
#### C. Predicate Devices
The subject NuVasive Mesh is substantially equivalent to the NuVasive Mesh currently manufactured and distributed commercially in the U.S. by NuVasive, Inc.
### D. Device Description
The NuVasive Mesh is an implantable titanium vertebral body replacement device indicated for use in the thoracic and lumbar spine (i.e., T1 to L5) to replace a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body.
A plurality of rhombic pattern holes are built in the wall, transverse to the longitudinal axis. The hollow core allows for packing of bone grafting materials to help promote a solid fusion. A ring of small spikes or teeth on each end of the device serves to grip the endplates of the adjacent vertebrae to resist expulsion.
The device is available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient.
{1}------------------------------------------------
### E. Intended Use
32476
The NuVasive Mesh is a vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased or damaged vertebral body caused by tumor or fracture, to restore height of a collapsed vertebral body, and to achieve decompression of the spinal cord and neural tissues. The NuVasive® Mesh is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft material may be used at the surgeon's discretion.
#### F. Comparison to Predicate Devices
As was established in this submission, the subject device is substantially equivalent to the current NuVasive Mesh cleared by the agency for commercial distribution in the United States.
Engineering drawings, labeling, and mechanical testing have demonstrated that the subject device is substantially equivalent, if not identical, to its predicate devices in terms of design, materials of composition, indications for use, and such other characteristics as may be associated with the manufacture of any medical device.
Due to this equivalency, the device raises no new safety or effectiveness issues.
#### G. Summary of Design Control Activities
Design control activities employed to control the development of the modification to the NuVasive Mesh included:
- a comprehensive Risk Analysis to identify potential risks and failures associated with ● operation of the device, any mitigations incorporated to reduce or eliminate those risks and failures, and an assessment of residual risk;
- a comprehensive program of verification and validation activities demonstrating that . acceptance criteria were met, and that design output satisfied design input.
#### H. Summary of Clinical Tests
(Not applicable.)
### H. Summary of Non-Clinical Tests
Mechanical testing was presented.
#### J. Conclusions
The subject device is substantially equivalent to the currently marketed predicate device, and its development has been adequately and appropriately conducted and validated under a comprehensive design control program complying with Title 21 CFR, §820.30.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized depiction of a human profile with three faces, representing the department's focus on health and human well-being. The seal is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 3 2003
Ms. Laetitia Bernard Manager of Regulatory Affairs & Quality Assurance NuVasive, Incorporated 10065 Old Grove Road San Diego, California 92131
Re: K032476
Trade/Device Name: NuVasive Mesh Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: MQP Dated: October 31, 2003 Received: November 3, 2003
Dear Ms. Bernard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Laetitia Bernard
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Mark N Millerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
NuVasive®, Inc.
# Page 1 of
#### D. Indications for Use Statement
510(k) Number (if known): K032476
Device Name: NuVasive Mesh
Indications for Use:
The NuVasive® Mesh is a vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased or damaged vertebral body caused by tumor or fracture, to restore height of a collapsed vertebral body, and to achieve decompression of the spinal cord and neural tissues. The NuVasive® Mesh is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft material may be used at the surgeon's discretion.
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use -(Per 21 CFR 801.109) OR
Over-The-Counter Use
R. Mark N. Milliken
on Sig A storative Division of Ge and Neurologic
510(k) Number: K032976
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.