CD HORIZON SPINAL SYSTEM

K050981 · Medtronic Sofamor Danek USA, Inc. · MNI · May 17, 2005 · Orthopedic

Device Facts

Record IDK050981
Device NameCD HORIZON SPINAL SYSTEM
ApplicantMedtronic Sofamor Danek USA, Inc.
Product CodeMNI · Orthopedic
Decision DateMay 17, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

The CD HORIZON® Spinal System is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion. When used in a percutaneous, non-cervical, posterior approach with the SEXTANT instrumentation, the CD HORIZON® screws are intended for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system, CD HORIZON® components such as ECLIPSE® components are intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) fracture, (6) pseudarthrosis; (7) tumor resection, and/or (8) failed previous fusion. The CD HORIZON® SPIRE Plate is posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1 - S1). It is intended for plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies; spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor. The CD HORIZON® LEGACY 3.5mm rod and associated components, when used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, are indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic impairment, (2) fracture, (3) dislocation, (4) scoliosis, (5) kyphosis, (6) spinal tumor, and/or (7) failed previous fusion (pseudarthrosis). In addition, when used as a pedicle screw fixation system, the CD HORIZON® LEGACY 3.5mm rod and associated components, are indicated for skeletally mature patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are receiving autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass. In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX™ Reconstruction System with the VERTEX™ rod connector. Refer to the VERTEX™ Reconstruction System Package Insert for a list of the VERTEX™ indications of use.

Device Story

Spinal stabilization system comprising rods, hooks, screws, plates, and connectors; used to build custom spinal constructs for immobilization/stabilization as adjunct to fusion. Components implanted by surgeons in OR; used in posterior, percutaneous, or anterolateral thoracic/lumbar approaches. System allows rigid locking of components in configurations tailored to individual patient anatomy. Can connect to VERTEX™ Reconstruction System. Benefits include spinal segment stabilization to facilitate fusion and address deformities or instability.

Clinical Evidence

No clinical data provided; substantial equivalence based on design and risk analysis.

Technological Characteristics

Spinal fixation system consisting of metallic rods, hooks, screws, and plates. Modular design allows for various construct configurations. Components are compatible with existing Medtronic spinal systems. Non-active, mechanical device.

Indications for Use

Indicated for non-cervical spinal stabilization/fixation in patients with degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal deformities (scoliosis, kyphosis, lordosis), tumor, pseudarthrosis, or failed previous fusion. Specific subsets indicated for skeletally mature patients with severe L5-S1 spondylolisthesis.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K050981 MAY 1 7 2005 # CD HORIZON® Spinal System Summary of Safety and Effectiveness April 2005 Medtronic Sofamor Danek, Inc. USA Company: I. 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133 > Richard W. Treharne, PhD Contact: Senior Vice President Regulatory Affairs - Proposed Proprietary Trade Name: CD HORIZON® Spinal System II. - Classification Name(s)/Product Code(s): Spinal Interlaminal Fixation and Spinal III. Intervertebral Fixation Orthosis and/or Pedicle Screw Spinal System (per 21 CFR Inter ver 888.3050, 888.3060 and/or 888.3070) Product Codes: MNI, MNH, KWP, KWQ and NKB ### Product Description IV. The CD HORIZON® Spinal System consists of a variety of rods, hooks, screws, CROSSLINK® plates, staples, and other connecting components used to build a spinal construct. Instrumentation is also available to facilitate implantation of the device components. The CD HORIZON® Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine. The CD HORIZON® Spinal System implant components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. If necessary, the CD Connegations System can be connected to the VERTEX™ Reconstruction System through a rod connector. Certain implant components from other Medtronic Sofamor Danek spinal systems can be used with the CD HORIZON® Spinal System. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples and washers; GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK PLUS® bolts; and Medtronic Sofamor Danek Multi-Axial rods and screws. CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORIZON® ECLIPSE® rods and screws are intended for anterior use only. However, for patients of smaller stature, CD HORIZON® 4.5mm rods and associated components may be used posteriorly. The purpose of this 510(k) submission is to add modified connecting components to the CD HORIZON® Spinal System. {1}------------------------------------------------ 2/2 #### Indications V. The CD HORIZON® Spinal System is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with 1010wing indivations: degentialed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion. When used in a percutaneous, non-cervical, posterior approach with the SEXTANT instrumentation, the CD HORIZON® screws are intended for the following indications: instrumentive disc disease (defined as back pain of discogenic origin with degeneration of the disc dogfired by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system, CD HORIZON® components such as ECLIPSE® components are intended for the following indications: (1) components such as DOD BEO veral by back pain of discogenic origin with degeneration of the disc degeneral ve unde alsease (a and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion. The CD HORIZON® SPIRE Plate is posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1 - S1). It is intended for plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies; spondylolisthesis; trauma (i.e., fracture or dislocation); and/or turnor. The CD HORIZON® LEGACY 3.5mm rod and associated components, when used as a pedicle The OD Iton system of the non-cervical posterior spine in skeletally mature patients, are indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic impairment, (2) fracture, (3) dislocation, (4) scoliosis, (5) kyphosis, (6) spinal tumor, and/or (7) failed previous fusion (pseudarthrosis). In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be in order to the VERTEX™ Reconstruction System with the VERTEX™ rod connector. Refer to the VERTEX™ Reconstruction System Package Insert for a list of the VERTEXTM indications of use. ### Substantial Equivalence VI. Documentation, including a Risk Analysis, was provided which demonstrated the subject Documents to be substantially equivalent to the previously cleared CD HORIZON® Spinal System connectors cleared in (K041862, K050079, K040583 AND K043488) for use for the identical indications. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with three lines representing the head, body, and legs. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA", which is arranged in a circular pattern around the figure. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 1 7 2005 Richard W. Treharne, Ph.D. Medtronic Sofamor Danek Incorporated 1800 Pyramid Place Memphis, Tennessee 38132 Re: K050981 Trade/Device Name: CD HORIZON Spinal System Trade/Dovies Number: 21 CFR 888.3070, 888.3050, 888.3060, Regulation Name: Pedicle Screw Spinal System, Spinal Interlaminal Fixation Orthosis, Spinal Intervertebral Body Fixation Orthosis Regulatory Class: III Rogalatory of: MNI, KWP, KWQ, MNH, NKB Dated: April 18, 2005 Received: April 19, 2005 Dear Dr. Treharne: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bector b r c(a) fee device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to regarly the enactment date of the Medical Device Amendments, or to commerce prior to May 20, 1978, in exact ance with the provisions of the Federal Food, DNA devices that have been require approval of a premarket approval application (PMA). and Cosment Act (Act) that to hot require appen of the general controls provisions of the Act. The You may, therefore, market the act include requirements for annual registration, listing of general controls provisions of use tice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (600 world). Existing major regulations affecting your device can may be subject to suen additions, Title 21, Parts 800 to 898. In addition, FDA may oc found in the Oous of reants concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I DA 3 issualite of two device complies with other requirements of the Act that FDA has made a decommances and regulations administered by other Federal agencies. You must or ally recetal statures and regaranents ancluding, but not limited to: registration and listing (21 comply with an the Hot 6 requirements) CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFK Fatt 807), falceinig (21 OFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Richard W. Treharne, Ph.D. This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to oegin manceinig your antial equivalence of your device to a legally premarket nothleadon. The PDA miang of easification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please r the contraction of the comment and and and on the recept of the recept le If you desire specific advice for your ac not of concerner and the regulation entitled, Conliact the Office of Comphaliso at (210) = 16 = 16 = 16 = 10 = 10 = 10 = 10 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 = 1 " Misoranding by relevelec to premarkst notification of Sct from the Division of Small on 1990 consisted information on your responsibilities under the Act from the 1990 Call other general information on your responsions Assistance at its toll-free number (800) 638-2041 or and index bird Manufacturers, Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Sincerely yours, Az. Miriam Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 April 2005 510(k) Number (if known): Device Name:_ CD HORIZON® Spinal System ## Indications for Use: The CD HORIZON® Spinal System is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion. When used in a percutaneous, non-cervical, posterior approach with the SEXTANT instrumentation, the CD HORIZON® screws are intended for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudoarthrosis; and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system, CD HORIZON® components such as ECLIPSE® components are intended for the following indications: (1) degenerative disc disease fas defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) fracture, (6) pseudarthrosis; (7) turnor resection, and/or (8) failed previous fusion. The CD HORIZON® SPIRE Plate is posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1 - S1). It is intended for plate fixation/attachment to spinous process for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies; spondylolisthesis; trauma (i.e., fracture or dislocation); and/or tumor. The CD HORIZON® LEGACY 3.5mm rod and associated components, when used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, are indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic imparment (2) fracture, (3) dislocation, (4) scoliosis, (5) kyphosis, (6) spinal tumor, and/or (7) failed previous fusion (pseudarthrosis). In addition, when used as a pedicle screw fixation system, the CD HORIZON® LEGACY 3.5mm rod and associated components, are indicated for skeletally mature patients: (a) having severe spondylolistlesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are receiving fusionesis autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacchil spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mpss. In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEXTM Reconstruction System with the VERTEX™ rod connector. Refer to the VERTEXTM Reconstruction System Package Insert for a list of the VERTEX™ indications of use, Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Divisions Signed for the below this line-continue on another page if needed Division of General, Restorative, ancurrence of CDRI **510(k) Number**
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...