INTRAMEDULLARY NAIL SYSTEM

K122994 · Weigao Orthopaedic Device Co., Ltd. · MNH · Apr 12, 2013 · Orthopedic

Device Facts

Record IDK122994
Device NameINTRAMEDULLARY NAIL SYSTEM
ApplicantWeigao Orthopaedic Device Co., Ltd.
Product CodeMNH · Orthopedic
Decision DateApr 12, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

The General Spinal System is intended for posterior, non-cervical, pedicle fixation for the following indications: severe spondylolisthesis (grade 3 or 4) of the L5-S1 vertebrae; trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The device is to be used in skeletally mature patients, and for stabilization and immobilization of the spine as an adjunct to fusion with bone graft. The levels of fixation are T8 - S1.

Device Story

General Spinal System is a posterior, non-cervical, pedicle fixation system; components include screws, rods, crosslink plates, set screws, and hooks. Device provides stabilization and immobilization of the spine as an adjunct to fusion with bone graft. Used in clinical settings by surgeons for skeletally mature patients. Components are manufactured from Titanium Alloy (Ti6Al4VELI) per ASTM F136-02a. System is provided non-sterile; requires hospital-based autoclave sterilization per ISO 17665-1:2006. Surgeon implants hardware to fixate T8-S1 spinal levels, facilitating fusion and patient stabilization.

Clinical Evidence

No clinical data. Bench testing only; compliance with ASTM F1717-04 standards for spinal implant constructs in a vertebrectomy model demonstrated mechanical equivalence.

Technological Characteristics

Materials: Titanium Alloy (Ti6Al4VELI) per ASTM F136-02a. Components: screws, rods, crosslink plates, set screws, hooks. Fixation levels: T8-S1. Sterilization: Autoclave (moist heat) per ISO 17665-1:2006. Mechanical testing: ASTM F1717-04 (static/dynamic compression bending, static torsion).

Indications for Use

Indicated for skeletally mature patients requiring posterior, non-cervical, pedicle fixation (T8-S1) for severe spondylolisthesis (grade 3-4, L5-S1), trauma (fracture/dislocation), spinal stenosis, curvatures (scoliosis/lordosis), tumor, pseudoarthrosis, or failed previous fusion; used for stabilization/immobilization as an adjunct to bone graft fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Exhibit #4 510(k) Summary *122994 - Page 1 of 3 ## 510(k) Summary # APR 1 2 2013 This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) Number: ... 1. Date of Submission: 04/03/2013 . 2. Sponsor Weigao Orthopaedic Device Co., Ltd. No 26 Xiangjiang Road, Tourist Resorts, Weihai, Shandong, 264203, China Establishment Registration Number: 3006639944 Contact Person: Han Wang Position: Quality & Technique Manager Tel: +86-631-5788966 Fax: +86-631-5660958 Email: wanghan@wegortho.com 3. Submission Correspondent > Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net {1}------------------------------------------------ - K122994 - 4. Proposed Device Identification Proposed Device Name: General Spinal System Classification Name: Pedicle screw spinal system Product Code: MNI, MNH, KWP Regulation Number: 21 CFR part 888.3070, 21 CFR part 888.3050 Review Panel: Orthopedic Intended Use Statement: The General Spinal System is intended for posterior, non-cervical, pedicle fixation for the following indications: severe spondylolisthesis (grade 3 or 4) of the L5-S1 vertebrae; trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The device is to be used in skeletally mature patients, and for stabilization and immobilization of the spine as an adjunct to fusion with bone graft. The levels of fixation are T8 - S1. - ડ. Predicate Device Identification 510(k) Number: K042790 Product Name: CD HORIZON ® Spinal System Predicate Device Name: CD HORIZON LEGACY 5.5mm Spinal System Manufacturer: Medtronic SofamorDanek, Inc.USA - 6. Device Description The spinal system consists of screws, rods, crosslink plates, set screws and hooks. It is made of Titanium Alloy (Ti6Al4VELI), which meets ASTM F136-02a, Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications, which are widely used for surgical implants with well known biocompatibility. The proposed devices are provided non-sterile. It is required to be sterilized via autoclave mèthod to reach a SAL of 10 by the hospital prior to surgery. The recommended sterilization method was validated per ISO 17665-1: 2006 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development. validation, and routine control of a sterilization process for medical devices {2}------------------------------------------------ - 7. Non-Clinical Test Conclusion Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: ASTM F1717-04, Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model, including the following items: - Static compression bending test; • - Dynamic compression bending test; - Static torsion test. . #### 8. Substantially Equivalent Conclusion The proposed device, Spinal System, is determined to be Substantially Equivalent (SE) to the predicate device, CD HORIZON LEGACY 5.5mm Spinal System (K042790), in respect of safety and effectiveness. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Letter dated: April 12, 2013 Weigao Orthopaedic Device Co., Ltd. % MID-LINK Consulting Co., Ltd. Ms. Diana Hong General Manager P.O. Box 237-023 200237 Shanghai China Re: K122994 Trade/Device Name: General Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH, KWP Dated: April 3, 2013 Received: April 8, 2013 Dear Ms. Hong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ #### Page 2 - Ms. Diana Hong device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark News Regson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Indications for Use 510(k) Number: K122994 Device Name: General Spinal System Indications for Use: The General Spinal System is intended for posterior, non-cervical, pedicle fixation for the following indications: severe spondylolisthesis (grade 3 or 4) of the L5-S1 vertebrae; trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The device is to be used in skeletally mature patients, and for stabilization and immobilization of the spine as an adjunct to fusion with bone graft. The levels of fixation are T8 - S1 ZPRESCRIPTION USE (Part 21 CFR 801 Subpart D) O OVER-THE-COUNTER USE (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Image /page/5/Picture/11 description: The image shows the name "Colin O'Neill" in a stylized font. The letters "C", "O", "N", "e", "i", "l", and "l" are in a standard font. The letters "l", "i", "O", and "N" are stylized with geometric shapes and lines. The name is written in black on a white background. (Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K122994
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...