K151458 · Double Medical Technology, Inc. · MNI · Aug 22, 2016 · Orthopedic
Device Facts
Record ID
K151458
Device Name
Double Medical Universal Spine System
Applicant
Double Medical Technology, Inc.
Product Code
MNI · Orthopedic
Decision Date
Aug 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Double Medical Universal Spine System is intended for posterior, non-cervical, pedicle fixation for the following indications: severe spondylolisthesis (grade 3 or 4) of the L5-S1 vertebrae; trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudarthrosis, and failed previous fusion. The device is to be used in skeletally mature patients, and for stabilization of the spine as an adjunct to fusion with bone graft. The levels of fixation are T8 - S1.
Device Story
Pedicle screw spinal system consisting of polyaxial and monoaxial pedicle screws, rods, transconnectors, and nuts; used for posterior, non-cervical spinal stabilization (T8-S1) as an adjunct to fusion. Implanted by surgeons during spinal surgery; provides mechanical support and immobilization of vertebral segments. Benefits patients by stabilizing the spine to facilitate fusion in cases of trauma, deformity, or degenerative conditions.
Clinical Evidence
Bench testing only. Testing performed per ASTM F1717-14 (static/dynamic compression bending, static torsion) and ASTM F1798-13 (axial gripping capacity) to verify design specifications and mechanical equivalence to the predicate.
Technological Characteristics
Components: pedicle screws, rods, transconnectors, nuts. Material: Ti-6Al-4V ELI per ASTM F 136. Design: Polyaxial and monoaxial pedicle screws. Energy source: None (mechanical implant). Connectivity: None. Sterilization: Non-sterile, user-sterilized.
Indications for Use
Indicated for skeletally mature patients requiring posterior, non-cervical, pedicle fixation (T8-S1) as an adjunct to fusion with bone graft for severe spondylolisthesis (grade 3 or 4) at L5-S1, trauma (fracture/dislocation), spinal stenosis, spinal curvatures (scoliosis, kyphosis, lordosis), tumor, pseudarthrosis, or failed previous fusion.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
K131785 — PAGODA PEDICLE SCREW SYSTEM · Ortho Development Corp. · Aug 20, 2013
K160820 — PressON Pro Spinal Fixation System · Nexus Spine, LLC · Jun 28, 2016
K230063 — G Surgical Marksman Spinal Deformity System · G Surgical, LLC · Feb 9, 2023
K043484 — LANX SPINAL FIXATION SYSTEM · Lanx, LLC · May 16, 2005
K201568 — Calvary Spine Pedicle Screw System · Calvary Spine, LLC · Nov 12, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and are set against a white background. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 22, 2016
Double Medical Technology Incorporated Mr. Da Zeng No. 18, Shanbianhong East Road, Haicang District Xiamen Fujian, 361026 CHINA
Re: K151458
Trade/Device Name: Double Medical Universal Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH Dated: July 16, 2016 Received: July 20, 2016
Dear Mr. Zeng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Vincent J. Devlin -S
for
- Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K151458
#### Device Name Double Medical Universal Spine System
#### Indications for Use (Describe)
Double Medical Universal Spine System is intended for posterior, non-cervical, pedicle fixation for the following indications: severe spondylolisthesis (grade 3 or 4) of the L5-S1 vertebrae; trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudarthrosis, and failed previous fusion. The device is to be used in skeletally mature patients, and for stabilization of the spine as an adjunct to fusion with bone graft. The levels of fixation are T8 - S1.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <input type="checkbox"/> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 3 510(k) Summary
This 510(k) summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number: K151458
## 3.1 Date of Submission
July 16th, 2016
#### 3.2 Submission Correspondent
Mr. Da Zeng Double Medical Technology Inc. No. 18, Shanbianhong East Road, Haicang District, Xiamen Fujian, China 361026 Tel: 86-592-6087078 Fax: 86-592-6587093 Email: zengda@Double-Engine.com; da52827@gmail.com
## 3.3 Proposed Device
Trade /Proprietary Name: Double Medical Universal Spine System Common Name: Pedicle screw spinal system Classification Name: Pedicle screw spinal system Device Classification: II Classification Panel: Orthopedic Panel Product Code: MNI, MNH Regulation Number: 21 CFR Part 888.3070
#### 3.4 Predicate Devices Information
Trauson General Spinal System (GSS) K082617
{4}------------------------------------------------
#### 3.5 Device Description
Double Medical Universal Spine System consists of pedicle screws, rods, transconnector and nut etc. It is made of Ti-6A1-4V ELI, which meets ASTM F 136 "Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications" and is widely used for surgical implants. The proposed devices are provided non-sterile but sterilized by the end user.
#### 3.6 Indication for Use/Intended Use
Double Medical Universal Spine System is intended for posterior, non-cervical, pedicle fixation for the following indications: severe spondylolisthesis (grade 3 or 4) of the L5-S1 vertebrae; trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scollosis, kyphosis, and/or lordosis), tumor, pseudarthrosis, and failed previous fusion. The device is to be used in skeletally mature patients, and for stabilization and immobilization of the spine as an adjunct to fusion with bone graft. The levels of fixation are T8 – S1.
#### 3.7 Technological Characteristics
Double Medical Universal Spine System components incorporate the same technological characteristics as the predicate device to stable the non-cervical posterior spine as an adjunct to fusion. At a high level, the subject and predicate device are based on the following same or equivalent technological elements:
- FDA Product Codes: MNI and MNH
- Equivalent intended uses
- Components are manufactured from Ti-6Al-4V ELI as per ASTM F 136
- Polyaxial and monoaxial pedicle screws are used to attach to the vertebrae
- Same pedicle screw thread form configuration
- Construct support mechanism
- Use of instrumentation to aid the placement of pedicle screws and rods ●
- Mechanical performance
#### 3.8 Non-Clinical Tests Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The following tests were performed per ASTM F 1717-14 "Standard Test Methods for Spinal Implant Constructs in a
{5}------------------------------------------------
Vertebrectomy Model":
- Static compression bending test
- Dynamic compression bending test ●
- Static torsion test. ●
And ASTM F 1798-13 "Standard Test Method for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants":
- Axial gripping capacity test ●
### 3.9 Substantially Equivalent Conclusion
The information submitted in this premarket notification has compared Double Medical Universal Spine System to the predicate device in regard to intended use, material, function, sizes and mechanical test results. These comparisons demonstrate sufficient evidence to conclude that Double Medical Universal Spine System is Substantially Equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.